Literature DB >> 20930618

Thrombogenic potential of Innovia polymer valves versus Carpentier-Edwards Perimount Magna aortic bioprosthetic valves.

Thomas E Claiborne1, Gaurav Girdhar, Siobhain Gallocher-Lowe, Jawaad Sheriff, Yasushi P Kato, Leonard Pinchuk, Richard T Schoephoerster, Jolyon Jesty, Danny Bluestein.   

Abstract

Trileaflet polymeric prosthetic aortic valves (AVs) produce hemodynamic characteristics akin to the natural AV and may be most suitable for applications such as transcatheter implantation and mechanical circulatory support (MCS) devices. Their success has not yet been realized due to problems of calcification, durability, and thrombosis. We address the latter by comparing the platelet activation rates (PARs) of an improved polymer valve design (Innovia LLC) made from poly(styrene-block-isobutylene-block-styrene) (SIBS) with the commercially available Carpentier-Edwards Perimount Magna Aortic Bioprosthetic Valve. We used our modified prothrombinase platelet activity state (PAS) assay and flow cytometry methods to measure platelet activation of a pair of 19 mm valves mounted inside a pulsatile Berlin left ventricular assist device (LVAD). The PAR of the polymer valve measured with the PAS assay was fivefold lower than that of the tissue valve (p = 0.005) and fourfold lower with flow cytometry measurements (p = 0.007). In vitro hydrodynamic tests showed clinically similar performance of the Innovia and Magna valves. These results demonstrate a significant improvement in thrombogenic performance of the polymer valve compared with our previous study of the former valve design and encourage further development of SIBS for use in heart valve prostheses.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 20930618     DOI: 10.1097/MAT.0b013e3181fcbd86

Source DB:  PubMed          Journal:  ASAIO J        ISSN: 1058-2916            Impact factor:   2.872


  17 in total

1.  Physical Characterization and Platelet Interactions under Shear Flows of a Novel Thermoset Polyisobutylene-based Co-polymer.

Authors:  Jawaad Sheriff; Thomas E Claiborne; Phat L Tran; Roshni Kothadia; Sheela George; Yasushi P Kato; Leonard Pinchuk; Marvin J Slepian; Danny Bluestein
Journal:  ACS Appl Mater Interfaces       Date:  2015-09-23       Impact factor: 9.229

2.  Design of a pulsatile flow facility to evaluate thrombogenic potential of implantable cardiac devices.

Authors:  Sivakkumar Arjunon; Pablo Hidalgo Ardana; Neelakantan Saikrishnan; Shalv Madhani; Brent Foster; Ari Glezer; Ajit P Yoganathan
Journal:  J Biomech Eng       Date:  2015-02-11       Impact factor: 2.097

3.  Routine clinical anti-platelet agents have limited efficacy in modulating hypershear-mediated platelet activation associated with mechanical circulatory support.

Authors:  Lorenzo Valerio; Jawaad Sheriff; Phat L Tran; William Brengle; Alberto Redaelli; Gianfranco B Fiore; Federico Pappalardo; Danny Bluestein; Marvin J Slepian
Journal:  Thromb Res       Date:  2017-12-05       Impact factor: 3.944

Review 4.  Principles of TAVR valve design, modelling, and testing.

Authors:  Oren M Rotman; Matteo Bianchi; Ram P Ghosh; Brandon Kovarovic; Danny Bluestein
Journal:  Expert Rev Med Devices       Date:  2018-10-29       Impact factor: 3.166

5.  Comparative efficacy of in vitro and in vivo metabolized aspirin in the DeBakey ventricular assist device.

Authors:  Jawaad Sheriff; Gaurav Girdhar; Wei-Che Chiu; Jolyon Jesty; Marvin J Slepian; Danny Bluestein
Journal:  J Thromb Thrombolysis       Date:  2014-05       Impact factor: 2.300

Review 6.  Polymeric trileaflet prosthetic heart valves: evolution and path to clinical reality.

Authors:  Thomas E Claiborne; Marvin J Slepian; Syed Hossainy; Danny Bluestein
Journal:  Expert Rev Med Devices       Date:  2012-11       Impact factor: 3.166

7.  In vitro evaluation of a novel hemodynamically optimized trileaflet polymeric prosthetic heart valve.

Authors:  Thomas E Claiborne; Jawaad Sheriff; Maximilian Kuetting; Ulrich Steinseifer; Marvin J Slepian; Danny Bluestein
Journal:  J Biomech Eng       Date:  2013-02       Impact factor: 2.097

8.  Hemodynamic and thrombogenic analysis of a trileaflet polymeric valve using a fluid-structure interaction approach.

Authors:  Filippo Piatti; Francesco Sturla; Gil Marom; Jawaad Sheriff; Thomas E Claiborne; Marvin J Slepian; Alberto Redaelli; Danny Bluestein
Journal:  J Biomech       Date:  2015-08-21       Impact factor: 2.712

9.  A novel mathematical model of activation and sensitization of platelets subjected to dynamic stress histories.

Authors:  João S Soares; Jawaad Sheriff; Danny Bluestein
Journal:  Biomech Model Mechanobiol       Date:  2013-01-29

10.  Device thrombogenicity emulation: a novel methodology for optimizing the thromboresistance of cardiovascular devices.

Authors:  Danny Bluestein; Gaurav Girdhar; Shmuel Einav; Marvin J Slepian
Journal:  J Biomech       Date:  2012-12-06       Impact factor: 2.712

View more