Literature DB >> 21084576

Clinical and radiographic results of metal-on-metal hip resurfacing with a minimum ten-year follow-up.

Harlan C Amstutz1, Michel J Le Duff, Patricia A Campbell, Thomas A Gruen, Lauren E Wisk.   

Abstract

BACKGROUND: There was a need for information about the long-term performance of the modern generation of hip resurfacing implants. A retrospective review of the first 100 hips that had implantation of one resurfacing design and had been followed for a minimum of ten years was performed.
METHODS: Between 1996 and 1998, 100 CONSERVE PLUS hip resurfacing devices were implanted by a single surgeon in eighty-nine patients. The mean age of the patients was 49.1 years, and fifty-nine patients were male. Primary osteoarthritis was the etiology for sixty-four hips. All patients were assessed clinically and radiographically.
RESULTS: The mean duration of follow-up was 11.7 years (range, 10.8 to 12.9 years). Two patients were lost to follow-up, and five patients died of causes unrelated to the surgery. Eleven hips had conversion to total hip arthroplasty because of loosening of the femoral component (eight), a femoral neck fracture (one), recurrent subluxation (one), and late infection (one). The Kaplan-Meier survivorship was 88.5% at ten years. None of the resurfacing arthroplasties failed in the twenty-eight hips that had a femoral component of >46 mm and no femoral head cystic or necrotic defects of >1 cm. Five hips had narrowing of the femoral neck, three had radiolucent zones interpreted as osteolysis, and twenty had signs of neck-socket impingement. Five hips had radiolucencies around the metaphyseal stem (two partial and three complete) that had been stable for 7.8 to 10.2 years. The mean scores on the University of California at Los Angeles (UCLA) system at the time of the latest follow-up were 9.5 points for pain, 9.3 points for walking, 8.9 points for function, and 6.8 points for activity; the mean scores on the physical and mental components of the Short Form-12 (SF-12) were 47.3 and 50.5 points, respectively; and the mean Harris hip score was 90 points.
CONCLUSIONS: The results of the present series constitute a reference point to which subsequent series should be compared. These ten-year results in a group of young patients are satisfactory, and the low rate of osteolysis is encouraging, but longer follow-up is required for comparison with conventional total hip arthroplasty.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 21084576     DOI: 10.2106/JBJS.I.01715

Source DB:  PubMed          Journal:  J Bone Joint Surg Am        ISSN: 0021-9355            Impact factor:   5.284


  24 in total

1.  Management of failed metal-on-metal total hip arthroplasty.

Authors:  Justin W Griffin; Michele D'Apuzzo; James A Browne
Journal:  World J Orthop       Date:  2012-06-18

2.  [Approach to painful hip resurfacing].

Authors:  L Gerdesmeyer; H Gollwitzer; P Diehl; M Fuerst; M Schmitt-Sody
Journal:  Orthopade       Date:  2011-06       Impact factor: 1.087

3.  Do revised hip resurfacing arthroplasties lead to outcomes comparable to those of primary and revised total hip arthroplasties?

Authors:  William Desloges; Isabelle Catelas; Toru Nishiwaki; Paul R Kim; Paul E Beaulé
Journal:  Clin Orthop Relat Res       Date:  2012-11       Impact factor: 4.176

Review 4.  The future role of metal-on-metal hip resurfacing.

Authors:  Gulraj S Matharu; Hemant G Pandit; David W Murray; Ronan B C Treacy
Journal:  Int Orthop       Date:  2015-02-24       Impact factor: 3.075

5.  The First SICOT Oral Presentation Award 2011: imageless computer-assisted femoral component positioning in hip resurfacing: a prospective randomised trial.

Authors:  Maik Stiehler; Jens Goronzy; Albrecht Hartmann; Frank Krummenauer; Klaus-Peter Günther
Journal:  Int Orthop       Date:  2013-02-06       Impact factor: 3.075

6.  Patients report improvement in quality of life and satisfaction after hip resurfacing arthroplasty.

Authors:  Wael A Rahman; Nelson V Greidanus; Alexander Siegmeth; Bassam A Masri; Clive P Duncan; Donald S Garbuz
Journal:  Clin Orthop Relat Res       Date:  2013-02       Impact factor: 4.176

Review 7.  Metal-on-metal hip resurfacing arthroplasty: an analysis of safety and revision rates.

Authors:  S Sehatzadeh; K Kaulback; L Levin
Journal:  Ont Health Technol Assess Ser       Date:  2012-08-01

8.  Metal-on-metal hip resurfacing: correlation between clinical and radiological assessment, metal ions and ultrasound findings.

Authors:  M Scaglione; L Fabbri; N Bianchi; D Dell'Omo; G Guido
Journal:  Musculoskelet Surg       Date:  2014-12-24

9.  Conversion of a failed hip resurfacing arthroplasty to total hip arthroplasty: pearls and pitfalls.

Authors:  Jacob A Haynes; Jeffrey B Stambough; Robert L Barrack; Denis Nam
Journal:  Curr Rev Musculoskelet Med       Date:  2016-03

10.  Prospective study on serum metal levels in patients with metal-on-metal lumbar disc arthroplasty.

Authors:  Matthew F Gornet; J K Burkus; M L Harper; F W Chan; A K Skipor; J J Jacobs
Journal:  Eur Spine J       Date:  2012-11-20       Impact factor: 3.134

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.