| Literature DB >> 20971721 |
Alan M Groves1, Gaia Chiesa, Giuliana Durighel, Stephen T Goldring, Julie A Fitzpatrick, Sergio Uribe, Reza Razavi, Jo V Hajnal, A David Edwards.
Abstract
OBJECTIVE: To use cardiac MRI techniques to assess ventricular function and systemic perfusion in preterm and term newborns, to compare techniques to echocardiographic methods, and to obtain initial reference data.Entities:
Mesh:
Year: 2010 PMID: 20971721 PMCID: PMC3093932 DOI: 10.1136/adc.2010.189142
Source DB: PubMed Journal: Arch Dis Child Fetal Neonatal Ed ISSN: 1359-2998 Impact factor: 5.747
Figure 1Four chamber (A) and short axis (B) views obtained with balanced fast field echo sequences in a newborn infant (view also video links at http://adc.bmj.com).
Figure 2Regression plot of external validation of phase contrast flow quantification versus flow phantom. Line of unity shown.
Phase contrast measures of left ventricular output (LVO), right ventricular output (RVO) and superior vena caval (SVC) and descending aortic (DAo) flow in 28 stable newborn infants
| LVO (ml/kg/min) | RVO (ml/kg/min) | SVC flow (ml/kg/min) | DAo flow (ml/kg/min) | SVC+DAo (ml/kg/min) | |
|---|---|---|---|---|---|
| Mean | 222 | 219 | 95 | 126 | 224 |
| SD | 45.8 | 46.9 | 27.0 | 32.1 | 51.2 |
Figure 3Normal ranges by corrected gestational age at scan for phase contrast assessment of left ventricular output (A), right ventricular output (B), superior vena caval flow (C) and descending aortic flow (D).
Repeatability of phase contrast measures of flow in newborn infants
| LVO | RVO | SVC | DAo | |
|---|---|---|---|---|
| n | 26 | 9 | 24 | 21 |
| Limits of agreement (ml/kg/min) | 50.2 | 55.5 | 20.9 | 26.2 |
| Repeatability index | 22.2% | 22.5% | 20.0% | 19.4% |
Figure 4Bland-Altman plots of scan-rescan repeatability of assessment of left ventricular output performed by phase contrast (A) and cine (B) imaging, and of comparison of phase contrast and echo measures of left ventricular output (C).
Left ventricular indices taken from stacks of cine images in 75 stable newborn infants
| LV EDV (ml/kg) | LV ESV (ml/kg) | LV SV (ml/kg) | LV EF (%) | HR (bpm) | LVO (ml/kg/min) | |
|---|---|---|---|---|---|---|
| Mean | 2.51 | 0.78 | 1.73 | 69.0 | 142 | 245 |
| SD | 0.40 | 0.21 | 0.28 | 5.7 | 17.1 | 47.1 |
EDV, end-diastolic volume; EF, ejection fraction; ESV, end-systolic volume; HR, heart rate; LV, left ventricle; LVO, left ventricular output; SV, stroke volume.
Figure 5Normal ranges by corrected gestational age at scan for bFFE assessment of left ventricular output (A), end-diastolic volume (B), end-systolic volume (C) and ejection fraction (D).
Repeatability of left ventricular indices in 22 newborn infants
| EDV | ESV | SV | EF | LVO | |
|---|---|---|---|---|---|
| Limits of agreement | 1.07 ml | 0.48 ml | 0.88 ml | 7.4% | 58.3 (ml/kg/min) |
| Repeatability index | 19.4% | 27.7% | 23.1% | 10.8% | 22.1% |
EDV, end-diastolic volume; EF, ejection fraction; ESV, end-systolic volume; LVO, left ventricular output; SV, stroke volume.