Literature DB >> 20860044

Effect of cup inclination angle during microseparation and rim loading on the wear of BIOLOX® delta ceramic-on-ceramic total hip replacement.

Mazen Al-Hajjar1, Ian J Leslie, Joanne Tipper, Sophie Williams, John Fisher, Louise M Jennings.   

Abstract

Ceramic-on-ceramic (CoC) bearings in total hip replacements (THRs) have shown low wear volumes under standard gait in hip simulator studies. However, clinical reports have indicated variations in wear rates and formation of stripe-like wear area on the ceramic femoral heads. The aim of this study was to investigate the influence of cup inclination angle and microseparation on the wear of CoC bearings in THRs. The six station Leeds II Physiological Anatomical Joint Simulator was used to investigate the wear of 28 mm diameter alumina matrix composite ceramic bearings (BIOLOX® delta). It was shown that increasing the cup inclination angle from 55° to 65° had no significant effect on the wear rate of BIOLOX® delta CoC under both standard gait and microseparation conditions in this in vitro study. Under standard gait conditions, the mean wear rate for both cup inclination angle conditions was very low at 0.05 mm(3)/million cycles. The introduction of microseparation to the standard gait cycle increased the mean wear rates to 0.13 mm(3)/million cycles for the cup inclination angle of 55° and 0.11 mm(3)/million cycles for that of 65°. The level of increased wear with microseparation was not dependent on cup angle. A stripe of wear on the head also formed, with corresponding superior rim wear on the cup. The wear rates obtained were low compared to the HIPed third generation alumina ceramic (BIOLOX® forte) tested under the same adverse conditions (1.84 mm(3)/million cycles). BIOLOX® delta has shown lower wear than previous ceramic materials used in THR under adverse conditions.
© 2010 Wiley Periodicals, Inc.

Entities:  

Mesh:

Year:  2010        PMID: 20860044     DOI: 10.1002/jbm.b.31708

Source DB:  PubMed          Journal:  J Biomed Mater Res B Appl Biomater        ISSN: 1552-4973            Impact factor:   3.368


  15 in total

1.  Biotribological study of large diameter ceramic-on-CFR-PEEK hip joint including fluid uptake, wear and frictional heating.

Authors:  Qian Qian Wang; Jun Jie Wu; Anthony Unsworth; Adam Briscoe; Marcus Jarman-Smith; Conor Lowry; David Simpson; Simon Collins
Journal:  J Mater Sci Mater Med       Date:  2012-03-28       Impact factor: 3.896

2.  Edge-loading severity as a function of cup lip radius in metal-on-metal total hips--a finite element analysis.

Authors:  Jacob M Elkins; Karen M Kruger; Douglas R Pedersen; John J Callaghan; Thomas D Brown
Journal:  J Orthop Res       Date:  2011-08-02       Impact factor: 3.494

3.  Squeaking and microcracks in a delta-delta ceramic coupling: pin-on-disc study.

Authors:  Kiyokazu Fukui; Ayumi Kaneuji; Tadami Matsumoto; Kazuhiro Shintani
Journal:  Eur J Orthop Surg Traumatol       Date:  2016-03-12

4.  Third generation delta ceramic-on-ceramic bearing for total hip arthroplasty at mid-term follow-up.

Authors:  Alessandro Bistolfi; Riccardo Ferracini; Alessandro Aprato; Alessandro Massè; Walter Daghino; Sara Lea; Stefano Artiaco; Gwo-Chin Lee
Journal:  J Orthop       Date:  2020-08-28

5.  A case-driven hypothesis for multi-stage crack growth mechanism in fourth-generation ceramic head fracture.

Authors:  Stefano Lucchini; Massimiliano Baleani; Federico Giardina; Andrea Martelli; Francesco Castagnini; Barbara Bordini; Francesco Traina
Journal:  J Orthop Surg Res       Date:  2022-06-03       Impact factor: 2.677

6.  (v) Simulation and measurement of wear in metal-on-metal bearings in vitro- understanding the reasons for increased wear.

Authors:  John Fisher; Mazen Al Hajjar; Sophie Williams; Joanne Tipper; Eileen Ingham; Louise Jennings
Journal:  Orthop Trauma       Date:  2012-08

7.  Effect of femoral head size on the wear of metal on metal bearings in total hip replacements under adverse edge-loading conditions.

Authors:  Mazen Al-Hajjar; John Fisher; Sophie Williams; Joanne L Tipper; Louise M Jennings
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2012-12-20       Impact factor: 3.368

8.  (iv) Enhancing the safety and reliability of joint replacement implants.

Authors:  Louise M Jennings; Mazen Al-Hajjar; Claire L Brockett; Sophie Williams; Joanne L Tipper; Eileen Ingham; John Fisher
Journal:  Orthop Trauma       Date:  2012-08

9.  Contact mechanics of modular metal-on-polyethylene total hip replacement under adverse edge loading conditions.

Authors:  Xijin Hua; Junyan Li; Ling Wang; Zhongmin Jin; Ruth Wilcox; John Fisher
Journal:  J Biomech       Date:  2014-09-01       Impact factor: 2.712

10.  Wear of composite ceramics in mixed-material combinations in total hip replacement under adverse edge loading conditions.

Authors:  Mazen Al-Hajjar; Silvia Carbone; Louise M Jennings; Sabine Begand; Thomas Oberbach; Daniel Delfosse; John Fisher
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2016-04-08       Impact factor: 3.368