| Literature DB >> 20963537 |
Yuan Xu1, Sean Hayes, Iftikhar Ali, Terrence D Ruddy, R Glenn Wells, Daniel S Berman, Guido Germano, Piotr J Slomka.
Abstract
BACKGROUND: We define the repeatability coefficients (RC) of key quantitative and visual perfusion and function parameters that can be derived by the QGS/QPS automated software and by expert visual observer from gated myocardial perfusion SPECT (MPS) scans.Entities:
Mesh:
Year: 2010 PMID: 20963537 PMCID: PMC2990014 DOI: 10.1007/s12350-010-9297-0
Source DB: PubMed Journal: J Nucl Cardiol ISSN: 1071-3581 Impact factor: 5.952
Study demographics (N = 100)
| Patient characteristic | Values |
|---|---|
| Gender (Male) | 74 |
| Age (mean ± SD) | 56 ± 11 |
| BMI (mean ± SD) | 28.3 ± 4.0 |
| Diabetes (N) | 14 |
| Hypertension (N) | 58 |
| Dyslipidemia (N) | 56 |
| Smoking (N) | 54 |
| Family history (N) | 41 |
| Pretest likelihood of CAD | |
| Low risk (<10%) | 20 |
| Moderate risk (10%–90%) | 72 |
| High risk (>90%) | 8 |
| Previous event (MI/PCI/CABG) | 20/25/8 |
Note: MI, Myocardial infarction; PCI, percutaneous coronary intervention; CABG, coronary artery bypass graft.
Global (A) and regional (B) reproducibility of perfusion parameters between test–retest studies in each vascular territory
| Bias ± SD | 95% Limits of agreement | Correlation | RC | |
|---|---|---|---|---|
| (A) Global | ||||
| VSSS | 0.0 ± 1.7 | −3.3, 3.3 | 0.95 | 3.3 |
| VSRS | −0.1 ± 1.3 | −2.7, 2.5 | 0.95 | 2.6 |
| VSDS | 0.08 ± 1.5 | −2.8, 3.0 | 0.81 | 2.9 |
| VSSS% | 0.01 ± 2.5 | −4.8, 4.8 | 0.95 | 4.8 |
| VSRS% | −0.10 ± 1.9 | −3.9, 3.7 | 0.95 | 3.8 |
| VSDS% | 0.12 ± 2.2 | −4.2, 4.4 | 0.81 | 4.3 |
| STPD (%) | 0.1 ± 1.7 | −3.2, 3.4 | 0.98* | 3.3* |
| RTPD (%) | −0.08 ± 0.9 | −1.9, 1.8 | 0.99* | 1.8* |
| ITPD (%) | 0.19 ± 1.6 | −3.0, 3.4 | 0.85 | 3.2* |
| (B) Regional | ||||
| LAD VSSS% | 0.11 ± 1.66 | −3.14, 3.36 | 0.96 | 3.3 |
| LCX VSSS% | −0.02 ± 1.48 | −2.91, 2.87 | 0.78 | 2.9 |
| RCA VSSS% | −0.07 ± 2.10 | −4.18, 4.04 | 0.81 | 4.1 |
| LAD STPD (%) | 0.2 ± 1.8 | −3.4, 3.8 | 0.96 | 3.5 |
| LCX STPD (%) | 0.08 ± 1.59 | −3.04, 3.20 | 0.80 | 3.1 |
| RCA STPD (%) | −0.12 ± 0.53 | −1.16, 0.91 | 0.98## | 1.0## |
| LAD VSRS% | 0.18 ± 1.03 | −1.84, 2.20 | 0.98 | 2.0 |
| LCX VSRS% | 0.04 ± 0.73 | −1.39, 1.48 | 0.72 | 1.4# |
| RCA VSRS% | −0.32 ± 1.39 | −3.05, 2.40 | 0.91 | 2.7 |
| LAD RTPD (%) | 0.0 ± 1.2 | −2.4, 2.4 | 0.98 | 2.4 |
| LCX RTPD (%) | −0.01 ± 1.04 | −2.1, 2.0 | 0.71 | 2.0 |
| RCA RTPD (%) | −0.07 ± 0.47 | −0.98, 0.85 | 0.98## | 0.9## |
* P ≤ .002 vs corresponding value for normalized visual scores.
