Prachi Pandit1, Yi Qi, Kevin F King, G Allan Johnson.
Abstract
A simple technique is implemented for correction of artifacts arising from nonuniform T(2) -weighting of k-space data in fast spin echo-based PROPELLER (periodically rotated overlapping parallel lines with enhanced reconstruction). An additional blade with no phase-encoding gradients is acquired to generate the scaling factor used for the correction. Results from simulations and phantom experiments, as well as in vivo experiments in free-breathing mice, demonstrate the advantages of the proposed method. This technique is developed specifically for high-field imaging applications where T(2) decay is rapid.Entities:
Mesh:
Year: 2010 PMID: 20928875 PMCID: PMC3026877 DOI: 10.1002/mrm.22624
Source DB: PubMed Journal: Magn Reson Med ISSN: 0740-3194 Impact factor: 4.668