Literature DB >> 20948977

Adaptive CT for High-Resolution, Controlled-Dose, Region-of-Interest Imaging.

Jared W Moore1, Harrison H Barrett, Lars R Furenlid.   

Abstract

We have developed a flexible x-ray micro-CT system, named FaCT, capable of changing its geometric configuration and acquisition protocol in order to best suit an object being imaged for a particular diagnostic task. High-performance computing technologies have been a major enabling factor for this adaptive CT system in terms of system control, fast reconstruction, and data analysis. In this work, we demonstrate an adaptive procedure in which a quick, sparse-projection pre-scan is performed, the data are reconstructed, and a region of interest is identified. Next, a diagnostic-quality scan is performed where, given the region of interest, the control computer calculates an illumination window for on-line control of an x-ray source masking aperture to transmit radiation only through the region of interest throughout the scan trajectory. Finally, the diagnostic scan data are reconstructed, with the region of interest being clearly resolved. We use a combination of a multi-core CPU and a pair of NVIDIA Tesla GPUs to perform these tasks.

Entities:  

Year:  2009        PMID: 20948977      PMCID: PMC2952940          DOI: 10.1109/NSSMIC.2009.5402313

Source DB:  PubMed          Journal:  IEEE Nucl Sci Symp Conf Rec (1997)        ISSN: 1095-7863


  4 in total

1.  A Task-Based Approach to Adaptive and Multimodality Imaging: Computation techniques are proposed for figures-of-merit to establish feasibility and optimize use of multiple imaging systems for disease diagnosis and treatment-monitoring.

Authors:  Eric Clarkson; Matthew A Kupinski; Harrison H Barrett; Lars Furenlid
Journal:  Proc IEEE Inst Electr Electron Eng       Date:  2008-03       Impact factor: 10.961

2.  Adaptive SPECT.

Authors:  Harrison H Barrett; Lars R Furenlid; Melanie Freed; Jacob Y Hesterman; Matthew A Kupinski; Eric Clarkson; Meredith K Whitaker
Journal:  IEEE Trans Med Imaging       Date:  2008-06       Impact factor: 10.048

3.  A prototype instrument for single pinhole small animal adaptive SPECT imaging.

Authors:  Melanie Freed; Matthew A Kupinski; Lars R Furenlid; Donald W Wilson; Harrison H Barrett
Journal:  Med Phys       Date:  2008-05       Impact factor: 4.071

4.  Instrumentation Design for Adaptive SPECT/CT.

Authors:  Jared W Moore; Lars R Furenlid; Harrison H Barrett
Journal:  IEEE Nucl Sci Symp Conf Rec (1997)       Date:  2008-10
  4 in total
  2 in total

1.  Adaptive Angular Sampling for SPECT Imaging.

Authors:  Nan Li; Ling-Jian Meng
Journal:  IEEE Trans Nucl Sci       Date:  2011-10-03       Impact factor: 1.679

2.  SPECT system optimization against a discrete parameter space.

Authors:  L J Meng; N Li
Journal:  Phys Med Biol       Date:  2013-04-15       Impact factor: 3.609