Accuracy of proton beam range verification using post-treatment positron emission tomography/computed tomography as function of treatment site.

Antje-Christin Knopf1, Katia Parodi, Harald Paganetti, Thomas Bortfeld, Juliane Daartz, Martijn Engelsman, Norbert Liebsch, Helen Shih.   

Abstract

PURPOSE: For 23 patients, an off-line positron emission tomography scan and a computed tomography scan after proton radiotherapy was performed at the Massachusetts General Hospital to assess in vivo treatment verification. A well-balanced population of patients was investigated to assess the effect of the tumor location on the accuracy of the technique. METHODS AND MATERIALS: Range verification was achieved by comparing the measured positron emission tomography activity distributions with the corresponding Monte Carlo-simulated distributions. Observed differences in the distal end of the activity distributions were analyzed as potential indicators for the range differences between the actual delivered and planned dose.
RESULTS: The average spatial agreement between the measured and simulated activity distribution was within ±3 mm, and the corresponding average absolute agreement was within ±45% (derived from gamma index analysis). The mean absolute range deviation at 93 randomly chosen positions in 17 treatment fields delivered to 11 patients was 3.6 mm. Characteristic differences in the agreement of the measured and simulated activity distribution for the different tumor/target sites were found. This resulted from the different effect of factors such as biologic washout effects, motion, or limitations in the Monte Carlo-simulated activity patterns.
CONCLUSION: We found that intracranial and cervical spine patients can greatly benefit from off-line positron emission tomography and computed tomography range verification. However, for the successful application of the method to patients with abdominopelvic tumors, major technological and methodologic improvements are needed. Among the intracranial and cervical spine target sites, patients with arteriovenous malformations or metal implants represent groups that could especially benefit from the approach.
Copyright © 2011 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20646839     DOI: 10.1016/j.ijrobp.2010.02.017

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  15 in total

1.  Monitoring proton therapy with PET.

Authors:  H Paganetti; G El Fakhri
Journal:  Br J Radiol       Date:  2015-05-20       Impact factor: 3.039

Review 2.  Current status and future prospects of multi-dimensional image-guided particle therapy.

Authors:  Shinichiro Mori; Silvan Zenklusen; Antje-Christin Knopf
Journal:  Radiol Phys Technol       Date:  2013-02-19

3.  Monitoring proton radiation therapy with in-room PET imaging.

Authors:  Xuping Zhu; Samuel España; Juliane Daartz; Norbert Liebsch; Jinsong Ouyang; Harald Paganetti; Thomas R Bortfeld; Georges El Fakhri
Journal:  Phys Med Biol       Date:  2011-06-15       Impact factor: 3.609

4.  A full-scale clinical prototype for proton range verification using prompt gamma-ray spectroscopy.

Authors:  Fernando Hueso-González; Moritz Rabe; Thomas A Ruggieri; Thomas Bortfeld; Joost M Verburg
Journal:  Phys Med Biol       Date:  2018-09-17       Impact factor: 3.609

5.  Feasibility of proton-activated implantable markers for proton range verification using PET.

Authors:  Jongmin Cho; Geoffrey Ibbott; Michael Gillin; Carlos Gonzalez-Lepera; Uwe Titt; Harald Paganetti; Matthew Kerr; Osama Mawlawi
Journal:  Phys Med Biol       Date:  2013-10-08       Impact factor: 3.609

6.  Determination of elemental tissue composition following proton treatment using positron emission tomography.

Authors:  Jongmin Cho; Geoffrey Ibbott; Michael Gillin; Carlos Gonzalez-Lepera; Chul Hee Min; Xuping Zhu; Georges El Fakhri; Harald Paganetti; Osama Mawlawi
Journal:  Phys Med Biol       Date:  2013-05-16       Impact factor: 3.609

7.  The rationale for intensity-modulated proton therapy in geometrically challenging cases.

Authors:  S Safai; A Trofimov; J A Adams; M Engelsman; T Bortfeld
Journal:  Phys Med Biol       Date:  2013-08-22       Impact factor: 3.609

Review 8.  Latest developments in in-vivo imaging for proton therapy.

Authors:  Katia Parodi
Journal:  Br J Radiol       Date:  2019-12-12       Impact factor: 3.039

Review 9.  Range uncertainties in proton therapy and the role of Monte Carlo simulations.

Authors:  Harald Paganetti
Journal:  Phys Med Biol       Date:  2012-05-09       Impact factor: 3.609

Review 10.  Range Verification Methods in Particle Therapy: Underlying Physics and Monte Carlo Modeling.

Authors:  Aafke Christine Kraan
Journal:  Front Oncol       Date:  2015-07-07       Impact factor: 6.244

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.