Effectiveness of four electronic apex locators to determine distance from the apical foramen.

R Stoll1, B Urban-Klein, M J Roggendorf, A Jablonski-Momeni, K Strauch, R Frankenberger.   

Abstract

AIM: To evaluate the accuracy of four electronic apex locators (EAL) in the apical region (0-3 mm short of the foramen) and to compare the precision of the readings on the display with the real position of the file in the root canal.
METHODOLOGY: Twenty single-rooted extracted teeth with round root canals were used. The canal orifices were preflared, and the length to the major foramen was determined visually under a microscope. Canals were enlarged, so that a size 15 file fitted well inside the canal. Teeth were mounted in acrylic test tubes filled with physiologic saline solution. Electronic length was determined in the region between the major foramen and 3 mm short of it in 0.5 mm steps with the Dentaport ZX, Root ZX mini, Elements Diagnostic Unit and Apex Locator and Raypex 5 using files of size 10 and size 15. The data were analysed using linear regression between true length and EAL reading, Bland-Altman plots and nonparametric tests at a significance level of alpha = 0.05.
RESULTS: The major foramen was detected by all EALs. With a measurement file positioned directly at the major foramen, meter readings were equivalent to a position 0.01-0.38 mm away. For the Dentaport ZX, a better accuracy using the size 15 file for the area 0-1.5 mm short of the apex was found. The differences in measurements between the two files were smaller for the other EALs. In linear regression, a good linearity for Dentaport ZX and Root ZX mini and moderate linearity for Elements Diagnostic Unit and Apex Locator and Raypex 5 were found. The slope of the measurement curve was too low (0.37-0.57) for the Raypex 5 and almost optimal for the Dentaport ZX (1.01-1.05). The Root ZX mini and the Elements Obturation Unit produced lower slope values and especially the Elements Obturation Unit revealed much higher SDs at the different measurement levels.
CONCLUSION: Amongst the four EALs, the Dentaport ZX and Root ZX mini had the best agreement between true lengths and meter readings. For the Raypex 5, an interpretation of the colour-coded zones as distance to the foramen cannot be recommended and might lead to erroneous interpretations.

Entities:  

Mesh:

Year:  2010        PMID: 20796049     DOI: 10.1111/j.1365-2591.2010.01765.x

Source DB:  PubMed          Journal:  Int Endod J        ISSN: 0143-2885            Impact factor:   5.264


  12 in total

1.  The use of cone beam computed tomography to predetermine root canal lengths in molar teeth: a comparison between two-dimensional and three-dimensional measurements.

Authors:  J P Tchorz; P J Poxleitner; S Stampf; S B M Patzelt; D Rottke; E Hellwig; M J Altenburger
Journal:  Clin Oral Investig       Date:  2013-07-30       Impact factor: 3.573

2.  Accuracy of five electronic foramen locators with different operating systems: an ex vivo study.

Authors:  Bruno Carvalho de Vasconcelos; Michelli de Medeiros Bueno; Suyane Maria Luna-Cruz; Marco Antonio Hungaro Duarte; Carlos Augusto de Oliveira Fernandes
Journal:  J Appl Oral Sci       Date:  2013 Mar-Apr       Impact factor: 2.698

3.  Effect of pre-flaring and file size on the accuracy of two electronic apex locators.

Authors:  Manoel Brito-Júnior; Carla Cristina Camilo; Gil Moreira-Júnior; Jesus Djalma Pecora; Manoel Damião Sousa-Neto
Journal:  J Appl Oral Sci       Date:  2012 Sep-Oct       Impact factor: 2.698

4.  Efficiency of 2 electronic apex locators on working length determination: A clinical study.

Authors:  Sibel Koçak; Mustafa Murat Koçak; Baran Can Sağlam
Journal:  J Conserv Dent       Date:  2013-05

5.  Calcium hydroxide dressing residues after different removal techniques affect the accuracy of Root-ZX apex locator.

Authors:  Emel Uzunoglu; Ayhan Eymirli; Mehmet Özgür Uyanik; Semra Çalt; Emre Nagas
Journal:  Restor Dent Endod       Date:  2014-11-05

6.  In vitro comparison of working length determination using three different electronic apex locators.

Authors:  Alper Kuştarci; Dilara Arslan; Demet Altunbaş
Journal:  Dent Res J (Isfahan)       Date:  2014-09

7.  Influence of root canal curvature on the accuracy of an electronic apex locator: An in vitro study.

Authors:  Lekha Santhosh; Pooja Raiththa; Srirekha Aswathanarayana; Srinivas Panchajanya; Jayakumar Thimmaraya Reddy; Shwetha Rajanna Susheela
Journal:  J Conserv Dent       Date:  2014-11

8.  Radicular anatomy of permanent mandibular second molars in an Iranian population: A preliminary study.

Authors:  Nahid M Akhlaghi; Fatemeh Mashadi Abbas; Mostafa Mohammadi; Mohammad Reza Karami Shamloo; Orkideh Radmehr; Ramin Kaviani; Vahid Rakhshan
Journal:  Dent Res J (Isfahan)       Date:  2016 Jul-Aug

9.  Evaluation of Conventional Radiography and an Electronic Apex Locator in Determining the Working Length in C-shaped Canals.

Authors:  Hamid Jafarzadeh; Masoud Beyrami; Maryam Forghani
Journal:  Iran Endod J       Date:  2017

10.  Comparison Between the Accuracy of NovApex Apex Locator and Radiographs in Determining Radiographic Apex.

Authors:  Shahriar Shahi; Saeed Rahimi; Amin Salem Milani; Mohammad Asghari Jafarabadi; Gholam Reza Djoorabbaf Naghsh
Journal:  Iran Endod J       Date:  2011-05-15