Literature DB >> 2105795

Measuring blood pressure in the elderly: does atrial fibrillation increase observer variability?

D Sykes1, R Dewar, K Mohanaruban, K Donovan, F Nicklason, D M Thomas, D Fisher.   

Abstract

OBJECTIVE: To compare the interobserver and intraobserver variability of blood pressure measurements in geriatric patients in atrial fibrillation and in sinus rhythm.
DESIGN: Prospective assessment of blood pressure measurements carried out in random order in two groups of elderly patients by five doctors unaware of the aims of the study.
SETTING: Acute assessment wards for geriatric medicine, Cardiff Royal Infirmary. PATIENTS: 50 Elderly patients in sinus rhythm and 50 in atrial fibrillation. MAIN OUTCOME MEASURES: Interobserver and intraobserver variability of blood pressure measurements in the two groups expressed as the coefficient of variability and compared by the Mann-Whitney U test.
RESULTS: Interobserver variability was significantly greater in the patients with atrial fibrillation for both systolic and diastolic pressures. Intraobserver variability was significantly greater in the atrial fibrillation group for diastolic pressures but the difference was not significant for systolic pressures. These differences were not related to pulse rate, age, or level of blood pressure.
CONCLUSIONS: The findings suggest that in the presence of atrial fibrillation physicians' interpretations of Korotkoff sounds are less uniform, which may have important clinical implications. Possibly a standardised methodology may overcome this problem.

Entities:  

Mesh:

Year:  1990        PMID: 2105795      PMCID: PMC1662156          DOI: 10.1136/bmj.300.6718.162

Source DB:  PubMed          Journal:  BMJ        ISSN: 0959-8138


  8 in total

1.  Recommendations on blood pressure measurement.

Authors:  J C Petrie; E T O'Brien; W A Littler; M de Swiet
Journal:  Br Med J (Clin Res Ed)       Date:  1986-09-06

Review 2.  Blood pressure measurement: current practice and future trends.

Authors:  E O'Brien; D Fitzgerald; K O'Malley
Journal:  Br Med J (Clin Res Ed)       Date:  1985-03-09

3.  Observer error in systolic blood pressure measurement in the elderly. A case for automatic recorders?

Authors:  K M Hla; K A Vokaty; J R Feussner
Journal:  Arch Intern Med       Date:  1986-12

4.  Sources of error in recording the blood pressure of patients with hypertension in general practice.

Authors:  H R Patterson
Journal:  Br Med J (Clin Res Ed)       Date:  1984-12-15

5.  Mortality and morbidity results from the European Working Party on High Blood Pressure in the Elderly trial.

Authors:  A Amery; W Birkenhäger; P Brixko; C Bulpitt; D Clement; M Deruyttere; A De Schaepdryver; C Dollery; R Fagard; F Forette
Journal:  Lancet       Date:  1985-06-15       Impact factor: 79.321

6.  Epidemiologic assessment of chronic atrial fibrillation and risk of stroke: the Framingham study.

Authors:  P A Wolf; T R Dawber; H E Thomas; W B Kannel
Journal:  Neurology       Date:  1978-10       Impact factor: 9.910

7.  Inappropriate antihypertensive therapy in the elderly.

Authors:  G Jackson; T A Pierscianowski; W Mahon; J Condon
Journal:  Lancet       Date:  1976-12-18       Impact factor: 79.321

8.  Observer bias in blood pressure studies.

Authors:  N G Bruce; A G Shaper; M Walker; G Wannamethee
Journal:  J Hypertens       Date:  1988-05       Impact factor: 4.844

  8 in total
  12 in total

1.  Measuring blood pressure in the elderly.

Authors:  I W Grant
Journal:  BMJ       Date:  1990-02-24

2.  Non-invasive blood pressure monitoring with an oscillometric brachial cuff: impact of arrhythmia.

Authors:  Karim Lakhal; Maëlle Martin; Stephan Ehrmann; Sofian Faiz; Bertrand Rozec; Thierry Boulain
Journal:  J Clin Monit Comput       Date:  2017-09-27       Impact factor: 2.502

Review 3.  Blood Pressure Assessment in Adults in Clinical Practice and Clinic-Based Research: JACC Scientific Expert Panel.

Authors:  Paul Muntner; Paula T Einhorn; William C Cushman; Paul K Whelton; Natalie A Bello; Paul E Drawz; Beverly B Green; Daniel W Jones; Stephen P Juraschek; Karen L Margolis; Edgar R Miller; Ann Marie Navar; Yechiam Ostchega; Michael K Rakotz; Bernard Rosner; Joseph E Schwartz; Daichi Shimbo; George S Stergiou; Raymond R Townsend; Jeff D Williamson; Jackson T Wright; Lawrence J Appel
Journal:  J Am Coll Cardiol       Date:  2019-01-29       Impact factor: 24.094

4.  Confounders of auscultatory blood pressure measurement.

Authors:  R H Baker; J Ende
Journal:  J Gen Intern Med       Date:  1995-04       Impact factor: 5.128

5.  Accuracy of automated blood pressure measurements in the presence of atrial fibrillation: systematic review and meta-analysis.

Authors:  Christopher E Clark; Sinead T J McDonagh; Richard J McManus
Journal:  J Hum Hypertens       Date:  2019-01-10       Impact factor: 3.012

6.  Blood Pressure in Patients With Atrial Fibrillation: Part 1--Measurement.

Authors:  Debbie L Cohen; Raymond R Townsend
Journal:  J Clin Hypertens (Greenwich)       Date:  2016-08-20       Impact factor: 3.738

7.  Atrial fibrillation influences automatic oscillometric ankle-brachial index measurement.

Authors:  Michał Dąbrowski; Jacek Lewandowski; Cezary Szmigielski; Maciej Siński
Journal:  Arch Med Sci       Date:  2019-07-12       Impact factor: 3.318

Review 8.  Techniques for Non-Invasive Monitoring of Arterial Blood Pressure.

Authors:  Agnes S Meidert; Bernd Saugel
Journal:  Front Med (Lausanne)       Date:  2018-01-08

9.  Comparing volume-clamp method and intra-arterial blood pressure measurements in patients with atrial fibrillation admitted to the intensive or medium care unit.

Authors:  G F N Berkelmans; S Kuipers; B E Westerhof; A M E Spoelstra-de Man; Y M Smulders
Journal:  J Clin Monit Comput       Date:  2017-07-07       Impact factor: 2.502

10.  Blood pressure measurement in special populations and circumstances.

Authors:  George S Stergiou; Eamon Dolan; Anastasios Kollias; Neil R Poulter; Andrew Shennan; Jan A Staessen; Zhen-Yu Zhang; Michael A Weber
Journal:  J Clin Hypertens (Greenwich)       Date:  2018-07       Impact factor: 3.738

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.