Literature DB >> 20681771

Feasibility and limitations of an automated 2D-3D rigid image registration system for complex endovascular aortic procedures.

Tom W G Carrell1, Bijan Modarai, James R I Brown, Graeme P Penney.   

Abstract

PURPOSE: To examine the feasibility of an automated 2-dimensional (2D) to 3- dimensional (3D) image registration system to simplify the navigational challenges faced in complex endovascular aortic procedures.
METHODS: An automated 2D-3D image registration system was used to overlay pre-acquired 3D computed tomography images onto fluoroscopy images taken during endovascular aneurysm repair. Errors between the 3D overlay and digital subtraction angiograms were measured and correlated with aortic neck angulation. A mean discrepancy < or =3 mm was considered clinically acceptable.
RESULTS: There was a strong correlation between maximum neck angulation and maximum registration error (Pearson's r = 0.75). Aortas with a maximum neck angulation < or =30 degrees had a mean error of 2.5+/-1.2 mm, whereas aortas with neck angulation >30 degrees had a mean error of 6.2+/-2.5 mm (p<0.0001).
CONCLUSION: The major source of registration errors is aortic deformation caused by the presence of the introducer and endovascular graft. Further work is required if this technology is to be routinely applied to severely angulated aortic anatomy.

Entities:  

Mesh:

Year:  2010        PMID: 20681771     DOI: 10.1583/09-2987MR.1

Source DB:  PubMed          Journal:  J Endovasc Ther        ISSN: 1526-6028            Impact factor:   3.487


  6 in total

Review 1.  Computer-aided surgery: concepts and applications in vascular surgery.

Authors:  Adrien Kaladji; Antoine Lucas; Alain Cardon; Pascal Haigron
Journal:  Perspect Vasc Surg Endovasc Ther       Date:  2012-04-18

Review 2.  Update on New Tools for Three-dimensional Navigation in Endovascular Procedures.

Authors:  Jos C van den Berg
Journal:  Aorta (Stamford)       Date:  2014-12-01

3.  Finite-element-based matching of pre- and intraoperative data for image-guided endovascular aneurysm repair.

Authors:  Aurélien Duménil; Adrien Kaladji; Miguel Castro; Simon Esneault; Antoine Lucas; Michel Rochette; Cemil Göksu; Pascal Haigron
Journal:  IEEE Trans Biomed Eng       Date:  2012-12-20       Impact factor: 4.538

4.  Accuracy of deformable image registration techniques for alignment of longitudinal cholangiocarcinoma CT images.

Authors:  Anando Sen; Brian M Anderson; Guillaume Cazoulat; Molly M McCulloch; Dalia Elganainy; Brigid A McDonald; Yulun He; Abdallah S R Mohamed; Baher A Elgohari; Mohamed Zaid; Eugene J Koay; Kristy K Brock
Journal:  Med Phys       Date:  2020-02-12       Impact factor: 4.071

Review 5.  Augmented Reality in Vascular and Endovascular Surgery: Scoping Review.

Authors:  Joshua Eves; Dimitri Amiras; Abhilash Sudarsanam; Joseph Shalhoub
Journal:  JMIR Serious Games       Date:  2022-09-23       Impact factor: 3.364

6.  Accuracy of registration techniques and vascular imaging modalities in fusion imaging for aortic endovascular interventions: a phantom study.

Authors:  M M Sieren; C Schareck; M Kaschwich; M Horn; F Matysiak; E Stahlberg; F Wegner; T H Oechtering; J Barkhausen; J Goltz
Journal:  CVIR Endovasc       Date:  2021-06-14
  6 in total