Defibrillation testing at the time of ICD insertion: an analysis from the Ontario ICD Registry.

J S Healey1, D H Birnie, D S Lee, A D Krahn, E Crystal, C S Simpson, P Dorian, Z Chen, D Cameron, A Verma, S J Connolly, L J Gula, E Lockwood, G Nair, J V Tu.   

Abstract

BACKGROUND: increasingly, ICD implantation is performed without defibrillation testing (DT).
OBJECTIVES: To determine the current frequency of DT, the risks associated with DT, and to understand how physicians select patients to have DT.
METHODS: between January 2007 and July 2008, all patients in Ontario, Canada who received an ICD were enrolled in this prospective registry.
RESULTS: a total of 2,173 patients were included; 58% had new ICD implants for primary prevention, 25% for secondary prevention, and 17% had pulse generator replacement. DT was carried out at the time of ICD implantation or predischarge in 65%, 67%, and 24% of primary, secondary, and replacement cases respectively (P = <0.0001). The multivariate predictors of a decision to conduct DT included: new ICD implant (OR = 13.9, P < 0.0001), dilated cardiomyopathy (OR = 1.8, P < 0.0001), amiodarone use (OR = 1.5, P = 0.004), and LVEF > 20% (OR = 1.3, P = 0.05). A history of atrial fibrillation (OR = 0.58, P = 0.0001) or oral anticoagulant use (OR = 0.75, P = 0.03) was associated with a lower likelihood of having DT. Age, gender, NYHA class, and history of stroke or TIA did not predict DT. Perioperative complications, including death, myocardial infarction, stroke, tamponade, pneumothorax, heart failure, infection, wound hematoma, and lead dislodgement, were similar among patients with (8.7%) and without (8.3%) DT (P = 0.7)
CONCLUSIONS: DT is performed in two-thirds of new ICD implants but only one-quarter of ICD replacements. Physicians favored performance of DT in patients who are at lower risk of DT-related complications and in those receiving amiodarone. DT was not associated with an increased risk of perioperative complications.

Entities:  

Mesh:

Year:  2010        PMID: 20662988     DOI: 10.1111/j.1540-8167.2010.01850.x

Source DB:  PubMed          Journal:  J Cardiovasc Electrophysiol        ISSN: 1045-3873


  7 in total

Review 1.  Complications of defibrillation testing. A review of the literature.

Authors:  Bishoy Deif; Jeff S Healey
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2015-11-30

2.  Shock or no shock - a question of philosophy or should intraoperative implantable cardioverter defibrillator testing be recommended?

Authors:  Andreas Keyser; Michael K Hilker; Sebastian Schmidt; Christian von Bary; Wolfgang Zink; Michael Ried; Christof Schmid; Claudius Diez
Journal:  Interact Cardiovasc Thorac Surg       Date:  2012-12-07

3.  How to Manage a High Defibrillation Threshold in ICD Patients: and Does it Really Matter?

Authors:  Maria Vittoria Matassini; Jeff S Healey
Journal:  Curr Treat Options Cardiovasc Med       Date:  2013-08

4.  2015 HRS/EHRA/APHRS/SOLAECE expert consensus statement on optimal implantable cardioverter-defibrillator programming and testing.

Authors:  Bruce L Wilkoff; Laurent Fauchier; Martin K Stiles; Carlos A Morillo; Sana M Al-Khatib; Jesœs Almendral; Luis Aguinaga; Ronald D Berger; Alejandro Cuesta; James P Daubert; Sergio Dubner; Kenneth A Ellenbogen; N A Mark Estes; Guilherme Fenelon; Fermin C Garcia; Maurizio Gasparini; David E Haines; Jeff S Healey; Jodie L Hurtwitz; Roberto Keegan; Christof Kolb; Karl-Heinz Kuck; Germanas Marinskis; Martino Martinelli; Mark McGuire; Luis G Molina; Ken Okumura; Alessandro Proclemer; Andrea M Russo; Jagmeet P Singh; Charles D Swerdlow; Wee Siong Teo; William Uribe; Sami Viskin; Chun-Chieh Wang; Shu Zhang
Journal:  J Arrhythm       Date:  2016-02-01

5.  Critical analysis of ineffective post implantation implantable cardioverter-defibrillator-testing.

Authors:  Markus Roos; J Christoph Geller; Marc-Alexander Ohlow
Journal:  World J Cardiol       Date:  2017-02-26

6.  Significance of intraoperative testing in right-sided implantable cardioverter-defibrillators.

Authors:  Andreas Keyser; Michael K Hilker; Ekrem Ucer; Sigrid Wittmann; Christof Schmid; Claudius Diez
Journal:  J Cardiothorac Surg       Date:  2013-04-11       Impact factor: 1.637

7.  Incidence of ineffective safety margin testing (<10 J) and efficacy of routine subcutaneous array insertion during implantable cardioverter defibrillator implantation.

Authors:  Marc-Alexander Ohlow; Marcus Roos; Bernward Lauer; J Christoph Geller
Journal:  Indian Pacing Electrophysiol J       Date:  2016-02-26
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.