Automatic tube current modulation technique for multidetector CT: is it effective with a 64-detector CT?

Yoshinori Funama1, Kazuo Awai, Masahiro Hatemura, Masamitchi Shimamura, Yumi Yanaga, Seitaro Oda, Yasuyuki Yamashita.   

Abstract

To investigate whether it is possible to obtain adequate images at uniform image noise levels and reduced radiation exposure with our automatic tube current modulation (ATCM) technique for 64-detector CT. The study population consisted of 64 patients with known or suspected lung or abdominal disease. We used a 64-detector CT scanner (LightSpeed VCT, GE Healthcare, Waukesha, WI, USA) and a combined angular and longitudinal tube current modulation technique (Smart mA, GE Healthcare, Waukesha, WI, USA) to examine 34 patients. The scanning parameters were identical; the minimum and maximum tube current thresholds were 50 and 800 mA, respectively. For study of the constant tube current technique, 30 additional patients were examined at 350 mA. The CT number and image noise (SD of the CT number) were measured in the 64 patients at six levels, i.e., the center of the left ventricle, the liver dome, the porta hepatis, the center of the spleen and the right and left renal pelvis. When we used the ATCM technique, the mean image noise ranged from 8.40 at the center of the left ventricle to 11.31 at the porta hepatis; the mean tube current ranged from 105.9 mAs at the center of the left ventricle to 169.6 mAs at the center of the spleen. The mean dose reduction rate per constant tube current at 175 mAs ranged from 3.1 to 39.5%. By use of the ATCM technique, it is possible to maintain a constant image noise level with a 64-detector CT.

Entities:  

Mesh:

Year:  2007        PMID: 20821161     DOI: 10.1007/s12194-007-0005-y

Source DB:  PubMed          Journal:  Radiol Phys Technol        ISSN: 1865-0333


  19 in total

1.  Determination of height, weight, body mass index, and body surface area with a single abdominal CT image.

Authors:  Estella M Geraghty; John M Boone
Journal:  Radiology       Date:  2003-07-24       Impact factor: 11.105

2.  Radiation dose and image quality in pediatric CT: effect of technical factors and phantom size and shape.

Authors:  Marilyn J Siegel; Bernhard Schmidt; David Bradley; Christoph Suess; Charles Hildebolt
Journal:  Radiology       Date:  2004-09-09       Impact factor: 11.105

3.  Use of an automatic exposure control mechanism for dose optimization in multi-detector row CT examinations: clinical evaluation.

Authors:  Tom H Mulkens; Patrick Bellinck; Michel Baeyaert; Dirk Ghysen; Xavier Van Dijck; Elvier Mussen; Caroline Venstermans; Jean-Luc Termote
Journal:  Radiology       Date:  2005-08-26       Impact factor: 11.105

4.  Metallic prosthesis: technique to avoid increase in CT radiation dose with automatic tube current modulation in a phantom and patients.

Authors:  Tejas Dalal; Mannudeep K Kalra; Stefania M R Rizzo; Bernhard Schmidt; Christoph Suess; Thomas Flohr; Michael A Blake; Sanjay Saini
Journal:  Radiology       Date:  2005-08       Impact factor: 11.105

5.  Multislice CT angiography in aortic stent grafting: relationship between image noise and body mass index.

Authors:  Zhonghua Sun
Journal:  Eur J Radiol       Date:  2006-11-07       Impact factor: 3.528

6.  Dose reduction in CT by anatomically adapted tube current modulation. II. Phantom measurements.

Authors:  W A Kalender; H Wolf; C Suess
Journal:  Med Phys       Date:  1999-11       Impact factor: 4.071

7.  Application of European Commission reference dose levels in CT examinations in Crete, Greece.

Authors:  V Tsapaki; S Kottou; D Papadimitriou
Journal:  Br J Radiol       Date:  2001-09       Impact factor: 3.039

8.  Comparison of patient doses in 256-slice CT and 16-slice CT scanners.

Authors:  S Mori; M Endo; K Nishizawa; K Murase; H Fujiwara; S Tanada
Journal:  Br J Radiol       Date:  2006-01       Impact factor: 3.039

9.  Artifacts in CT: recognition and avoidance.

Authors:  Julia F Barrett; Nicholas Keat
Journal:  Radiographics       Date:  2004 Nov-Dec       Impact factor: 5.333

Review 10.  Techniques and applications of automatic tube current modulation for CT.

Authors:  Mannudeep K Kalra; Michael M Maher; Thomas L Toth; Bernhard Schmidt; Bryan L Westerman; Hugh T Morgan; Sanjay Saini
Journal:  Radiology       Date:  2004-10-21       Impact factor: 11.105

View more
  2 in total

1.  A new low-dose multi-phase trauma CT protocol and its impact on diagnostic assessment and radiation dose in multi-trauma patients.

Authors:  Zlatan Alagic; Andreas Eriksson; Erika Drageryd; Sara Rezaei Motamed; Marius C Wick
Journal:  Emerg Radiol       Date:  2017-04-05

2.  Effects of the Use of Automatic Tube Current Modulation on Patient Dose and Image Quality in Computed Tomography

Authors:  Ayşegül Yurt; İsmail Özsoykal; Funda Obuz
Journal:  Mol Imaging Radionucl Ther       Date:  2019-09-06
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.