Agreement between directly measured blood pressure and pressures obtained with three veterinary-specific oscillometric units in cats.

Mark J Acierno1, Diana Seaton, Mark A Mitchell, Anderson da Cunha.   

Abstract

OBJECTIVE: To determine whether veterinary-specific oscillometric blood pressure units yield measurements that are in good agreement with directly measured blood pressures in cats.
DESIGN: Evaluation study. ANIMALS: 21 cats undergoing routine spaying or neutering. PROCEDURES: A 24-gauge catheter was inserted in a dorsal pedal artery, and systolic, diastolic, and mean arterial pressures were directly measured with a validated pressure measurement system. Values were compared with indirect blood pressure measurements obtained with 3 veterinary-specific oscillometric blood pressure units.
RESULTS: There was poor agreement between indirectly and directly measured blood pressures. For unit 1, bias between indirectly and directly measured values was -14.9 mm Hg (95% limits of agreement [LOA], -52.2 to 22.4 mm Hg), 4.4 mm Hg (95% LOA, -26.0 to 34.8 mm Hg), and -1.3 mm Hg (95% LOA, -26.7 to 24.1 mm Hg) for systolic, diastolic, and mean arterial pressures, respectively. For unit 2, bias was -10.3 mm Hg (95% LOA, -52.9 to 32.2 mm Hg), 13.0 mm Hg (95% LOA, -32.1 to 58.0 mm Hg), and 9.1 mm Hg (95% LOA, -32.9 to 51.2 mm Hg) for systolic, diastolic, and mean arterial pressures, respectively. For unit 3, bias was -13.4 mm Hg (95% LOA, -51.8 to 25.1 mm Hg), 8.0 mm Hg (95% LOA, -25.5 to 41.6 mm Hg), and -3.6 mm Hg (95% LOA, -31.6 to 24.5 mm Hg) for systolic, diastolic, and mean arterial pressures, respectively. CONCLUSIONS AND CLINICAL RELEVANCE: Results suggested that none of the 3 veterinary-specific oscillometric blood pressure units could be recommended for indirect measurement of blood pressure in cats.

Entities:  

Mesh:

Year:  2010        PMID: 20707750     DOI: 10.2460/javma.237.4.402

Source DB:  PubMed          Journal:  J Am Vet Med Assoc        ISSN: 0003-1488            Impact factor:   1.936


  6 in total

1.  Comparison of invasive and oscillometric blood pressure measurement techniques in anesthetized camelids.

Authors:  Turi K Aarnes; John A E Hubbell; Phillip Lerche; Richard M Bednarski
Journal:  Can Vet J       Date:  2012-08       Impact factor: 1.008

2.  Comparison of Doppler ultrasonic and oscillometric devices (with or without proprietary optimisations) for non-invasive blood pressure measurement in conscious cats.

Authors:  Petra Cerna; Panos E Archontakis; Hester Ok Cheuk; Danièlle A Gunn-Moore
Journal:  J Feline Med Surg       Date:  2020-06-25       Impact factor: 2.015

3.  Agreement of High-Definition Oscillometry (HDO) and Invasive Blood Pressure Measurements at a Metatarsal Artery in Isoflurane-Anaesthetised Horses.

Authors:  Lara Twele; Stephan Neudeck; Julien Delarocque; Nicole Verhaar; Julia Reiners; Mike Noll; Julia Tünsmeyer; Sabine B R Kästner
Journal:  Animals (Basel)       Date:  2022-02-02       Impact factor: 2.752

4.  'The Mercury Challenge': feline systolic blood pressure in primary care practice - a European survey.

Authors:  Andrew Sparkes; Catherine Garelli-Paar; Thomas Blondel; Emilie Guillot
Journal:  J Feline Med Surg       Date:  2022-06-27       Impact factor: 1.971

5.  Indirect arterial blood pressure measurement in healthy anesthetized cats using a device that combines oscillometry with photoplethysmography.

Authors:  Yasuhiro Heishima; Yasutomo Hori; Seishiro Chikazawa; Kazutaka Kanai; Fumio Hoshi; Naoyuki Itoh
Journal:  J Vet Med Sci       Date:  2016-03-18       Impact factor: 1.267

6.  Acute pharmacodynamic effects of pimobendan in client-owned cats with subclinical hypertrophic cardiomyopathy.

Authors:  Maureen S Oldach; Yu Ueda; Eric S Ontiveros; Samantha L Fousse; Lance C Visser; Joshua A Stern
Journal:  BMC Vet Res       Date:  2021-02-23       Impact factor: 2.741

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.