Noise variance analysis using a flat panel x-ray detector: a method for additive noise assessment with application to breast CT applications.

Kai Yang1, Shih-Ying Huang, Nathan J Packard, John M Boone.   

Abstract

PURPOSE: A simplified linear model approach was proposed to accurately model the response of a flat panel detector used for breast CT (bCT).
METHODS: Individual detector pixel mean and variance were measured from bCT projection images acquired both in air and with a polyethylene cylinder, with the detector operating in both fixed low gain and dynamic gain mode. Once the coefficients of the linear model are determined, the fractional additive noise can be used as a quantitative metric to evaluate the system's efficiency in utilizing x-ray photons, including the performance of different gain modes of the detector.
RESULTS: Fractional additive noise increases as the object thickness increases or as the radiation dose to the detector decreases. For bCT scan techniques on the UC Davis prototype scanner (80 kVp, 500 views total, 30 frames/s), in the low gain mode, additive noise contributes 21% of the total pixel noise variance for a 10 cm object and 44% for a 17 cm object. With the dynamic gain mode, additive noise only represents approximately 2.6% of the total pixel noise variance for a 10 cm object and 7.3% for a 17 cm object.
CONCLUSIONS: The existence of the signal-independent additive noise is the primary cause for a quadratic relationship between bCT noise variance and the inverse of radiation dose at the detector. With the knowledge of the additive noise contribution to experimentally acquired images, system modifications can be made to reduce the impact of additive noise and improve the quantum noise efficiency of the bCT system.

Entities:  

Mesh:

Year:  2010        PMID: 20831059      PMCID: PMC2902539          DOI: 10.1118/1.3447720

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  7 in total

1.  On the noise variance of a digital mammography system.

Authors:  Arthur Burgess
Journal:  Med Phys       Date:  2004-07       Impact factor: 4.071

2.  A time-delay integration charge-coupled device camera for slot-scanned digital radiography.

Authors:  D W Holdsworth; R K Gerson; A Fenster
Journal:  Med Phys       Date:  1990 Sep-Oct       Impact factor: 4.071

3.  Determination of the conversion gain and the accuracy of its measurement for detector elements and arrays.

Authors:  B P Beecken; E R Fossum
Journal:  Appl Opt       Date:  1996-07-01       Impact factor: 1.980

4.  Properties of preprocessed sinogram data in x-ray computed tomography.

Authors:  Bruce R Whiting; Parinaz Massoumzadeh; Orville A Earl; Joseph A O'Sullivan; Donald L Snyder; Jeffrey F Williamson
Journal:  Med Phys       Date:  2006-09       Impact factor: 4.071

5.  A practical exposure-equivalent metric for instrumentation noise in x-ray imaging systems.

Authors:  G K Yadava; A T Kuhls-Gilcrist; S Rudin; V K Patel; K R Hoffmann; D R Bednarek
Journal:  Phys Med Biol       Date:  2008-08-22       Impact factor: 3.609

6.  Noise power properties of a cone-beam CT system for breast cancer detection.

Authors:  Kai Yang; Alexander L C Kwan; Shih-Ying Huang; Nathan J Packard; John M Boone
Journal:  Med Phys       Date:  2008-12       Impact factor: 4.071

7.  Empirical and theoretical investigation of the noise performance of indirect detection, active matrix flat-panel imagers (AMFPIs) for diagnostic radiology.

Authors:  J H Siewerdsen; L E Antonuk; Y el-Mohri; J Yorkston; W Huang; J M Boudry; I A Cunningham
Journal:  Med Phys       Date:  1997-01       Impact factor: 4.071

  7 in total
  10 in total

1.  Cascaded systems analysis of noise and detectability in dual-energy cone-beam CT.

Authors:  Grace J Gang; Wojciech Zbijewski; J Webster Stayman; Jeffrey H Siewerdsen
Journal:  Med Phys       Date:  2012-08       Impact factor: 4.071

2.  Optical mammography: Diffuse optical imaging of breast cancer.

Authors:  Kijoon Lee
Journal:  World J Clin Oncol       Date:  2011-01-10

3.  Cupping artifact correction and automated classification for high-resolution dedicated breast CT images.

Authors:  Xiaofeng Yang; Shengyong Wu; Ioannis Sechopoulos; Baowei Fei
Journal:  Med Phys       Date:  2012-10       Impact factor: 4.071

4.  Accelerated barrier optimization compressed sensing (ABOCS) reconstruction for cone-beam CT: phantom studies.

Authors:  Tianye Niu; Lei Zhu
Journal:  Med Phys       Date:  2012-07       Impact factor: 4.071

5.  A photon recycling approach to the denoising of ultra-low dose X-ray sequences.

Authors:  Sai Gokul Hariharan; Norbert Strobel; Christian Kaethner; Markus Kowarschik; Stefanie Demirci; Shadi Albarqouni; Rebecca Fahrig; Nassir Navab
Journal:  Int J Comput Assist Radiol Surg       Date:  2018-04-10       Impact factor: 2.924

6.  Experimentally determined spectral optimization for dedicated breast computed tomography.

Authors:  Nicolas D Prionas; Shih-Ying Huang; John M Boone
Journal:  Med Phys       Date:  2011-02       Impact factor: 4.071

7.  A semiempirical linear model of indirect, flat-panel x-ray detectors.

Authors:  Shih-Ying Huang; Kai Yang; Craig K Abbey; John M Boone
Journal:  Med Phys       Date:  2012-04       Impact factor: 4.071

8.  Cone-beam breast CT using an offset detector: effect of detector offset and image reconstruction algorithm.

Authors:  Hsin Wu Tseng; Andrew Karellas; Srinivasan Vedantham
Journal:  Phys Med Biol       Date:  2022-04-07       Impact factor: 4.174

9.  Dedicated cone-beam breast CT using laterally-shifted detector geometry: Quantitative analysis of feasibility for clinical translation.

Authors:  Srinivasan Vedantham; Hsin-Wu Tseng; Souleymane Konate; Linxi Shi; Andrew Karellas
Journal:  J Xray Sci Technol       Date:  2020       Impact factor: 1.535

10.  Radiation dosimetry of a clinical prototype dedicated cone-beam breast CT system with offset detector.

Authors:  Hsin Wu Tseng; Andrew Karellas; Srinivasan Vedantham
Journal:  Med Phys       Date:  2021-01-26       Impact factor: 4.506

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.