Annick C Weustink1, Pim J de Feyter.
Abstract
Computed tomographic (CT) imaging involves exposure to ionizing radiation, with a consequent potential risk to the patient. Here we review the technical factors that affect the radiation dose in cardiac CT diagnostic procedures, the parameters that are used to measure and compare radiation doses, the magnitude of radiation exposure, and how to efficiently reduce it.Entities:
Year: 2009 PMID: 20948773 PMCID: PMC2920698 DOI: 10.3410/M1-1
Source DB: PubMed Journal: F1000 Med Rep ISSN: 1757-5931
Parameters of radiation dose
| Parameter | What it represents | SI unit |
|---|---|---|
| CT dose index (CTDIvol) | Mean absorbed radiation dose over total volume scanned | Gray (Gy) |
| Dose-length product (DLP) | Overall dose per scan | mGy x cm |
| Effective dose (E) | Biological effect | Sievert (Sv) |
Variables associated with radiation exposure
| Variable | Measures to minimize radiation |
|---|---|
| Number of CT scans | Alternative diagnostic modality |
| Avoid repeat scans | |
| Justification of CT scan (avoid young age) | |
| Tube current (mA) | Optimize settings according to patient BMI |
| Tube voltage (kV) | |
| Pitch | Adjust to patient's heart rate |
| Tube current modulation | |
| ECG-controlled | Adjust data acquisition to optimal cardiac phase (ECG pulsing window) |
| Angular-longitudinal | Adjust to patient attenuation profile |
| Scan length | Minimize |
| Scan mode | |
| Prospective gating axial (PGA) | Only X-ray switched on during optimal cardiac phase |
| Retrospective gating helical (RGH) | Optimize tube current modulation |
Estimates of effective dose (E)
| E range in mSv (mean) | ECG tube modulation | |
|---|---|---|
| Average background radiation (United States) | 1.0-10 (3.6) | - |
| Conventional coronary angiography | 3.1-21.8 (5.6) | - |
| Nuclear MPI studies | 2.2-31.5 | - |
| Ca scoring (retrospectively gated helical) | 1.0-6.2 | - |
| Ca scoring (prospectively gated axial) | 1.0-3.0 | - |
| CTCA (retrospectively gated helical) | ||
| 16-slice | 5.7-13 | No |
| 16-slice | 2.9-5.6 | Yes |
| 64-slice | 9.4-21.4 | Yes |
| 64-slice dual source | 7.8-16.9 | Yes |
| 320-slice | 10.1-18.1 | No |
| CTCA (prospectively gated axial) | ||
| 64-slice | 0.75-6.7 (2.8) | - |
| 320-slice | 4.9-16.5 | - |
Data from [9-19].
Lifetime-attributable risk of cancer following 64-slice CT coronary angiography
| Age (years) | Standard CT | Tube current modulation | Standard CT | Tube current modulation |
|---|---|---|---|---|
| Female | Female | Male | Male | |
| 40 | 1 in 284 | 1 in 435 | 1 in 1007 | 1 in 1534 |
| 60 | 1 in 466 | 1 in 715 | 1 in 1241 | 1 in 1911 |
| 80 | 1 in 1388 | 1 in 2082 | 1 in 1284 | 1 in 1834 |
Data from [7].