| Literature DB >> 32580610 |
Petra Cerna1,2, Panos E Archontakis3, Hester Ok Cheuk4, Danièlle A Gunn-Moore1,5.
Abstract
OBJECTIVES: This study compared Doppler and oscillometric (PetMAP+) devices (with or without proprietary optimisations) for the non-invasive measurement of blood pressure in conscious cats.Entities:
Keywords: Feline; NIBP; blood pressure; hypertension; indirect
Year: 2020 PMID: 32580610 PMCID: PMC7838334 DOI: 10.1177/1098612X20932407
Source DB: PubMed Journal: J Feline Med Surg ISSN: 1098-612X Impact factor: 2.015
Published arterial blood pressure (mmHg) values obtained from cats
| Study | Method | n | Systolic | Diastolic | Mean | Machine |
|---|---|---|---|---|---|---|
| Intra-arterial | ||||||
| Belew et al (1999)[ | Conscious | 13 | 126 ± 9 | 106 ± 10 | 91 ± 11 | Dinamap 8300 (Critikon) |
| Mishina et al (2006)[ | Conscious | 20 | 118 ± 11 | 78 ± 9 | 95 ± 10 | Model TA11 PA-C40 (Data Sciences International) |
| Slingerland et al (2008)[ | Conscious | 21 | 132 ± 9 | 115 ± 8 | 96 ± 8 | Gabarith (Becton Dickinson BV) |
| Jenkins et al (2015)[ | Anaesthetised | 6 | 160 ± 12 | 138 ± 11 | 116 ± 8 | Model TA11 PA-C40 (Data Sciences International) |
| Oscillometry | ||||||
| Bodey and Sansom (1998)[ | Conscious | 104 | 139 ± 27 | 99 ± 27 | 77 ± 25 | Dinamap 1846SX (Critikon) |
| Mishina et al (1998)[ | Conscious | 60 | 115 ± 10 | 96 ± 12 | 74 ± 11 | USM-700GTM (Ueda Electronic Works) |
| Haberman et al (2004)[ | Conscious and anaesthetised | 13 | 132 ± 24 | 111 ± 24 | 96 ± 22 | Dinamap 8300 (Critikon) |
| Doppler ultrasonography | ||||||
| Sparkes et al (1999)[ | Conscious | 50 | 162 ± 19 | Parks model 811-BTS (Parks Medical) | ||
| Haberman et al (2004)[ | Conscious and anaesthetised | 13 | 161 ± 46 | Parks model 811 (Parks Medical) | ||
| Lin et al (2006)[ | Conscious | 53 | 134 ± 16 | Parks model 811-B (Parks Medical) | ||
| Dos Reis et al (2014)[ | Conscious | 30 | 125 ± 16 | DV2000 | ||
| Chetboul et al (2014)[ | Conscious | 20 | 151 ± 5 | Parks model 811-BL (Parks Medical) | ||
| Payne et al (2017)[ | Conscious | 780 | 121 ± 16 | Parks model 811-B (Parks Medical) | ||
Data are mean ± SD unless otherwise indicated
IQR = interquartile range
Average of blood pressure measurements obtained from the 18 cats included in the study
| Signalment | Condition | Cuff size | Cuff location | Doppler | PetMAP+ | PetMAP+ | |||
|---|---|---|---|---|---|---|---|---|---|
| Age (years) | Type | Sex | |||||||
| 1 | 17 | Bengal | FN | Healthy | 2.0 | LF | 170 | 209/160 (178) | 183/134 (149) |
| 2 | 11 | Bengal | M | Vomiting | 3.0 | LF | 160 | 156/104 (124) | 182/126 (146) |
| 3 | 8 | DSH | MN | Healthy | 3.0 | LF | 139 | 138/92 (109) | 130/98 (108) |
| 4 | 3 | DSH | FN | Healthy | 2.5 | RF | 146 | 181/124[ | 189/99[ |
| 5 | 10 | DSH | MN | HCM | 2.5 | LF | 152 | 171/113 (132) | 163/114 (130) |
| 6 | 3 | DSH | FN | Lethargy | 3.0 | LF | 131 | 159/108 (126) | 148/101 (118) |
| 7 | 10 | DSH | MN | Jaundice | 2.5 | LF | 120 | 157/107 (128) | 159/99 (118) |
| 8 | 8 | DSH | MN | HCM | 3.0 | RF | 148 | 150/118 (133) | 160/104 (124) |
| 9 | 13 | DSH | FN | Skin wound | 2.5 | LF | 141 | 189/127 (149) | 175/131 (144) |
| 10 | 5 | DSH | MN | IMHA | 2.5 | LF | 148 | 167/82 (107) | 153/98 (116) |
| 11 | 4 | DSH | FN | Healthy | 2.5 | LF | 156 | 199/129 (155) | 185/99 (129) |
| 12 | 3 | DSH | M | RTA | 3.0 | RF | 139 | 163/116 (122) | 159/104 (120) |
