Literature DB >> 20587254

Simulation Study of Distal Balloon Protection Systems during Carotid Artery Stenting.

T Kuroiwa1, N Sakai, M Sakaguchi, H Adachi, H Imamura, C Sakai, A Morizane, H Ishihara, T Yano, S Nakao, H Kikuchi.   

Abstract

SUMMARY: The purpose of our experimental study was to assess the pitfalls of distal balloon protection systems and to learn any technique tips to increase safety. Silicone carotid artery models were connected to a circulatory system to simulate arterial flow. A distal balloon protection device, PercuSurge GuardWire Plus (GWP, Medtronic Vascular) was delivered to the internal carotid artery (ICA), then was inflated to occlude ICA flow temporarily. A debris aspiration catheter (Export catheter) was delivered just proximal to the GuardWire Plus balloon coaxially, in order to introduce and diffuse particulate debris (20-500micro meter in diameter) in the ICA stump. Then, after debris in the stump was aspirated, the GWP balloon was deflated. We recorded all the processes of our simulation experiments on a digital video and observed the movements of debris during these experiments. Exp 1) We simulated the movements of debris in the ICA stump when the GWP balloon was gradually deflated to produce a crevice between the balloon and vessel wall, simulating accidental movement of the GWP balloon during the procedure. Exp 2) In order to assess the optimal placement of the tip of aspiration catheter, the debris in the ICA stump was aspirated from three different sites(from just below the GuardWire balloon, from 2cm below it, and from 5cm below it). Exp 1) When the crevice appeared between GuardWire balloon and silicone tube, simulated debris began to concentrate just below the balloon. Then, some debris migrated distally from the crevice, and another part crowded in the crevice so that was impossible to aspirate and migrated in the end. Exp. 2) Debris aspiration was most effective from immediately below the GuardWire balloon, and the aspiration ability declined as the distance between the balloon and aspiration catheter became longer. According to our simulation studies, when the GWP balloon was moved accidentally during CAS procedures, or when the aspiration catheter was not delivered all the way to the GuardWire Plus balloon, distal embolization might still occur, even under protection.

Entities:  

Year:  2008        PMID: 20587254      PMCID: PMC3522226          DOI: 10.1177/15910199040100S215

Source DB:  PubMed          Journal:  Interv Neuroradiol        ISSN: 1591-0199            Impact factor:   1.610


  14 in total

Review 1.  Critical analysis of distal protection devices.

Authors:  Takao Ohki; Frank J Veith
Journal:  Semin Vasc Surg       Date:  2003-12       Impact factor: 1.000

2.  Cerebral protection with filter devices during carotid artery stenting.

Authors:  B Reimers; N Corvaja; S Moshiri; S Saccà; R Albiero; C Di Mario; P Pascotto; A Colombo
Journal:  Circulation       Date:  2001-07-03       Impact factor: 29.690

3.  Carotid artery stenting protected with an emboli containment system.

Authors:  Patrick L Whitlow; Pedro Lylyk; Hugo Londero; Oscar A Mendiz; Klaus Mathias; Horst Jaeger; Juan Parodi; Claudio Schönholz; Jose Milei
Journal:  Stroke       Date:  2002-05       Impact factor: 7.914

4.  Angioplasty and stenting of the extracranial carotid arteries.

Authors:  M Henry; M Amor; C Klonaris; I Henry; I Masson; Z Chati; E Leborgne; M Hugel
Journal:  Tex Heart Inst J       Date:  2000

5.  Use of the Parodi anti-embolism system in carotid stenting: Italian trial results.

Authors:  Carlo A Adami; Alberto Scuro; Luca Spinamano; Elisabetta Galvagni; Davide Antoniucci; Gian A Farello; Franco Maglione; Stefano Manfrini; Nicola Mangialardi; Gian C Mansueto; Francesco Mascoli; Ettore Nardelli; Domenico Tealdi
Journal:  J Endovasc Ther       Date:  2002-04       Impact factor: 3.487

6.  Balloon-protected carotid angioplasty.

Authors:  F C Albuquerque; G P Teitelbaum; S D Lavine; D W Larsen; S L Giannotta
Journal:  Neurosurgery       Date:  2000-04       Impact factor: 4.654

7.  Carotid artery stenosis: treatment with protected balloon angioplasty and stent placement.

Authors:  J G Theron; G G Payelle; O Coskun; H F Huet; L Guimaraens
Journal:  Radiology       Date:  1996-12       Impact factor: 11.105

8.  Protected carotid stenting: clinical advantages and complications of embolic protection devices in 442 consecutive patients.

Authors:  Alberto Cremonesi; Raffaella Manetti; Francesco Setacci; Carlo Setacci; Fausto Castriota
Journal:  Stroke       Date:  2003-07-03       Impact factor: 7.914

9.  Feasibility and efficacy of balloon-based neuroprotection during carotid artery stenting in a single-center setting.

Authors:  Michael Schlüter; Thilo Tübler; Detlef G Mathey; Joachim Schofer
Journal:  J Am Coll Cardiol       Date:  2002-09-04       Impact factor: 24.094

10.  Focal ischemia of the brain after neuroprotected carotid artery stenting.

Authors:  Michael Schlüter; Thilo Tübler; Johann C Steffens; Detlef G Mathey; Joachim Schofer
Journal:  J Am Coll Cardiol       Date:  2003-09-17       Impact factor: 24.094

View more