Literature DB >> 20963545

Experimental analysis of torque characteristics of orthodontic wires.

Samira Partowi1, Ludger Keilig, Susanne Reimann, Andreas Jäger, Christoph Bourauel.   

Abstract

BACKGROUND AND
OBJECTIVE: A major requirement for orthodontic wires is superior elasticity. The question arises whether today's nickel titanium wires display a distinctively superelastic plateau also under torsional loading, and whether typical torsion angles are sufficient to produce adequate torquing moments with superelastic wires. The objective of this study was to compare various nickel titanium and steel wires in a pure torsion experiment.
MATERIAL AND METHODS: To experimentally examine the torque characteristic of orthodontic wires, we used nickel titanium wires manufactured by Dentaurum (Tensic®) and ODS (Euro Arch NiTi) measuring 0.40 × 0.40 mm2, 0.40 × 0.56 mm2, 0.43 × 0.64 mm2, 0.46 × 0.64 mm2, 0.48 × 0.64 mm2, 0.51 × 0.51 mm2, and steel wires made by Dentaurum (remanium®), 3M Unitek (HI-T™ II Stainless Steel) measuring 0.40 × 0.40 mm2, 0.40 × 0.56 mm2, 0.43 × 0.56 mm2, 0.43 × 0.64 mm2, 0.46 × 0.64 mm2, 0.48 × 0.64 mm2, 0.51 × 0.51 mm2. Torque was measured at an ambient temperature of 37 ° C for the nickel titanium wires. The distance between the bearing points was 10 mm for all measurements. The steel wires' torquing moments were determined at a torsion angle of 20 °, the torquing moments of the nickel titanium wires were investigated on the plateau, and we compared the results of the means and the standard deviations of all wires. Starting and end points of the loading and unloading plateau were determined for all nickel titanium wires.
RESULTS: The torquing moment/torsion angle diagrams of the nickel titanium wires by different manufacturers displayed the curves typical of superelastic wires. The torquing moments of the wires with a smaller cross-section yielded values below 5 Nmm and thus appear obviously too small (e.g. Dentaurum Tensic®, 0.40 × 0.40 mm2: 1.3 Nmm). The mean values of the loading plateau's starting point yielded values of about 20 °.
CONCLUSIONS: As the torsional play in the bracket slot of the wires with a smaller cross-section is typically about 10 °, we doubt whether these wires reach the superelastic plateau. Moreover, we observed production-associated variations in the material properties of various nickel titanium wires. It is thus difficult for the practitioner to draw a correlation between a wire's crosssection and the actual torquing moment delivered.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20963545     DOI: 10.1007/s00056-010-1028-2

Source DB:  PubMed          Journal:  J Orofac Orthop        ISSN: 1434-5293            Impact factor:   1.938


  11 in total

1.  [Comparison of mechanical properties of orthodontic nickel-titanium wires].

Authors:  D Kayser; C Bourauel; B Braumann; A Jäger
Journal:  Biomed Tech (Berl)       Date:  2002-12       Impact factor: 1.411

2.  [Biomechanical analysis of orthodontic brackets with different closing mechanisms].

Authors:  Enver Morina; Ludger Keilig; Andreas Jäger; Christoph Bourauel
Journal:  Biomed Tech (Berl)       Date:  2009-04       Impact factor: 1.411

3.  [The materials engineering characteristics of orthodontic nickel-titanium wires].

Authors:  D Drescher; C Bourauel; M Thier
Journal:  Fortschr Kieferorthop       Date:  1990-12

4.  An evaluation of 55 cobalt substituted Nitinol wire for use in orthodontics.

Authors:  G F Andreasen; T B Hilleman
Journal:  J Am Dent Assoc       Date:  1971-06       Impact factor: 3.634

5.  [A computer-controlled flexing test for determining the elastic parameters of highly flexible orthodontic wires].

Authors:  R Plietsch; C Bourauel; D Drescher; B Nellen
Journal:  Fortschr Kieferorthop       Date:  1994-04

6.  Properties of superelastic wires and their relevance to orthodontic treatment.

Authors:  D Segner; D Ibe
Journal:  Eur J Orthod       Date:  1995-10       Impact factor: 3.075

7.  [The computer-aided development of orthodontic treatment elements made from NiTi memory alloys exemplified by a pseudoelastic retraction spring].

Authors:  C Bourauel; D Drescher; L P Nolte
Journal:  Fortschr Kieferorthop       Date:  1993-02

8.  [The long-term fracture resistance of orthodontic nickel-titanium wires].

Authors:  D Drescher; C Bourauel; W Sonneborn; G P Schmuth
Journal:  Schweiz Monatsschr Zahnmed       Date:  1994

9.  [Torque characteristics of steel and nitinol wires].

Authors:  J Bachmann
Journal:  Fortschr Kieferorthop       Date:  1983-08

10.  Torque expression of self-ligating brackets compared with conventional metallic, ceramic, and plastic brackets.

Authors:  Enver Morina; Theodore Eliades; Nikolaos Pandis; Andreas Jäger; Christoph Bourauel
Journal:  Eur J Orthod       Date:  2008-06       Impact factor: 3.075

View more
  1 in total

1.  Torque efficiency of square and rectangular archwires into 0.018 and 0.022 in. conventional brackets.

Authors:  Spyridon N Papageorgiou; Iosif Sifakakis; Ioannis Doulis; Theodore Eliades; Christoph Bourauel
Journal:  Prog Orthod       Date:  2016-01-15       Impact factor: 2.750

  1 in total