Literature DB >> 21099578

Retrograde trans-posterior communicating artery snare-assisted rescue of lost access to a foreshortened pipeline embolization device: complication management.

Erik F Hauck1, Sabareesh K Natarajan, David J Langer, L Nelson Hopkins, Adnan H Siddiqui, Elad I Levy.   

Abstract

BACKGROUND AND IMPORTANCE: The Pipeline embolization device (PED; Covidien Vascular Therapies, Mansfield, Massachusetts) is a promising, yet experimental, vascular reconstruction device for the treatment of complex intracranial aneurysms. We present a PED-related complication and describe a salvage strategy. CLINICAL
PRESENTATION: A 64-year-old woman underwent PED-assisted parent vessel reconstruction for her giant cavernous internal carotid artery (ICA) aneurysm. During placement of the first PED, the proximal part of the PED foreshortened and was displaced into the aneurysm sac. Multiple subsequent attempts to recatheterize the PED failed, and, ultimately, distal access through and beyond the PED was lost. Therefore, completion of the Pipeline construct by stacking PEDs for definitive treatment was prevented. Retrograde access of the PED was gained from the distal ICA through a microwire that was advanced from the basilar artery through the posterior communicating artery. The microwire from the distal ICA was grasped with a snare from the proximal ICA and pulled down to the cervical ICA. The opened snare around the microwire was used as a lasso to advance a microcatheter from the cervical ICA through the PED to regain distal access. Five more PEDs were used to achieve complete parent vessel reconstruction and aneurysm obliteration.
CONCLUSION: Maintaining distal access is critical until the entire parent vessel is reconstructed, especially when multiple PEDs are required. The salvage technique described may help regain distal access if it is lost during the procedure.

Entities:  

Mesh:

Year:  2010        PMID: 21099578     DOI: 10.1227/NEU.0b013e3181f8530d

Source DB:  PubMed          Journal:  Neurosurgery        ISSN: 0148-396X            Impact factor:   4.654


  13 in total

1.  Computation of the change in length of a braided device when deployed in realistic vessel models.

Authors:  Hector Fernandez; Juan M Macho; Jordi Blasco; Luis San Roman; Werner Mailaender; Luis Serra; Ignacio Larrabide
Journal:  Int J Comput Assist Radiol Surg       Date:  2015-06-11       Impact factor: 2.924

Review 2.  Treatment of direct carotid-cavernous fistulas with flow diversion - does it work?

Authors:  Zeynep Hüseyinoglu; Marvin D Oppong; Andrew S Griffin; Erik Hauck
Journal:  Interv Neuroradiol       Date:  2018-10-31       Impact factor: 1.610

3.  What's coming down the pipe--and should we be excited, concerned, or both?

Authors:  J L Brisman
Journal:  AJNR Am J Neuroradiol       Date:  2012-11-22       Impact factor: 3.825

4.  Deployment of flow diverter devices: prediction of foreshortening and validation of the simulation in 18 clinical cases.

Authors:  Robert Kellermann; Steffen Serowy; Oliver Beuing; Martin Skalej
Journal:  Neuroradiology       Date:  2019-08-31       Impact factor: 2.804

5.  Retrograde trans-anterior communicating artery rescue of unopened Pipeline Embolization Device with balloon dilation: complication management.

Authors:  Ramon Navarro; Jang Yoon; Tanya Dixon; David A Miller; Ricardo A Hanel; Rabih G Tawk
Journal:  BMJ Case Rep       Date:  2014-01-27

6.  Spontaneous delayed migration/shortening of the pipeline embolization device: report of 5 cases.

Authors:  N Chalouhi; S I Tjoumakaris; L F Gonzalez; D Hasan; P J Pema; G Gould; R H Rosenwasser; P M Jabbour
Journal:  AJNR Am J Neuroradiol       Date:  2013-06-27       Impact factor: 3.825

7.  Novel balloon application for rescue and realignment of a proximal end migrated pipeline flex embolization device into the aneurysmal sac: complication management.

Authors:  Mario Martínez-Galdámez; Joaquin Ortega-Quintanilla; Antonio Hermosín; Eduardo Crespo-Vallejo; Juan José Ailagas; Santiago Pérez
Journal:  BMJ Case Rep       Date:  2016-03-23

8.  Salvage of Herniated Flow Diverters Using Stent and Balloon Anchoring Techniques: A Technical Note.

Authors:  Bharathi D Jagadeesan; Divyajot Sandhu; Kim Jae Hong; Sara Roshan Khanipour; Andrew W Grande; Ramachandra P Tummala
Journal:  Interv Neurol       Date:  2016-11-11

9.  Treatment of carotid cavernous sinus fistulas with flow diverters. A case report and systematic review.

Authors:  Theodosios Stamatopoulos; Evangelos Anagnostou; Sotirios Plakas; Konstantinos Papachristou; Panagiotis Lagos; Apostolos Samelis; Shahram Derakhshani; Aristotelis Mitsos
Journal:  Interv Neuroradiol       Date:  2021-05-10       Impact factor: 1.610

Review 10.  Bailout Strategies and Complications Associated with the Use of Flow-Diverting Stents for Treating Intracranial Aneurysms.

Authors:  Fawaz Al-Mufti; Eric R Cohen; Krishna Amuluru; Vikas Patel; Mohammad El-Ghanem; Rolla Nuoman; Neil Majmundar; Neha S Dangayach; Philip M Meyers
Journal:  Interv Neurol       Date:  2018-10-16
View more