Dosimetric properties of Gafchromic EBT films in monoenergetic medical ion beams.

Mária Martisíková1, Oliver Jäkel.   

Abstract

For quality assurance measurements in radiotherapy with scanned ion beams, such as beam profile measurements and dosimetrical verification of patient plans, high spatial accuracy is required. We investigated the Gafchromic EBT film, which is the ancestor of the currently available EBT2 film, for this purpose. The excellent spatial resolution and improvements in sensitivity as compared to older types of radiochromic films make it a suitable candidate. In this contribution a study of the dose-response relation for monoenergetic beams (plateau), based on the absolute dose, is presented. We found the film response to be comparable for protons and photons. For carbon ions we observed a decrease in the darkening efficiency of about 30% with respect to photons. This effect is likely to be caused by the ion track structure. We did not find any dependence of the relative efficiency on the applied dose and energy over typical ranges used in ion beam therapy. Furthermore, it is shown that overlaps of ion tracks are not an issue for EBT film dosimetry in medical ion beams. The above findings enable the use of EBT films calibrated in a beam of a particular ion type for dose measurements at any other dose and energy. The uncertainty of the film darkening determination in ion beams is also discussed.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20530848     DOI: 10.1088/0031-9155/55/13/011

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  7 in total

1.  Investigation of EBT2 and EBT3 films for proton dosimetry in the 4-20 MeV energy range.

Authors:  S Reinhardt; M Würl; C Greubel; N Humble; J J Wilkens; M Hillbrand; A Mairani; W Assmann; K Parodi
Journal:  Radiat Environ Biophys       Date:  2015-01-09       Impact factor: 1.925

2.  Experimental and Monte Carlo characterization of a dynamic collimation system prototype for pencil beam scanning proton therapy.

Authors:  Blake R Smith; Mark Pankuch; Daniel E Hyer; Wesley S Culberson
Journal:  Med Phys       Date:  2020-09-09       Impact factor: 4.071

3.  LET dependent response of GafChromic films investigated with MeV ion beams.

Authors:  V Grilj; D J Brenner
Journal:  Phys Med Biol       Date:  2018-12-18       Impact factor: 3.609

4.  The accuracy of helium ion CT based particle therapy range prediction: an experimental study comparing different particle and x-ray CT modalities.

Authors:  L Volz; C-A Collins-Fekete; E Bär; S Brons; C Graeff; R P Johnson; A Runz; C Sarosiek; R W Schulte; J Seco
Journal:  Phys Med Biol       Date:  2021-11-29       Impact factor: 3.609

5.  Radiological properties of nanocomposite Fricke gel dosimeters for heavy ion beams.

Authors:  Takuya Maeyama; Nobuhisa Fukunishi; Kenichi L Ishikawa; Kazuaki Fukasaku; Shigekazu Fukuda
Journal:  J Radiat Res       Date:  2016-03-10       Impact factor: 2.724

6.  Linear energy transfer-independent calibration of radiochromic film for carbon-ion beams.

Authors:  Mutsumi Tashiro; Motohiro Kawashima
Journal:  Phys Imaging Radiat Oncol       Date:  2022-08-08

7.  Clinical and Research Activities at the CATANA Facility of INFN-LNS: From the Conventional Hadrontherapy to the Laser-Driven Approach.

Authors:  Giuseppe A P Cirrone; Giacomo Cuttone; Luigi Raffaele; Vincenzo Salamone; Teresio Avitabile; Giuseppe Privitera; Corrado Spatola; Antonio G Amico; Giuseppina Larosa; Renata Leanza; Daniele Margarone; Giuliana Milluzzo; Valeria Patti; Giada Petringa; Francesco Romano; Andrea Russo; Antonio Russo; Maria G Sabini; Valentina Scuderi; Francesco Schillaci; Lucia M Valastro
Journal:  Front Oncol       Date:  2017-09-19       Impact factor: 6.244

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.