# P < .001 vs corresponding values for TPD.
## P < .0001 vs corresponding values for normalized visual scores.
Figure 1Differences of global perfusion defects. The left column shows the differences of visual perfusion results between test and retest; right column shows the differences of TPD results between test and retest. The first row shows the differences of normalized visual stress perfusion and STPD between test and retest; the second row shows the differences of normalized visual rest perfusion and RTPD between test and retest; the third row shows the differences of normalized visual ischemic perfusion and ITPD between test and retest
Figure 2RC of stress and ischemic perfusion measures for all studies and for abnormal studies Note: CAD includes all cases with prior myocardial infarction, percutaneous coronary intervention, or coronary artery bypass graft. Abnr defined by STPD ≥ 3% includes abnormal cases who have STPD ≥ 3%, Abnr defined by STPD ≥ 5% includes abnormal cases who have STPD ≥ 5%, VSSS% means normalized visual summed scores for stress, and VSDS% means normalized visual summed difference score
Global (A) and regional (B) reproducibility of functional parameters for each vascular territory
| Bias ± SD | 95% Limits of agreement | Correlation | RC | |
|---|---|---|---|---|
| (A) Global | ||||
| VM | 0.2 ± 5.5 | −10.4, 10.9 | 0.74 | 10.7 |
| VT | −0.2 ± 1.9 | −3.9, 3.5 | 0.94 | 3.7 |
| WM (mm) | −0.13 ± 0.63 | −1.38, 1.11 | 0.89 | 1.3 |
| WT (%) | −0.8 ± 4.3 | −9.2, 7.7 | 0.91 | 8.6 |
| Stress EF (%) | −1.36 ± 3.6 | −8.5, 5.8 | 0.94 | 7.1 |
| Rest EF (%) | −0.07 ± 5.2 | −10.2, 10.1 | 0.87 | 10.1 |
| Stress EF (%)* | −1.88 ± 3.6 | −8.9, 5.1 | 0.94 | 6.4 |
| Rest EF (%)* | −0.66 ± 4.4 | −9.2, 7.9 | 0.91 | 8.7 |
| (B) Regional | ||||
| LAD VM | 0.2 ± 3.3 | −6.3, 6.6 | 0.84 | 6.5 |
| LCX VM | 0.0 ± 0.8 | −1.6, 1.5 | 0.41 | 1.6 |
| RCA VM | 0.1 ± 2.4 | −4.5, 4.7 | 0.56 | 4.7 |
| LAD WM (mm) | −0.09 ± 0.75 | −1.55, 1.37 | 0.90 | 1.5 |
| LCX WM (mm) | 0.06 ± 0.94 | −2.04, 1.65 | 0.87 | 1.9 |
| RCA WM (mm) | −0.12 ± 0.86 | −1.8, 1.57 | 0.84 | 1.7 |
| LAD VT | 0.0 ± 0.9 | −1.6, 1.7 | 0.98 | 1.8 |
| LCX VT | 0.0 ± 0.7 | −1.3, 1.3 | 0.72 | 1.4 |
| RCA VT | −0.2 ± 1.3 | −2.8, 2.4 | 0.85 | 2.5 |
| LAD WT (%) | 1.0 ± 5.2 | −11.2, 9.2 | 0.90 | 10.4 |
| LCX WT (%) | −0.8 ± 5.2 | −10.9, 9.3 | 0.87 | 10.4 |
| RCA WT (%) | −0.8 ± 4.5 | −9.6, 8.1 | 0.89 | 9.0 |
* The results without any observer’s contour adjustment on either scan (N = 65).