M Tomasevic1, V Vukcevic, G Stankovic, D Orlic, M Ostojic.
Abstract
Coronary artery bifurcations are one of the largest challenges in interventional cardiology. Presented is the case of a patient in whom restenosis of a drug-eluting stent (DES) occurred as a consequence of guide wire re-crossing between the main vessel stent struts and the vessel wall in the proximal part of DES, and consequential balloon crushing of the proximal portion of the DES. Initially, the complication was not recognized because of a good angiographic result and absence of intravascular ultrasound (IVUS) guidance during the procedure. During the second procedure, IVUS analysis explained the mechanism of the DES failure. The problem was solved with the implantation of a new DES.Entities:
Mesh:
Year: 2010 PMID: 20857077 DOI: 10.1007/s00059-010-3387-2