Effect of image-guidance frequency on geometric accuracy and setup margins in radiotherapy for locally advanced lung cancer.

Jane Higgins1, Andrea Bezjak, Andrew Hope, Tony Panzarella, Winnie Li, John B C Cho, Tim Craig, Anthony Brade, Alexander Sun, Jean-Pierre Bissonnette.   

Abstract

PURPOSE: To assess the relative effectiveness of five image-guidance (IG) frequencies on reducing patient positioning inaccuracies and setup margins for locally advanced lung cancer patients. METHODS AND MATERIALS: Daily cone-beam computed tomography data for 100 patients (4,237 scans) were analyzed. Subsequently, four less-than-daily IG protocols were simulated using these data (no IG, first 5-day IG, weekly IG, and alternate-day IG). The frequency and magnitude of residual setup error were determined. The less-than-daily IG protocols were compared against the daily IG, the assumed reference standard. Finally, the population-based setup margins were calculated.
RESULTS: With the less-than-daily IG protocols, 20-43% of fractions incurred residual setup errors ≥ 5 mm; daily IG reduced this to 6%. With the exception of the first 5-day IG, reductions in systematic error (∑) occurred as the imaging frequency increased and only daily IG provided notable random error (σ) reductions (∑ = 1.5-2.2 mm, σ = 2.5-3.7 mm; ∑ = 1.8-2.6 mm, σ = 2.5-3.7 mm; and ∑ = 0.7-1.0 mm, σ = 1.7-2.0 mm for no IG, first 5-day IG, and daily IG, respectively. An overall significant difference in the mean setup error was present between the first 5-day IG and daily IG (p < .0001). The derived setup margins were 5-9 mm for less-than-daily IG and were 3-4 mm with daily IG.
CONCLUSION: Daily cone-beam computed tomography substantially reduced the setup error and could permit setup margin reduction and lead to a reduction in normal tissue toxicity for patients undergoing conventionally fractionated lung radiotherapy. Using first 5-day cone-beam computed tomography was suboptimal for lung patients, given the inability to reduce the random error and the potential for the systematic error to increase throughout the treatment course.
Copyright © 2011 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2010        PMID: 20643515     DOI: 10.1016/j.ijrobp.2010.04.006

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  26 in total

1.  Localization accuracy from automatic and semi-automatic rigid registration of locally-advanced lung cancer targets during image-guided radiation therapy.

Authors:  Scott P Robertson; Elisabeth Weiss; Geoffrey D Hugo
Journal:  Med Phys       Date:  2012-01       Impact factor: 4.071

2.  Risk factors related to interfractional variation in whole pelvic irradiation for locally advanced pelvic malignancies.

Authors:  W S Yoon; D S Yang; J A Lee; S Lee; Y J Park; C Y Kim
Journal:  Strahlenther Onkol       Date:  2012-02-10       Impact factor: 3.621

3.  Motion management strategies and technical issues associated with stereotactic body radiotherapy of thoracic and upper abdominal tumors: A review from NRG oncology.

Authors:  Edward D Brandner; Indrin J Chetty; Tawfik G Giaddui; Ying Xiao; M Saiful Huq
Journal:  Med Phys       Date:  2017-04-20       Impact factor: 4.071

4.  Respiratory triggered 4D cone-beam computed tomography: a novel method to reduce imaging dose.

Authors:  Benjamin J Cooper; Ricky T O'Brien; Salim Balik; Geoffrey D Hugo; Paul J Keall
Journal:  Med Phys       Date:  2013-04       Impact factor: 4.071

5.  Accuracy in automatic image registration between MV cone beam computed tomography and planning kV computed tomography in image guided radiotherapy.

Authors:  Nithya Kanakavelu; E James Jebaseelan Samuel
Journal:  Rep Pract Oncol Radiother       Date:  2016-07-22

6.  Modern radiotherapy using image guidance for unresectable non-small cell lung cancer can improve outcomes in patients treated with chemoradiation therapy.

Authors:  Matthew P Deek; Sinae Kim; Ning Yue; Rekha Baby; Inaya Ahmed; Wei Zou; John Langenfeld; Joseph Aisner; Salma K Jabbour
Journal:  J Thorac Dis       Date:  2016-09       Impact factor: 2.895

Review 7.  Image-guided radiation therapy (IGRT): practical recommendations of Italian Association of Radiation Oncology (AIRO).

Authors:  Paola Franzone; Alba Fiorentino; Salvina Barra; Domenico Cante; Laura Masini; Elena Cazzulo; Liana Todisco; Pietro Gabriele; Elisabetta Garibaldi; Anna Merlotti; Maria Grazia Ruo Redda; Filippo Alongi; Renzo Corvò
Journal:  Radiol Med       Date:  2016-09-06       Impact factor: 3.469

Review 8.  Magnetic resonance imaging in precision radiation therapy for lung cancer.

Authors:  Hannah Bainbridge; Ahmed Salem; Rob H N Tijssen; Michael Dubec; Andreas Wetscherek; Corinne Van Es; Jose Belderbos; Corinne Faivre-Finn; Fiona McDonald
Journal:  Transl Lung Cancer Res       Date:  2017-12

9.  A study of respiration-correlated cone-beam CT scans to correct target positioning errors in radiotherapy of thoracic cancer.

Authors:  J P Santoro; J McNamara; E Yorke; H Pham; A Rimner; K E Rosenzweig; G S Mageras
Journal:  Med Phys       Date:  2012-10       Impact factor: 4.071

10.  Technical note: improved positioning protocol for patient setup accuracy in conventional radiotherapy for lung cancer.

Authors:  Hongbo Chai; Yuichiro Narita; Masafumi Takagi; Mikiko Kudo; Tomomi Kimura; Keiichi Kattou
Journal:  Radiol Phys Technol       Date:  2019-09-23
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.