Randomized trial of infusion set function: steel versus teflon.

Parul J Patel1, Kari Benasi, Gina Ferrari, Mark G Evans, Satya Shanmugham, Darrell M Wilson, Bruce A Buckingham.   

Abstract

BACKGROUND: This study compared infusion set function for up to 1 week using either a Teflon(®) (Dupont(™), Wilmington, DE) catheter or a steel catheter for insulin pump therapy in type 1 diabetes mellitus. SUBJECTS AND METHODS: Twenty subjects participating in a randomized, open-labeled, crossover study were asked to wear two Quick-Set(®) and two Sure-T(®) infusion sets (both from Medtronic Minimed, Northridge, CA) until the infusion set failed or was worn for 1 week. All subjects wore a MiniMed continuous glucose monitoring system for the duration of the study.
RESULTS: One subject withdrew from the study. There were 38 weeks of Sure-T wear and 39 weeks of Quick-Set wear with no difference in the survival curves of the infusion sets. There was, however, a 15% initial failure rate with the Teflon infusion set. After 7 days, both types of infusion sets had a 64% failure rate. Overall, 30% failed because of hyperglycemia and a failed correction dose, 13% were removed for pain, 10% were pulled out by accident, 10% had erythema and/or induration of>10 mm, 5% fell out because of loss of adhesion, and 4% were removed for infection. The main predictor of length of wear was the individual subject. There was no increase in hyperglycemia or daily insulin requirements when an infusion set was successfully used for 7 days (n=25 of 77 weeks).
CONCLUSIONS: We found no difference between steel and Teflon infusion sets in their function over 7 days, although 15% of Teflon sets failed because of kinking on insertion. The strongest predictor of prolonged 7-day infusion set function was the individual subject, not the type of infusion set.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 24090124      PMCID: PMC3887416          DOI: 10.1089/dia.2013.0119

Source DB:  PubMed          Journal:  Diabetes Technol Ther        ISSN: 1520-9156            Impact factor:   6.118


  17 in total

1.  Metabolic decompensation in pump users due to lispro insulin precipitation.

Authors:  Howard A Wolpert; Raquel N Faradji; Susan Bonner-Weir; Myra A Lipes
Journal:  BMJ       Date:  2002-05-25

2.  Patient management of long-term continuous subcutaneous insulin infusion.

Authors:  Unn-Britt Johansson; Ulf Adamson; Per-Eric Lins; Regina Wredling
Journal:  J Adv Nurs       Date:  2005-07       Impact factor: 3.187

3.  Glycaemic control with continuous subcutaneous insulin infusion compared with intensive insulin injections in patients with type 1 diabetes: meta-analysis of randomised controlled trials.

Authors:  John Pickup; Martin Mattock; Sally Kerry
Journal:  BMJ       Date:  2002-03-23

4.  Durability of insulin pump use in pediatric patients with type 1 diabetes.

Authors:  Jamie R Wood; Elaine C Moreland; Lisa K Volkening; Britta M Svoren; Deborah A Butler; Lori M B Laffel
Journal:  Diabetes Care       Date:  2006-11       Impact factor: 19.112

5.  Long-term metabolic control with insulin pump therapy. Report of experience with 127 patients.

Authors:  R S Mecklenburg; E A Benson; J W Benson; B A Blumenstein; P N Fredlund; T S Guinn; R J Metz; R L Nielsen
Journal:  N Engl J Med       Date:  1985-08-22       Impact factor: 91.245

6.  Dermatological side effects and complications of continuous subcutaneous insulin infusion in preschool-age and school-age children.

Authors:  Edith Schober; Birgit Rami
Journal:  Pediatr Diabetes       Date:  2008-11-03       Impact factor: 4.866

7.  Acute complications associated with insulin infusion pump therapy. Report of experience with 161 patients.

Authors:  R S Mecklenburg; E A Benson; J W Benson; P N Fredlund; T Guinn; R J Metz; R L Nielsen; C A Sanner
Journal:  JAMA       Date:  1984-12-21       Impact factor: 56.272

8.  Effect of steel and teflon infusion catheters on subcutaneous adipose tissue blood flow and infusion counter pressure in humans.

