Literature DB >> 20178608

Two stage fracture of a polyethylene post in a 9-year-old posterior-stabilized knee prosthesis: a case report.

Fabio D'Angelo1, Daniele Marcolli, Paolo Bulgheroni, Luigi Murena, Terenzio Congiu, Paolo Cherubino.   

Abstract

INTRODUCTION: Several cases of tibial post breakage are reported in the literature. To the best of our knowledge, only three cases of NexGen knee prosthesis (Zimmer, Warsaw, Indiana, USA) tibial post failure have been reported. CASE
PRESENTATION: In November 1999, a 63-year-old Caucasian woman from Italy with a history of symptomatic left knee osteoarthritis underwent a total knee arthroplasty. In March 2008, while rising from a chair, she felt a sudden pain and instability in her left knee. She reported a fracture of the polyethylene post of the tibial insert. No malposition or malalignment of either the femoral or tibial components were identified. The polyethylene tibial insert was studied under light microscopy and scanning electron microscopy. The fracture was also noted to have occurred without any notable polyethylene wear.
CONCLUSION: Scanning electron microscopy revealed two different damage patterns that could be explained with a two-stage rupture of our patient's polyethylene post. This could have been caused by a non-optimal ligamentous balancing during first implant surgery. Her knee probably developed a varus instability that weakened the post, and then a posterior anterior stress finally broke the polyethylene.

Entities:  

Year:  2010        PMID: 20178608      PMCID: PMC2843710          DOI: 10.1186/1752-1947-4-65

Source DB:  PubMed          Journal:  J Med Case Rep        ISSN: 1752-1947


  18 in total

1.  Fracture of posterior-stablized tibial insert in a Genesis knee prosthesis.

Authors:  M Mariconda; G Lotti; C Milano
Journal:  J Arthroplasty       Date:  2000-06       Impact factor: 4.757

2.  Tibial post impingement in posterior-stabilized total knee arthroplasty.

Authors:  John J Callaghan; Michael R O'Rourke; Devon D Goetz; Thomas P Schmalzried; Patricia A Campbell; Richard C Johnston
Journal:  Clin Orthop Relat Res       Date:  2002-11       Impact factor: 4.176

3.  Recurrent dislocation of total knee arthroplasty: an unusual cause.

Authors:  T P Ng; K Y Chiu
Journal:  J Arthroplasty       Date:  2003-12       Impact factor: 4.757

4.  Early instability with mobile-bearing total knee arthroplasty: a series of 25 cases.

Authors:  Stephen Ridgeway; Joseph T Moskal
Journal:  J Arthroplasty       Date:  2004-09       Impact factor: 4.757

5.  Failure of a metal-reinforced tibial post in total knee arthroplasty.

Authors:  B Sonny Bal; David Greenberg
Journal:  J Arthroplasty       Date:  2007-04       Impact factor: 4.757

6.  Mechanisms of failure in total knee arthroplasty.

Authors:  J R Moreland
Journal:  Clin Orthop Relat Res       Date:  1988-01       Impact factor: 4.176

7.  Prodromes of failure in total knee arthroplasty.

Authors:  J H Lonner; J M Siliski; R D Scott
Journal:  J Arthroplasty       Date:  1999-06       Impact factor: 4.757

8.  Fracture of the polyethylene tibial post in a NexGen posterior-stabilized knee prosthesis.

Authors:  Yen-Shuo Chiu; Wei-Ming Chen; Ching-Kuei Huang; Chao-Ching Chiang; Tain-Hsiung Chen
Journal:  J Arthroplasty       Date:  2004-12       Impact factor: 4.757

9.  Fracture of the polyethylene tibial post in a posterior cruciate-substituting total knee arthroplasty mimicking patellar clunk syndrome: a report of 5 cases.

Authors:  David R Mauerhan
Journal:  J Arthroplasty       Date:  2003-10       Impact factor: 4.757

10.  Early nontraumatic fracture of the polyethylene tibial post in a NexGen LPS-Flex posterior stabilized knee prosthesis.

Authors:  Chi-Sheng Lee; Wei-Ming Chen; Hung-Chan Kou; Wai-Hee Lo; Chien-Lung Chen
Journal:  J Arthroplasty       Date:  2009-03-20       Impact factor: 4.757

View more
  4 in total

1.  The 2012 Mark Coventry award: a retrieval analysis of high flexion versus posterior-stabilized tibial inserts.

Authors:  Nicholas R Paterson; Matthew G Teeter; Steven J MacDonald; Richard W McCalden; Douglas D R Naudie
Journal:  Clin Orthop Relat Res       Date:  2013-01       Impact factor: 4.176

Review 2.  How to treat a tibial post fracture in total knee arthroplasty? A systematic review.

Authors:  Paul F Lachiewicz
Journal:  Clin Orthop Relat Res       Date:  2010-10-21       Impact factor: 4.176

3.  Fracture of the insert cone of a polyethylene liner in a failed posterior-stabilized, rotating-platform total knee arthroplasty.

Authors:  Marc R Angerame; Jason M Jennings; Douglas A Dennis
Journal:  Arthroplast Today       Date:  2017-11-07

4.  Nontraumatic Tibial Polyethylene Insert Cone Fracture in Rotating-Platform Total Knee Arthroplasty.

Authors:  Cillian J Keogh; David Keohane; James A Harty; David Mulcahy
Journal:  Arthroplast Today       Date:  2021-04-08
  4 in total