Literature DB >> 20851567

Crack initiation in retrieved cross-linked highly cross-linked ultrahigh-molecular-weight polyethylene acetabular liners: an investigation of 9 cases.

Jevan Furmanski1, Matthew J Kraay, Clare M Rimnac.   

Abstract

Nine cross-linked highly cross-linked ultrahigh-molecular weight polyethylene acetabular liners were retrieved at revision surgery. Eight of the liners were fully intact and functional at retrieval. Six cases contained shallow initiated cracks at the root of rim notches; 1 crack had propagated several millimeters. Optical and electron microscopic inspection of the crack surfaces revealed clam shell markings, which are characteristic of fatigue crack initiation. Crack initiation at notches has been identified in reports of catastrophic cross-linked liner failures, with crack initiation sites exhibiting similar morphology and clam shell markings. Thus, we believe that the shallow cracks identified in this case series are precursors to catastrophic rim fracture. The results of this study recommend further investigations to clarify the etiology and prevalence of crack initiation in cross-linked acetabular liners.
Copyright © 2011 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20851567      PMCID: PMC3053431          DOI: 10.1016/j.arth.2010.07.016

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  12 in total

1.  A novel method of cross-linking ultra-high-molecular-weight polyethylene to improve wear, reduce oxidation, and retain mechanical properties. Recipient of the 1999 HAP Paul Award.

Authors:  O K Muratoglu; C R Bragdon; D O O'Connor; M Jasty; W H Harris
Journal:  J Arthroplasty       Date:  2001-02       Impact factor: 4.757

2.  Recurrent dislocation after revision total hip replacement with a large prosthetic femoral head. A case report.

Authors:  David Halley; Andrew Glassman; Roy D Crowninshield
Journal:  J Bone Joint Surg Am       Date:  2004-04       Impact factor: 5.284

3.  Fatigue crack propagation resistance of virgin and highly crosslinked, thermally treated ultra-high molecular weight polyethylene.

Authors:  Sara J Gencur; Clare M Rimnac; Steven M Kurtz
Journal:  Biomaterials       Date:  2005-11-21       Impact factor: 12.479

4.  Influence of femoral head size on impingement, dislocation and stress distribution in total hip replacement.

Authors:  Daniel Kluess; Heiner Martin; Wolfram Mittelmeier; Klaus-Peter Schmitz; Rainer Bader
Journal:  Med Eng Phys       Date:  2006-08-09       Impact factor: 2.242

5.  Rim cracking of the cross-linked longevity polyethylene acetabular liner after total hip arthroplasty.

Authors:  Stephen S Tower; John H Currier; Barbara H Currier; Kimberly A Lyford; Douglas W Van Citters; Michael B Mayor
Journal:  J Bone Joint Surg Am       Date:  2007-10       Impact factor: 5.284

6.  How prevalent are implant wear and osteolysis, and how has the scope of osteolysis changed since 2000?

Authors:  Amanda Marshall; Michael D Ries; Wayne Paprosky
Journal:  J Am Acad Orthop Surg       Date:  2008       Impact factor: 3.020

7.  Anteversion of the acetabular cup. Measurement of angle after total hip replacement.

Authors:  M K Ackland; W B Bourne; H K Uhthoff
Journal:  J Bone Joint Surg Br       Date:  1986-05

8.  Dislocations after total hip-replacement arthroplasties.

Authors:  G E Lewinnek; J L Lewis; R Tarr; C L Compere; J R Zimmerman
Journal:  J Bone Joint Surg Am       Date:  1978-03       Impact factor: 5.284

9.  Deformation, yielding, fracture and fatigue behavior of conventional and highly cross-linked ultra high molecular weight polyethylene.

Authors:  Lisa A Pruitt
Journal:  Biomaterials       Date:  2005-03       Impact factor: 12.479

10.  Clinical fracture of cross-linked UHMWPE acetabular liners.

Authors:  Jevan Furmanski; Martin Anderson; Sonny Bal; A Seth Greenwald; David Halley; Brad Penenberg; Michael Ries; Lisa Pruitt
Journal:  Biomaterials       Date:  2009-07-29       Impact factor: 12.479

View more
  16 in total

1.  Crack propagation resistance is similar under static and cyclic loading in crosslinked UHMWPE: a pilot study.

Authors:  Jevan Furmanski; Clare M Rimnac
Journal:  Clin Orthop Relat Res       Date:  2011-08       Impact factor: 4.176

2.  Early failure of a non-cemented femoral stem after minimal-invasive total hip arthroplasty: cause analysis and classification.

Authors:  Jan P Benthien; Wolf Siepen
Journal:  Musculoskelet Surg       Date:  2012-05-06

3.  Equivalent wear performance of dual mobility bearing compared with standard bearing in total hip arthroplasty: in vitro study.

Authors:  Gaël Gaudin; André Ferreira; Romain Gaillard; Jean Louis Prudhon; Jacques H Caton; Sébastien Lustig
Journal:  Int Orthop       Date:  2016-11-22       Impact factor: 3.075

Review 4.  Safe clinical use of carbon nanotubes as innovative biomaterials.

Authors:  Naoto Saito; Hisao Haniu; Yuki Usui; Kaoru Aoki; Kazuo Hara; Seiji Takanashi; Masayuki Shimizu; Nobuyo Narita; Masanori Okamoto; Shinsuke Kobayashi; Hiroki Nomura; Hiroyuki Kato; Naoyuki Nishimura; Seiichi Taruta; Morinobu Endo
Journal:  Chem Rev       Date:  2014-04-10       Impact factor: 60.622

5.  Highly crosslinked polyethylene: a safe alternative to conventional polyethylene for dual mobility cup mobile component. A biomechanical validation.

Authors:  Matthieu Malatray; Jean-Paul Roux; Stanislas Gunst; Vincent Pibarot; Julien Wegrzyn
Journal:  Int Orthop       Date:  2016-11-11       Impact factor: 3.075

6.  Backside Wear Is Not Dependent on the Acetabular Socket Design in Crosslinked Polyethylene Liners.

Authors:  Kamal Bali; Richard W McCalden; Douglas D R Naudie; Steven J MacDonald; Matthew G Teeter
Journal:  Clin Orthop Relat Res       Date:  2016-02       Impact factor: 4.176

7.  Gait instability may indicate liner failure in patients with total hip arthroplasty. A report of three cases.

Authors:  F Díaz-Dilernia; M Lattore; G Zanotti; F Comba; F Piccaluga; M Buttaro
Journal:  Ann R Coll Surg Engl       Date:  2021-08-20       Impact factor: 1.951

8.  Natural polyphenols enhance stability of crosslinked UHMWPE for joint implants.

Authors:  Jie Shen; Guorong Gao; Xincai Liu; Jun Fu
Journal:  Clin Orthop Relat Res       Date:  2015-03       Impact factor: 4.176

9.  Application of viscoelastic fracture model and non-uniform crack initiation at clinically relevant notches in crosslinked UHMWPE.

Authors:  P Abhiram Sirimamilla; Jevan Furmanski; Clare M Rimnac
Journal:  J Mech Behav Biomed Mater       Date:  2012-08-01

10.  Viscoplastic crack initiation and propagation in crosslinked UHMWPE from clinically relevant notches up to 0.5mm radius.

Authors:  P Abhiram Sirimamilla; Clare M Rimnac; Jevan Furmanski
Journal:  J Mech Behav Biomed Mater       Date:  2017-09-01
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.