Literature DB >> 20587649

Evaluation of demineralized bone and bone transplants in vitro and in vivo with cone beam computed tomography imaging.

F G Draenert1, F Gebhart, M Berthold, M Gosau, W Wagner.   

Abstract

OBJECTIVES: The objective of this study was to determine the ability of two flat panel cone beam CT (CBCT) devices to identify demineralized bone and bone transplants in vivo and in vitro.
METHODS: Datasets from patients with autologous bone grafts (n = 9, KaVo 3DeXam (KaVo, Biberach, Germany); n = 38, Accuitomo 40 (Morita, Osaka, Japan)) were retrospectively evaluated. Demineralized and non-demineralized porcine cancellous bone blocks were examined with the two CBCT devices. A SawBone skull (Pacific Research Laboratories, Vashon, WA) was used as a positioning tool for the bone blocks. Descriptive evaluation and image quality assessment were conducted on the KaVo 3DeXam data (voxel size 0.3 mm) using the OsiriX viewer as well as on the Morita Accuitomo data (voxel size 0.25 mm) using proprietary viewer software.
RESULTS: Both in vivo and in vitro, the descriptive analysis of the images of the two devices showed well-visualized bone transplants with clearly defined cancellous bones and well-defined single bone trabeculae in all cross-sections. In vitro, demineralized samples showed lower radiographic opacity but no significant loss of quality compared with fresh bone (P = 0.070). Single cancellous bone trabeculae were significantly better visualized with the Morita 3D Accuitomo device than with the KaVo 3DeXam device (P = 0.038).
CONCLUSION: Both the KaVo 3DeXam and Morita 3D Accuitomo devices produce good-quality images of cancellous bones in in vivo remodelling as well as after in vitro demineralization.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20587649      PMCID: PMC3520252          DOI: 10.1259/dmfr/147454468

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  18 in total

Review 1.  E.A.O. guidelines fo the use of diagnostic imaging in implant dentistry. A consensus workshop organized by the European Association for Osseointegration in Trinity College Dublin.

Authors:  David Harris; Daniel Buser; Karl Dula; Kerstin Grondahl; David Haris; Reinhilde Jacobs; Ulf Lekholm; Richard Nakielny; Daniel van Steenberghe; Paul van der Stelt
Journal:  Clin Oral Implants Res       Date:  2002-10       Impact factor: 5.977

2.  Evaluation of quality of reformatted images from two cone-beam computed tomographic systems.

Authors:  Dirk Schulze; Max Heiland; Felix Blake; Uwe Rother; Rainer Schmelzle
Journal:  J Craniomaxillofac Surg       Date:  2005-01-12       Impact factor: 2.078

3.  Structural and radiological parameters for the characterization of jawbone.

Authors:  Nele Stoppie; Veerle Pattijn; Tim Van Cleynenbreugel; Martine Wevers; Jos Vander Sloten; Naert Ignace
Journal:  Clin Oral Implants Res       Date:  2006-04       Impact factor: 5.977

4.  Flat panel computed tomography for non-invasive flow measurement: initial results in in-vitro studies.

Authors:  Andreas H Mahnken; Michael Grasruck; Bernhard Schmidt; Rolf W Günther; Joachim E Wildberger
Journal:  Eur Radiol       Date:  2007-12-04       Impact factor: 5.315

5.  [Osirix: free and open-source software for medical imagery].

Authors:  F Jalbert; J R Paoli
Journal:  Rev Stomatol Chir Maxillofac       Date:  2008-02-21

6.  Cervical soft tissue imaging using a mobile CBCT scanner with a flat panel detector in comparison with corresponding CT and MRI data sets.

Authors:  Max Heiland; Philipp Pohlenz; Marco Blessmann; Christian R Habermann; Lars Oesterhelweg; Philipp C Begemann; Christian Schmidgunst; Felix A S Blake; Klaus Püschel; Rainer Schmelzle; Dirk Schulze
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod       Date:  2007-09-27

Review 7.  New challenges in MDCT.

Authors:  Mathias Prokop
Journal:  Eur Radiol       Date:  2005-12       Impact factor: 5.315

8.  Imaging of bone transplants in the maxillofacial area by NewTom 9000 cone-beam computed tomography: a quality assessment.

Authors:  Florian G Draenert; Florian Gebhart; Christoph Neugebauer; Eva Coppenrath; Ulrich Mueller-Lisse
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod       Date:  2008-07

9.  A new volumetric CT machine for dental imaging based on the cone-beam technique: preliminary results.

Authors:  P Mozzo; C Procacci; A Tacconi; P T Martini; I A Andreis
Journal:  Eur Radiol       Date:  1998       Impact factor: 5.315

10.  Demineralization with EDTA by constant replacement.

Authors:  L F Bélanger; D H Copp; M A Morton
Journal:  Anat Rec       Date:  1965-09
View more
  2 in total

1.  Assessment of buccal marginal alveolar peri-implant and periodontal defects using a cone beam CT system with and without the application of metal artefact reduction mode.

Authors:  K Kamburoglu; E Kolsuz; S Murat; H Eren; S Yüksel; C S Paksoy
Journal:  Dentomaxillofac Radiol       Date:  2013       Impact factor: 2.419

2.  Comparative assessment of 3D reconstruction technique and Cavalieri's principle in predicting the mandibular bone defect volumes.

Authors:  Mehmet Ali Altay; Faisal A Quereshy; Sumit K Nijhawan; Jose F Teppa; Michael P Horan; Nelli Yıldırımyan; Dale A Baur
Journal:  Eur Oral Res       Date:  2018-05-01
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.