Literature DB >> 21116893

Decreased accuracy of acetabular cup placement for imageless navigation in obese patients.

Sachiyuki Tsukada1, Motohiro Wakui.   

Abstract

BACKGROUND: Optimal acetabular cup position is an important determinant of the success of total hip arthroplasty (THA), and navigation systems have been developed and applied to improve placement precision. However, the registration method touching bony landmarks through soft tissues may decrease the accuracy in obese cases and in patients with acetabular dysplasia.
METHODS: Placement accuracy in 69 joints in which THA was performed with imageless navigation was calculated by comparing the placement angle in the anterior pelvic plane and the placement angle in the x-ray table plane with the patient in the supine position. We defined the difference between the placement angle in the anterior pelvic plane and the angle displayed on the navigation screen as the error and the difference between the placement angle in the plane of the x-ray table and the target angle of acetabular component position as the difference in target angle. Patients were divided into obese (BMI ≥ 25) and nonobese (BMI < 25) groups and into acetabular dysplasia and nondysplasia groups.
RESULTS: The mean ± SD navigation errors for all cases were 2.4° ± 2.0° for inclination and 3.7° ± 2.3° for anteversion. The mean difference in target angle for all cases was 2.8° ± 2.5° for inclination and 4.2° ± 3.0° for anteversion. The error in anteversion was significantly higher in the obese group (4.8° ± 2.5°) than in the nonobese group (3.2° ± 2.6°) (P = 0.01). No significant difference was observed between the acetabular dysplasia and nondysplasia groups.
CONCLUSIONS: Imageless navigation can be performed with an error of approximately 3° at the anterior pelvic plane. However, accuracy was found to decrease in obese cases.

Entities:  

Mesh:

Year:  2010        PMID: 21116893     DOI: 10.1007/s00776-010-1546-1

Source DB:  PubMed          Journal:  J Orthop Sci        ISSN: 0949-2658            Impact factor:   1.601


  14 in total

1.  Evaluation of the accuracy of femoral component orientation by the CT-based fluoro-matched navigation system.

Authors:  Shinya Hayashi; Takayuki Nishiyama; Takaaki Fujishiro; Shingo Hashimoto; Noriyuki Kanzaki; Kotaro Nishida; Ryosuke Kuroda; Masahiro Kurosaka
Journal:  Int Orthop       Date:  2013-03-20       Impact factor: 3.075

Review 2.  [Total hip arthroplasty in overweight osteoarthritis patients].

Authors:  M Stiehler; J Goronzy; K-P Günther
Journal:  Orthopade       Date:  2015-07       Impact factor: 1.087

3.  [Analysis of rising treatment cost of elevated BMI in patients with proximal femoral fracture].

Authors:  Alexander Gutwerk; Michael Müller; Moritz Crönlein; Chlodwig Kirchhoff; Peter Biberthaler; Dominik Pförringer; Karl Braun
Journal:  Unfallchirurgie (Heidelb)       Date:  2022-05-23

4.  Abnormal spinopelvic mobility as a risk factor for acetabular placement error in total hip arthroplasty using optical computer-assisted surgical navigation system.

Authors:  Seong J Jang; Jonathan M Vigdorchik; Eric W Windsor; Ran Schwarzkopf; David J Mayman; Peter K Sculco
Journal:  Bone Jt Open       Date:  2022-06

5.  Total hip arthroplasty via an anterolateral supine approach for obese patients increases the risk of greater trochanteric fracture.

Authors:  Hirokazu Iwata; Kosuke Sakata; Eiji Sogo; Katsuhiko Nanno; Sanae Kuroda; Tsuyoshi Nakai
Journal:  J Orthop       Date:  2018-03-17

6.  Obesity does not influence acetabular component accuracy when using a 3D optical computer navigation system.

Authors:  Mohamad Sharan; Alex Tang; Lauren Schoof; Alexander Gaukhman; Morteza Meftah; Peter Sculco; Ran Schwarzkopf
Journal:  J Clin Orthop Trauma       Date:  2020-10-03

7.  Lower Dislocation Rate Following Total Hip Arthroplasty via Direct Anterior Approach than via Posterior Approach: Five-Year-Average Follow-Up Results.

Authors:  Sachiyuki Tsukada; Motohiro Wakui
Journal:  Open Orthop J       Date:  2015-05-15

8.  Accuracy of Component Orientation and Leg Length Adjustment in Total Hip Arthroplasty Using Image-free Navigation.

Authors:  Yu Takeda; Shigeo Fukunishi; Shoji Nishio; Yuki Fujihara; Shinichi Yoshiya
Journal:  Open Orthop J       Date:  2017-12-28

Review 9.  Robotic-assisted compared with conventional total hip arthroplasty: systematic review and meta-analysis.

Authors:  Xi Chen; Jianping Xiong; Peipei Wang; Shibai Zhu; Wenting Qi; Huiming Peng; Lingjia Yu; Wenwei Qian
Journal:  Postgrad Med J       Date:  2018-05-18       Impact factor: 2.401

10.  Correlation of tilt of the anterior pelvic plane angle with anatomical pelvic tilt and morphological configuration of the acetabulum in patients with developmental dysplasia of the hip: a cross-sectional study.

Authors:  Norio Imai; Hayato Suzuki; Asami Nozaki; Yuki Hirano; Naoto Endo
Journal:  J Orthop Surg Res       Date:  2019-10-17       Impact factor: 2.359