Literature DB >> 20963534

How to treat a tibial post fracture in total knee arthroplasty? A systematic review.

Paul F Lachiewicz1.   

Abstract

BACKGROUND: Posterior-stabilized TKAs, which use a polyethylene tibial post to articulate against a metal femoral cam, are used regularly. Reported complications are related to the patellofemoral articulation or the tibial post-cam mechanism. Fracture of the tibial post is an uncommon but disabling complication after posterior-stabilized TKA that requires operative treatment. QUESTIONS/PURPOSES: The literature was reviewed to determine the frequency of tibial post fracture and address three questions: (1) Is there a specific prosthetic design or patient demographics in knees with a fracture of the tibial post? (2) What are the common presenting complaints and methods of diagnosis? (3) What methods of treatment have been used?
METHODS: A PubMed search of English language articles from February 1982 to April 2010 was performed and 20 articles, all Level IV studies, were identified.
RESULTS: One specific design of posterior-stabilized tibial post with a central screw hole had a 12.4% incidence of fracture. Tibial post fracture has been reported with other designs, but with an incidence of 1% or less. The most common presenting symptoms include effusion, instability, or patella clunk syndrome. The most common method of diagnosis was clinical examination followed by arthroscopic examination. Treatment with revision to a new tibial polyethylene liner generally has been successful at short-term followup.
CONCLUSIONS: Tibial post fracture is a relatively uncommon complication after posterior-stabilized TKA that usually is treated successfully with liner exchange. The low quality of available literature makes it difficult to recommend a specific treatment protocol.

Entities:  

Mesh:

Year:  2010        PMID: 20963534      PMCID: PMC3094630          DOI: 10.1007/s11999-010-1609-9

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.176


  33 in total

1.  Tibial post wear in posterior stabilized total knee arthroplasty. An unrecognized source of polyethylene debris.

Authors:  S K Puloski; R W McCalden; S J MacDonald; C H Rorabeck; R B Bourne
Journal:  J Bone Joint Surg Am       Date:  2001-03       Impact factor: 5.284

2.  Fracture of a second-generation highly cross-linked UHMWPE tibial post in a posterior-stabilized scorpio knee system.

Authors:  Kwang Am Jung; Su Chan Lee; Seung Hyun Hwang; Sung Min Kim
Journal:  Orthopedics       Date:  2008-11       Impact factor: 1.390

3.  Fifteen-year to 19-year follow-up of the Insall-Burstein-1 total knee arthroplasty.

Authors:  Ayesha R Abdeen; Stacy R Collen; Stacy B Collen; Kelly G Vince
Journal:  J Arthroplasty       Date:  2009-02-05       Impact factor: 4.757

4.  Tibial post failures in a condylar posterior cruciate substituting total knee arthroplasty.

Authors:  B Sonny Bal; David Greenberg; Stephen Li; David R Mauerhan; Loren Schultz; Kenneth Cherry
Journal:  J Arthroplasty       Date:  2007-11-09       Impact factor: 4.757

5.  Survivorship of cemented total knee arthroplasty.

Authors:  D E Font-Rodriguez; G R Scuderi; J N Insall
Journal:  Clin Orthop Relat Res       Date:  1997-12       Impact factor: 4.176

6.  Two stage fracture of a polyethylene post in a 9-year-old posterior-stabilized knee prosthesis: a case report.

Authors:  Fabio D'Angelo; Daniele Marcolli; Paolo Bulgheroni; Luigi Murena; Terenzio Congiu; Paolo Cherubino
Journal:  J Med Case Rep       Date:  2010-02-23

7.  Fracture of the polyethylene tibial post in a NexGen posterior-stabilized knee prosthesis.

Authors:  Yen-Shuo Chiu; Wei-Ming Chen; Ching-Kuei Huang; Chao-Ching Chiang; Tain-Hsiung Chen
Journal:  J Arthroplasty       Date:  2004-12       Impact factor: 4.757

8.  Fracture of the polyethylene tibial post in a posterior cruciate-substituting total knee arthroplasty mimicking patellar clunk syndrome: a report of 5 cases.

Authors:  David R Mauerhan
Journal:  J Arthroplasty       Date:  2003-10       Impact factor: 4.757

9.  Fifteen-year survival and osteolysis associated with a modular posterior stabilized knee replacement. A concise follow-up of a previous report.

Authors:  Paul F Lachiewicz; Elizabeth S Soileau
Journal:  J Bone Joint Surg Am       Date:  2009-06       Impact factor: 5.284

10.  Early nontraumatic fracture of the polyethylene tibial post in a NexGen LPS-Flex posterior stabilized knee prosthesis.

Authors:  Chi-Sheng Lee; Wei-Ming Chen; Hung-Chan Kou; Wai-Hee Lo; Chien-Lung Chen
Journal:  J Arthroplasty       Date:  2009-03-20       Impact factor: 4.757

View more
  6 in total

1.  The 2012 Mark Coventry award: a retrieval analysis of high flexion versus posterior-stabilized tibial inserts.

Authors:  Nicholas R Paterson; Matthew G Teeter; Steven J MacDonald; Richard W McCalden; Douglas D R Naudie
Journal:  Clin Orthop Relat Res       Date:  2013-01       Impact factor: 4.176

2.  Vitamin E-enriched polyethylene bearings are not inferior to Arcom bearings in primary total knee arthroplasty at medium-term follow-up.

Authors:  Samy Ftaita; Aurélie Vanden Berghe; Emmanuel Thienpont
Journal:  Arch Orthop Trauma Surg       Date:  2021-01-08       Impact factor: 3.067

3.  Fracture of the polyethylene tibial post in the posterior-stabilized total knee prosthesis: arthrographic and CT arthrographic diagnosis.

Authors:  Yung Hsu; Chien-Hung Lin; Ginger H F Shu; Tsyh-Jyi Hsieh; Clement Kuen-Huang Chen
Journal:  Skeletal Radiol       Date:  2019-02-07       Impact factor: 2.199

Review 4.  Implant survival of 3rd-condyle and post-cam posterior-stabilised total knee arthroplasty are comparable at follow-up > 10 years: a systematic review.

Authors:  David H Dejour; Jacobus H Müller; Mo Saffarini; Michel Timoteo; Pierre Chambat; Gerard Deschamps; Michel P Bonnin
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2021-03-04       Impact factor: 4.342

5.  Catastrophic Failures of the Tibial Post in a Bicruciate Substituting Total Knee Prosthesis.

Authors:  Dominic Marino; Zethariah Zelinski; Daniel Mesko; Jason Cochran
Journal:  Arthroplast Today       Date:  2021-04-15

6.  Mechanics of post-cam engagement during simulated dynamic activity.

Authors:  Clare K Fitzpatrick; Chadd W Clary; Adam J Cyr; Lorin P Maletsky; Paul J Rullkoetter
Journal:  J Orthop Res       Date:  2013-04-19       Impact factor: 3.494

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.