Yi Yang1, Dongyu Gu, Haji Akber Aisa, Yoichiro Ito.
Abstract
Optimal positions of coiled separation columns on the type-I centrifuge were determined for four typical two-phase solvent systems to obtain the best separation efficiency (resolution and retention of stationary phase) for each with a suitable set of test samples. A set of short coiled columns is connected in series and mounted around the holder hub in four different ways: (model A) the tail of one unit with left-handedness was connected to the head of the next unit with right-handedness (TL-HR); (model B) the tail of one unit with left-handedness was connected to the tail of the next unit with right-handedness (TL-TR); (model C) the tail of one unit with left-handedness was connected to the tail of the next unit with left-handedness (TL-TL); (model D) the tail of one unit with left-handedness was connected to the head of the next unit with left-handedness (TL-HL). The results indicated that the performance of model D was the best among the four models. High revolution speed (800 rpm) is favorable to separation using the moderately hydrophobic solvent system of hexane-ethyl acetate-methanol-0.1M HCl (1:1:1:1, v/v) (HEMW), while lower revolution speed (600 rpm) is beneficial to the separation with polar solvent system of 1-butanol-acetic acid-water (19:1:20, v/v) (BAW). Published by Elsevier B.V.Entities:
Mesh:
Substances:
Year: 2010 PMID: 21030029 PMCID: PMC2975402 DOI: 10.1016/j.chroma.2010.10.012
Source DB: PubMed Journal: J Chromatogr A ISSN: 0021-9673 Impact factor: 4.759