Authors:  Lise Højbjerre; Camilla Skov-Jensen; Peter Kaastrup; Per Elgård Pedersen; Bente Stallknecht
Journal:  Diabetes Technol Ther       Date:  2009-05       Impact factor: 6.118

9.  Dermatological complications of continuous subcutaneous insulin infusion in children and adolescents.

Authors:  Louise S Conwell; Elena Pope; Ana M Artiles; Arun Mohanta; Alan Daneman; Denis Daneman
Journal:  J Pediatr       Date:  2008-01-22       Impact factor: 4.406

Review 10.  Severe hypoglycaemia and glycaemic control in Type 1 diabetes: meta-analysis of multiple daily insulin injections compared with continuous subcutaneous insulin infusion.

Authors:  J C Pickup; A J Sutton
Journal:  Diabet Med       Date:  2008-07       Impact factor: 4.359

View more
  29 in total

1.  Evaluation of Intradermal and Subcutaneous Infusion Set Performance Under 24-Hour Basal and Bolus Conditions.

Authors:  Elaine McVey; Steven Keith; Joshua K Herr; Diane Sutter; Ronald J Pettis
Journal:  J Diabetes Sci Technol       Date:  2015-08-27

Review 2.  Moving Toward a Unified Platform for Insulin Delivery and Sensing of Inputs Relevant to an Artificial Pancreas.

Authors:  Anneke Graf; Sybil A McAuley; Catriona Sims; Johanna Ulloa; Alicia J Jenkins; Gayane Voskanyan; David N O'Neal
Journal:  J Diabetes Sci Technol       Date:  2016-12-13

3.  Tissue Response to Subcutaneous Infusion Catheter.

Authors:  Ershuai Zhang; Zhiqiang Cao
Journal:  J Diabetes Sci Technol       Date:  2019-03-31

4.  Insulin Infusion Sets and Continuous Glucose Monitoring Sensors: Where the Artificial Pancreas Meets the Patient.

Authors:  Gregory P Forlenza
Journal:  Diabetes Technol Ther       Date:  2017-04       Impact factor: 6.118

5.  Analysis of "Randomized Cross-Over Study Comparing Two Infusion Sets for CSII in Daily Life".

Authors:  Eric Renard
Journal:  J Diabetes Sci Technol       Date:  2016-10-23

6.  Application of Zone Model Predictive Control Artificial Pancreas During Extended Use of Infusion Set and Sensor: A Randomized Crossover-Controlled Home-Use Trial.

Authors:  Gregory P Forlenza; Sunil Deshpande; Trang T Ly; Daniel P Howsmon; Faye Cameron; Nihat Baysal; Eric Mauritzen; Tatiana Marcal; Lindsey Towers; B Wayne Bequette; Lauren M Huyett; Jordan E Pinsker; Ravi Gondhalekar; Francis J Doyle; David M Maahs; Bruce A Buckingham; Eyal Dassau
Journal:  Diabetes Care       Date:  2017-06-05       Impact factor: 19.112

7.  We Need More Research and Better Designs for Insulin Infusion Sets.

Authors:  Lutz Heinemann; John Walsh; Ruth Roberts
Journal:  J Diabetes Sci Technol       Date:  2014-02-27

8.  Skin Oil Staining to Avoid Infusion Set Cannula Crimping: A Personal Observation.

Authors:  Antonio Carlo Bossi
Journal:  J Diabetes Sci Technol       Date:  2016-08-22

9.  Real-Time Detection of Infusion Site Failures in a Closed-Loop Artificial Pancreas.

Authors:  Daniel P Howsmon; Nihat Baysal; Bruce A Buckingham; Gregory P Forlenza; Trang T Ly; David M Maahs; Tatiana Marcal; Lindsey Towers; Eric Mauritzen; Sunil Deshpande; Lauren M Huyett; Jordan E Pinsker; Ravi Gondhalekar; Francis J Doyle; Eyal Dassau; Juergen Hahn; B Wayne Bequette
Journal:  J Diabetes Sci Technol       Date:  2018-02-01

10.  Randomized Cross-Over Study Comparing Two Infusion Sets for CSII in Daily Life.

Authors:  Guido Freckmann; Stephan Arndt; Albrecht Fießelmann; Gerhard Klausmann; Kristina Pralle; Thomas Künsting; Bettina Petersen
Journal:  J Diabetes Sci Technol       Date:  2016-09-25
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.