Evaluation of central corneal thickness measurement with RTVue spectral domain optical coherence tomography in normal subjects.

Harsha Laxmana Rao1, Addepalli Uday Kumar, Anjul Kumar, Swathi Chary, Sirisha Senthil, Pravin Krishna Vaddavalli, Chandra Sekhar Garudadri.   

Abstract

PURPOSE: To determine (1) repeatability of central corneal thickness (CCT) measurements by spectral domain optical coherence tomography (RTVue; Optovue, Inc, Fremont, CA) and (2) agreement between CCT measurements by RTVue and those by ultrasonic pachymetry, Orbscan, and anterior segment optical coherence tomography (ASOCT).
METHODS: In a prospective observational study, 2 cohorts of normal subjects were recruited. In the first cohort (51 subjects), 5 measurements of CCT were obtained by RTVue during the same visit to determine the repeatability. In the second cohort (65 subjects), CCT measurements were obtained by RTVue, ultrasonic pachymetry, Orbscan, and ASOCT during the same visit to determine the agreement among these instruments. Repeatability was assessed by intraclass correlation coefficient (ICC), within-subject standard deviation, coefficient of repeatability, and within-subject coefficient of variation. Agreement was assessed by ICC and Bland and Altman plots.
RESULTS: Repeatability of CCT measurements by RTVue as assessed by ICC, within-subject standard deviation, coefficient of repeatability, and within-subject coefficient of variation was 0.99 (0.99-0.99), 2.2 (1.9-2.5), 4.2 μm (3.6-4.8), and 0.4% (0.3-0.5), respectively. The average CCT by RTVue (529 μm) was comparable to that by ultrasonic pachymetry (539 μm; P = 0.15), Orbscan (536 μm; P = 0.54), and ASOCT (526 μm; P = 0.77). The 95% limits of agreement on Bland and Altman plots ranged from 20 μm (between RTVue and ASOCT) to 33 μm (between RTVue and Orbscan).
CONCLUSIONS: CCT measurements by RTVue have an excellent repeatability. Although CCT measurements by RTVue are comparable to that by ultrasonic pachymetry, Orbscan, and ASOCT, the difference between instruments can be significant depending on the clinical situation considered.

Entities:  

Mesh:

Year:  2011        PMID: 20885314     DOI: 10.1097/ICO.0b013e3181e16c65

Source DB:  PubMed          Journal:  Cornea        ISSN: 0277-3740            Impact factor:   2.651


  12 in total

1.  Comparison of central corneal thickness measurements with three new optical devices and a standard ultrasonic pachymeter.

Authors:  Hasan Ali Bayhan; Seray Aslan Bayhan; Izzet Can
Journal:  Int J Ophthalmol       Date:  2014-04-18       Impact factor: 1.779

2.  Repeatability assessment of anterior segment biometric measurements under accommodative and nonaccommodative conditions using an anterior segment OCT.

Authors:  Noelia Martínez-Albert; Jose J Esteve-Taboada; Robert Montés-Micó
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2017-11-02       Impact factor: 3.117

3.  An investigation of the ocular toxic effects of levetiracetam therapy in children with epilepsy.

Authors:  Betul Diler Durgut; Adem Turk; Elif Acar Arslan; Tulay Kamasak; Sevim Sahin; Beril Dilber; Tugce Turkcan Soguksulu; Ali Cansu
Journal:  Childs Nerv Syst       Date:  2019-02-20       Impact factor: 1.475

4.  Comparison of artemis 2 ultrasound and Visante optical coherence tomography corneal thickness profiles.

Authors:  Roxana Ursea; Matthew Feng; Raksha Urs; Arindam RoyChoudhury; Ronald H Silverman
Journal:  J Refract Surg       Date:  2012-11-30       Impact factor: 3.573

Review 5.  The genetic and environmental factors for keratoconus.

Authors:  Ariela Gordon-Shaag; Michel Millodot; Einat Shneor; Yutao Liu
Journal:  Biomed Res Int       Date:  2015-05-17       Impact factor: 3.411

6.  Reproducibility and repeatability of central corneal thickness measurement in healthy eyes using four different optical devices.

Authors:  Remzi Karadag; Murat Unluzeybek; Ozgur Cakici; Ayse Yagmur Kanra; Huseyin Bayramlar
Journal:  Int Ophthalmol       Date:  2016-10-08       Impact factor: 2.031

7.  Quantitative OCT-based corneal topography in keratoconus with intracorneal ring segments.

Authors:  Sergio Ortiz; Pablo Pérez-Merino; Nicolas Alejandre; E Gambra; I Jimenez-Alfaro; Susana Marcos
Journal:  Biomed Opt Express       Date:  2012-04-02       Impact factor: 3.732

8.  Comparison of Measurement of Central Corneal Thickness with Spectral Domain Optical Coherence Tomography and Standard Ultrasonic Pachymeter in Premature Infants.

Authors:  Emre Hekimoglu; Muhammet Kazım Erol; Devrim Toslak; Deniz Turgut Coban; Berna Doğan; Ozgur Yucel
Journal:  J Ophthalmol       Date:  2015-07-22       Impact factor: 1.909

9.  Intraoperative pachymetry using spectral-domain optical coherence tomography during accelerated corneal collagen crosslinking.

Authors:  Vanissa W S Chow; Sayantan Biswas; Marco Yu; Victoria W Y Wong; Vishal Jhanji
Journal:  Biomed Res Int       Date:  2013-07-25       Impact factor: 3.411

10.  Repeatability and agreement of five imaging systems for measuring anterior segment parameters in healthy eyes.

Authors:  Mukesh Kumar; Rohit Shetty; Chaitra Jayadev; Harsha L Rao; Debarun Dutta
Journal:  Indian J Ophthalmol       Date:  2017-04       Impact factor: 1.848

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.