Literature DB >> 20831091

Designing passive MRI-safe implantable conducting leads with electrodes.

Paul A Bottomley1, Ananda Kumar, William A Edelstein, Justin M Allen, Parag V Karmarkar.   

Abstract

PURPOSE: The presence of implanted electronic devices with conducting leads and electrodes are contraindicated for magnetic resonance imaging (MRI), denying many patients its potential benefits. The prime concern is MRI's radio frequency (RF) fields, which can cause elevated local specific absorption rates (SARs) and potential heat injury. The purpose of this article is to develop and compare a range of passive implantable "MRI-safe" lead designs.
METHODS: Conducting leads incorporating different lengths (3-75 cm), insulation thicknesses (0-105 microm), resistances (100-3000 omega), coiled conductors (inner diameter < or = 1.2 mm), high-impedance (135-2700 omega) RF traps, and single-coiled and triple-coiled coaxial-wound "billabong" leads with reversed coil sections that oppose and reduce the induced current, are investigated both experimentally using local temperature measurements, and by numerical full-wave electromagnetic field analysis of the local SAR, in three different-sized bioanalogous model saline-gel phantoms at 1.5 T MRI and 4 W/kg exposure.
RESULTS: In all designs, the maximum computed 1 g average SAR and experimental temperature rise occur at the bare electrodes. Electrode heating increases with lead insulation thickness and peaks for uncoiled leads 25-50 cm long. A reasonable match between computed SAR and the point SAR estimated from thermal sensors obtained by approximating the computation volume to that of the thermal probes. Factors that maximize the impedance of leads with resistive, coiled, RF trap and billabong elements can effectively limit heating below 1-2 degrees, but folded lead configurations can be a concern. The RF trap and billabong designs can both support multiple conductors and electrodes, with billabong prototype leads also heating <1 degrees C when tested for 3 T MRI.
CONCLUSIONS: Lead insulation and length strongly affect implanted lead safety to RF exposure during MRI. Lead designs employing impedance and reversed winding sections offer hope for the development of passive, MRI-safe, implantable conducting leads for future human use.

Entities:  

Mesh:

Year:  2010        PMID: 20831091     DOI: 10.1118/1.3439590

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  27 in total

1.  Interventional loopless antenna at 7 T.

Authors:  Mehmet Arcan Ertürk; Abdel-Monem M El-Sharkawy; Paul A Bottomley
Journal:  Magn Reson Med       Date:  2011-12-12       Impact factor: 4.668

2.  MRI-guided vascular access with an active visualization needle.

Authors:  Christina E Saikus; Kanishka Ratnayaka; Israel M Barbash; Jessica H Colyer; Ozgur Kocaturk; Anthony Z Faranesh; Robert J Lederman
Journal:  J Magn Reson Imaging       Date:  2011-11       Impact factor: 4.813

3.  It's the little things: On the complexity of planar electrode heating in MRI.

Authors:  Johannes B Erhardt; Thomas Lottner; Jessica Martinez; Ali C Özen; Martin Schuettler; Thomas Stieglitz; Daniel B Ennis; Michael Bock
Journal:  Neuroimage       Date:  2019-03-29       Impact factor: 6.556

4.  A multichannel, real-time MRI RF power monitor for independent SAR determination.

Authors:  Abdel-Monem M El-Sharkawy; Di Qian; Paul A Bottomley; William A Edelstein
Journal:  Med Phys       Date:  2012-05       Impact factor: 4.071

5.  Controlling radiofrequency-induced currents in guidewires using parallel transmit.

Authors:  Maryam Etezadi-Amoli; Pascal Stang; Adam Kerr; John Pauly; Greig Scott
Journal:  Magn Reson Med       Date:  2014-12-17       Impact factor: 4.668

6.  Monitoring local heating around an interventional MRI antenna with RF radiometry.

Authors:  M Arcan Ertürk; AbdEl-Monem M El-Sharkawy; Paul A Bottomley
Journal:  Med Phys       Date:  2015-03       Impact factor: 4.071

Review 7.  A Review of Numerical Simulation and Analytical Modeling for Medical Devices Safety in MRI.

Authors:  J Kabil; L Belguerras; S Trattnig; C Pasquier; J Felblinger; A Missoffe
Journal:  Yearb Med Inform       Date:  2016-11-10

8.  Using a low-amplitude RF pulse at echo time (LARFET) for device localization in MRI.

Authors:  Murat Tümer; Baykal Sarioglu; Senol Mutlu; Yekta Ulgen; Arda Yalcinkaya; Cengizhan Ozturk
Journal:  Med Biol Eng Comput       Date:  2014-08-31       Impact factor: 2.602

9.  A Multi-Channel, High Dynamic Range, Real Time RF Power Deposition Monitor.

Authors:  A M El-Sharkawy; D Qian; P A Bottomley; W A Edelstein
Journal:  Proc Int Soc Magn Reson Med Sci Meet Exhib Int Soc Magn Reson Med Sci Meet Exhib       Date:  2011

10.  An RF dosimeter for independent SAR measurement in MRI scanners.

Authors:  Di Qian; Abdel-Monem M El-Sharkawy; Paul A Bottomley; William A Edelstein
Journal:  Med Phys       Date:  2013-12       Impact factor: 4.071

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.