Assessment of central corneal thickness using OCT, ultrasound, optical low coherence reflectometry and Scheimpflug pachymetry.

Sven C Beutelspacher1, Nermin Serbecic, Alexander F Scheuerle.   

Abstract

PURPOSE: Accurate measurement of central corneal thickness (CCT) is essential in refractive surgery and advanced glaucoma diagnostics. The gold standard for pachymetry is full-contact ultrasound-based pachymetry. As this method is associated with potential sources of error, noncontact methods have been introduced. The aim of this study was to compare CCT results measured using 4 different techniques.
METHODS: In this analysis of 20 patients (40 eyes) at the University Eye Hospital Heidelberg, Germany, we compared a slit-lamp-mounted optical coherence tomography (OCT) system (SL-OCT, Heidelberg Engineering, Heidelberg, Germany), conventional ultrasound pachymetry (IOPac, Heidelberg Engineering), optical low coherence reflectometry (OLCR, Haag-Streit, Germany), and scanning-slit pachymetry (Orbscan).
RESULTS: Comparison among the 4 groups did not show significant differences, except the comparison of OLCR to Orbscan; the mean was significantly different (p=0.0247) and the Orbscan detected slightly thicker values than the other methods.
CONCLUSIONS: Orbscan, SL-OCT, and OLCR provide non-touch technology, without the need for local anesthesia, and limiting the risk of infection or artifacts. Extreme care must be used interpreting the results obtained from Orbscan, as this technique may overestimate the CCT significantly.

Entities:  

Mesh:

Year:  2011        PMID: 20872361     DOI: 10.5301/ejo.2010.1093

Source DB:  PubMed          Journal:  Eur J Ophthalmol        ISSN: 1120-6721            Impact factor:   2.597


  9 in total

1.  Lamina cribrosa thickness is not correlated with central corneal thickness or axial length in healthy eyes: central corneal thickness, axial length, and lamina cribrosa thickness.

Authors:  Eun Ji Lee; Tae-Woo Kim; Robert N Weinreb; Min Hee Suh; Hyunjoong Kim
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2012-09-20       Impact factor: 3.117

2.  Comparison of central corneal thickness measurements with three new optical devices and a standard ultrasonic pachymeter.

Authors:  Hasan Ali Bayhan; Seray Aslan Bayhan; Izzet Can
Journal:  Int J Ophthalmol       Date:  2014-04-18       Impact factor: 1.779

3.  Monitoring of central corneal thickness after phacoemulsification-comparison of statical and rotating Scheimpflug pachymetry, and spectral-domain OCT.

Authors:  Daniel M Handzel; Carsten H Meyer; Alfred Wegener
Journal:  Int J Ophthalmol       Date:  2022-08-18       Impact factor: 1.645

4.  Comparison of central corneal thickness: ultrasound pachymetry versus slit-lamp optical coherence tomography, specular microscopy, and Orbscan.

Authors:  Wassia A Khaja; Sandeep Grover; Amy T Kelmenson; Lee R Ferguson; Kumar Sambhav; Kakarla V Chalam
Journal:  Clin Ophthalmol       Date:  2015-06-12

5.  Comparison of central corneal thickness with four noncontact devices: An agreement analysis of swept-source technology.

Authors:  Erhan Ozyol; Pelin Özyol
Journal:  Indian J Ophthalmol       Date:  2017-06       Impact factor: 1.848

6.  Accuracy of Cirrus HD-OCT and Topcon SP-3000P for measuring central corneal thickness.

Authors:  Jorge A Calvo-Sanz; Javier Ruiz-Alcocer; Miguel A Sánchez-Tena
Journal:  J Optom       Date:  2017-02-21

7.  Comparison of central corneal thickness measurements using different imaging devices and ultrasound pachymetry.

Authors:  Ertugrul Can; Hilal Eser-Ozturk; Mustafa Duran; Tugba Cetinkaya; Nursen Arıturk
Journal:  Indian J Ophthalmol       Date:  2019-04       Impact factor: 1.848

8.  Comparison of central corneal thickness measurements with three different optical devices.

Authors:  Mehmet Barış Üçer; Erdinç Bozkurt
Journal:  Ther Adv Ophthalmol       Date:  2021-02-23

9.  Repeatability and reproducibility of post-mortem central corneal thickness measurements using a portable optical coherence tomography system in humans: a prospective multicenter study.

Authors:  Pietro Emanuele Napoli; Matteo Nioi; Letizia Gabiati; Michela Laurenzo; Fabio De-Giorgio; Vincenzo Scorcia; Simone Grassi; Ernesto d'Aloja; Maurizio Fossarello
Journal:  Sci Rep       Date:  2020-09-02       Impact factor: 4.379