Greta Snellman1, Håkan Melhus, Rolf Gedeborg, Liisa Byberg, Lars Berglund, Lisa Wernroth, Karl Michaëlsson.
Abstract
BACKGROUND: Vitamin D is not only important for bone health but can also affect the development of several non-bone diseases. The definition of vitamin D insufficiency by serum levels of 25-hydroxyvitamin D depends on the clinical outcome but might also be a consequence of analytical methods used for the definition. Although numerous 25-hydroxyvitamin D assays are available, their comparability is uncertain. We therefore aim to investigate the precision, accuracy and clinical consequences of differences in performance between three common commercially available assays. METHODOLOGY/PRINCIPALEntities:
Mesh:
Substances:
Year: 2010 PMID: 20644628 PMCID: PMC2903481 DOI: 10.1371/journal.pone.0011555
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Mean 25-hydroxyvitamin D (25(OH)D)) values and characteristics of the twins as a function of season.
| Total (n = 204) | ||
| Mean (SD) | Range | |
|
| 85.0 (27.4) | 21.4−181.3 |
|
| 83.3 (27.4) | 21.4−174.4 |
|
| 7.7 (2.8) | 4.6−13.9 |
|
| 70 (24.0) | 26.1−156.5 |
|
| 59.9 (26.1) | 10.0−172.2 |
|
| 1.9 (0.9) | 0.4−5.3 |
|
| 57.5 (9.7) | 37.8−84.5 |
|
| 74 (12.2) | 47.9−122.6 |
|
| 170.5 (9.7) | 149.0−194.5 |
|
| 25.4 (3.2) | 18.9−39.6 |
Mean 25-hydroxyvitamin D levels are adjusted for the twin-ship dependence.
*Results are based on the 40 (20%) participants who had measurable levels of 25-hydroxyvitamin D2.
Figure 1Mean serum 25-hydroxyvitamin D by assay.
The error bars indicate 95% confidence intervals.
Figure 2Seasonal differences in 25-hydroxyvitamin D levels for the HPLC-APCI-MS, RIA and CLIA assays.
The error bars indicate 95% confidence intervals.
Figure 3Cumulative proportion of the subjects who are classified as insufficient using a 50 nmol/L cut-off.
HPLC-APCI-MS 8%, RIA 22%, CLIA 43%.
Intraclass correlation coefficients (ICC) for serum 25-hydroxyvitamin D levels between twins in a pair with 95% confidence intervals.
| Variable | ICC (95% CI) | P-value | |
|
| 0.66 (0.54−0.76) | ref | NA |
|
| 0.54 (0.39−0.66) | 0.004 | ref |
|
| 0.40 (0.22−0.55) | <0.001 | <0.001 |
The p-values correspond to a test of equality of the observed correlations.
Sample coefficient of variation (SCV) values with 95% confidence intervals.
| Variable | SCV% (95% CI) | P-value | |
|
| 32.3 (28.5−36.0) | ref | NA |
|
| 34.2 (29.8−38.3) | 0.243 | ref |
|
| 43.5 (37.7−48.9) | <0.001 | 0.001 |
*Adjusted for twin-ship dependence.
Figure 4Bland Altman plots for the difference in 25-hydroxyvitamin D level between the assays.
HPLC-APCI-MS vs. RIA (panel A), HPLC-APCI-MS vs. CLIA (panel B) and RIA vs. CLIA (panel C). Each circle represents one twin.
Linear regression analyses for the relationship between the absolute differences85 in serum values between the methods against the mean of the two values.
| Variable | Slope | 95% CI | P-value |
|
| 0.19 | 0.09−0.29 | <0.0001 |
|
| 0.22 | 0.12−0.31 | <0.0001 |
*Adjusted for twin-ship dependence.