Literature DB >> 20809682

Lower rate of initial failures and reduced occurrence of adverse events with a new catheter model for continuous subcutaneous insulin infusion: prospective, two-period, observational, multicenter study.

Eric Renard1, Bruno Guerci, Anne-Marie Leguerrier, Robert Boizel.   

Abstract

BACKGROUND: No recent clinical data on the incidence of catheter-related adverse events under insulin pump therapy have been reported.
METHODS: This was a prospective, two-period, observational, multicenter study in 45 diabetes outpatients (mean continuous subcutaneous insulin infusion [CSII] use, 6 years; mean hemoglobin A1c, 7.7%, at baseline). During the initial 1-month period (P1), the patients used their current catheter model, including a soft cannula in 98% of cases. They moved then to the new Accu-Chek FlexLink catheter model (Disetronic Medical Systems AG, Burgdorf, Switzerland) for a 3-month period. The primary end point, including insertion failures and unexplained hyperglycemia within the first 6h after catheter placement, was assessed from logbook records during P1 and the last month of the second period (P2). Secondary end points were catheter replacements for unexplained hyperglycemia and/or events at risk for immediate insulin delivery failure after the first 6h.
RESULTS: Forty-five initial infusion failures occurred in 14 patients among 507 catheter insertions (8.9% of cases) during P1, whereas 15 similar events were seen in nine patients during P2 among 488 catheter insertions (3.1% of cases) (P<0.001). Catheters were replaced for later infusion troubles in 8% of cases during both P1 and P2. The overall rate of late cumulative events was, however, 113 of 507 (P1) versus 66 of 488 (P2) (P<0.001). The occurrence of pain, skin reaction, or redness at the infusion site was lower during P2.
CONCLUSION: Incidences both of initial failures and of premature catheter replacements were 8–9% with current CSII catheters. Significantly reduced failures after insertion and adverse events at the infusion site were observed with the new catheter model.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20809682     DOI: 10.1089/dia.2010.0073

Source DB:  PubMed          Journal:  Diabetes Technol Ther        ISSN: 1520-9156            Impact factor:   6.118


  19 in total

Review 1.  Clinical requirements for closed-loop control systems.

Authors:  William L Clarke; Eric Renard
Journal:  J Diabetes Sci Technol       Date:  2012-03-01

2.  Randomized trial of infusion set function: steel versus teflon.

Authors:  Parul J Patel; Kari Benasi; Gina Ferrari; Mark G Evans; Satya Shanmugham; Darrell M Wilson; Bruce A Buckingham
Journal:  Diabetes Technol Ther       Date:  2013-10-03       Impact factor: 6.118

3.  A New Animal Model of Insulin-Glucose Dynamics in the Intraperitoneal Space Enhances Closed-Loop Control Performance.

Authors:  Ankush Chakrabarty; Justin M Gregory; L Merkle Moore; Philip E Williams; Ben Farmer; Alan D Cherrington; Peter Lord; Brian Shelton; Don Cohen; Howard C Zisser; Francis J Doyle; Eyal Dassau
Journal:  J Process Control       Date:  2019-02-23       Impact factor: 3.666

4.  Concept and Implementation of a Novel Patch Pump for Insulin Delivery.

Authors:  Sina Ulbrich; Delia Waldenmaier; Cornelia Haug; Guido Freckmann; Til Rendschmidt; Thomas Künsting
Journal:  J Diabetes Sci Technol       Date:  2019-08-27

5.  Analysis of "Randomized Cross-Over Study Comparing Two Infusion Sets for CSII in Daily Life".

Authors:  Eric Renard
Journal:  J Diabetes Sci Technol       Date:  2016-10-23

6.  Multicenter closed-loop insulin delivery study points to challenges for keeping blood glucose in a safe range by a control algorithm in adults and adolescents with type 1 diabetes from various sites.

Authors:  Howard Zisser; Eric Renard; Boris Kovatchev; Claudio Cobelli; Angelo Avogaro; Revital Nimri; Lalo Magni; Bruce A Buckingham; H Peter Chase; Francis J Doyle; John Lum; Peter Calhoun; Craig Kollman; Eyal Dassau; Anne Farret; Jerome Place; Marc Breton; Stacey M Anderson; Chiara Dalla Man; Simone Del Favero; Daniela Bruttomesso; Alessio Filippi; Rachele Scotton; Moshe Phillip; Eran Atlas; Ido Muller; Shahar Miller; Chiara Toffanin; Davide Martino Raimondo; Giuseppe De Nicolao; Roy W Beck
Journal:  Diabetes Technol Ther       Date:  2014-07-08       Impact factor: 6.118

7.  We Need More Research and Better Designs for Insulin Infusion Sets.

Authors:  Lutz Heinemann; John Walsh; Ruth Roberts
Journal:  J Diabetes Sci Technol       Date:  2014-02-27

8.  DiAs user interface: a patient-centric interface for mobile artificial pancreas systems.

Authors:  Patrick Keith-Hynes; Stephanie Guerlain; Benton Mize; Colleen Hughes-Karvetski; Momin Khan; Molly McElwee-Malloy; Boris P Kovatchev
Journal:  J Diabetes Sci Technol       Date:  2013-11-01

9.  Insulin Pump and CGM Usage in the United States and Germany: Results of a Real-World Survey With 985 Subjects.

Authors:  John Walsh; Ruth Roberts; Dietmar Weber; Gabriele Faber-Heinemann; Lutz Heinemann
Journal:  J Diabetes Sci Technol       Date:  2015-06-12

10.  Randomized Cross-Over Study Comparing Two Infusion Sets for CSII in Daily Life.

Authors:  Guido Freckmann; Stephan Arndt; Albrecht Fießelmann; Gerhard Klausmann; Kristina Pralle; Thomas Künsting; Bettina Petersen
Journal:  J Diabetes Sci Technol       Date:  2016-09-25
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.