| 13 | 12 | DSH | MN | Anaemia | 2.5 | RF | 147 | 142/104 (118) | 151/115 (124) |
| 14 | 4 | DSH | FN | RTA | 3.0 | LF | 130 | 123/53 (88) | 113/65 (84) |
| 15 | ‘Adult’ | DSH | − | Healthy | 3.0 | RF | 132 | 139/85 (120) | 155/76 (117) |
| 16 | 9 months | Bengal | F | Pelvic fracture | 2.5 | LF | 147 | 175/127 (144) | 166/116 (134) |
| 17 | 11 months | DSH | M | Patellar luxation | 3.0 | RF | 141 | 181/114 (139) | 131/85 (102) |
| 18 | 6 | DSH | MN | Vomiting | 2.5 | LF | 135 | 156/75 (110) | 162/113 (129) |
Mode A: Proprietary Optimizations (PPO); Mode B: Optimized None (ON)
Used first
Mean arterial pressure not recorded
FN = female neutered; LF = left forelimb; M = male; DSH = domestic shorthair; MN = male neutered; HCM = hypertrophic cardiomyopathy; RF = right forelimb; IMHA = immune-mediated anaemia; RTA = road traffic accident; F = female
A summary of Bland–Altman analyses of PetMAP+ Proprietary Optimizations (PPO) (mode A) and Optimized None (ON) (mode B) function vs Doppler in conscious cats
| Device | BP | Mean ± SD | Bias | Precision | LOA | Range |
|---|---|---|---|---|---|---|
| PetMAP+ | SAP | 164.17 ± 22.36 | + 21.28 | 17.67 | −13.36 to 55.92 | 72–243 |
| PetMAP+mode B | SAP | 159.17 ± 20.25 | + 15.72 | 16.61 | −16.83 to 48.28 | 50–245 |
| Doppler | SAP | 143.44 ± 11.99 | 110–190 | |||
| PetMAP+ | MAP | 128.35 ± 20.96 | −14.94 | 16.96 | −48.18 to 18.30 | 55–195 |
| PetMAP+ | MAP | 123.06 ± 16.24 | −20.24 | 12.72 | −45.17 to 4.70 | 30–176 |
Lowest and highest measurement recorded
BP = blood pressure; LOA = limits of agreement; SAP = systolic arterial pressure; MAP = mean arterial pressure
Figure 1Bland–Altman plot of the differences between systolic arterial blood pressures (SAPs) measured by the Doppler and PetMAP+ Proprietary Optimizations (PPO; mode A) oscillometric device; measurements from the two devices were plotted against the mean of the Doppler SAP measurements. Dashed line indicates the bias (average difference); dotted lines indicate the upper limit of agreement and lower limit of agreement
Figure 2Bland–Altman plot of differences between systolic arterial blood pressures (SAPs) measured by the Doppler and PetMAP+ Optimized None (ON; mode B) oscillometric device; measurements from the two devices were plotted against the mean of the Doppler SAP measurements. Dashed line indicates the bias (average difference); dotted lines indicate the upper limit of agreement and lower limit of agreement
Figure 3Bland–Altman plot of the differences between systolic arterial blood pressures (SAPs) measured by the Doppler and mean arterial pressure (MAP) by PetMAP+ Proprietary Optimizations (PPO; mode A) oscillometric device; measurements from the two devices were plotted against the mean of the Doppler SAP measurements. Dashed line indicates the bias (average difference); dotted lines indicate the upper limit of agreement and lower limit of agreement
Figure 4Bland–Altman plot of differences between systolic arterial blood pressures (SAPs) measured by the Doppler and mean arterial pressure (MAP) by PetMAP+ Optimized None (ON; mode B) oscillometric device; measurements from the two devices were plotted against the mean of the Doppler SAP measurements. Dashed line indicates the bias (average difference); dotted lines indicate the upper limit of agreement and lower limit of agreement