Literature DB >> 20528892

Histological, histomorphometrical, and radiological evaluation of an experimental implant design with a high insertion torque.

Joke Duyck1, Livía Corpas, Stephanie Vermeiren, Toru Ogawa, Marc Quirynen, Katleen Vandamme, Reinhilde Jacobs, Ignace Naert.   

Abstract

OBJECTIVES: The aim of this study was to compare bone behaviour around an experimental implant design with a high insertion torque with the Astra-Tech implant (control).
MATERIALS AND METHODS: In ten 18-month-old male minipigs, the last premolars and first molars were extracted to provide space for two implants in each quadrant. A first set of 40 implants were placed 3 months after the extraction and 40 additional implants were installed another 2 months later. The animals were sacrificed 3 months after the first implant installation so that half of the implants had healed for 1 month and the other half for 3 months. Radiological evaluation was performed at baseline, 1 month, 2 months, and 3 months after implant installation. Bone defect depth and area, bone level changes, bone-to-implant contact density, and peri-implant bone fraction were measured histomorphometrically.
RESULTS: Radiological analyses revealed a significantly higher bone loss around the experimental implants. Histomorphometric analyses confirmed significantly more bone loss, larger marginal bone defects, and a lower overall peri-implant bone fraction around the experimental implants.
CONCLUSION: The experimental implant design caused significantly more peri-implant bone loss compared with the control implant. As strain gauge measurements indicate excessive marginal strains around the experimental implants, osseocompression might have played a role in the observed marginal bone loss.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20528892     DOI: 10.1111/j.1600-0501.2010.01895.x

Source DB:  PubMed          Journal:  Clin Oral Implants Res        ISSN: 0905-7161            Impact factor:   5.977


  11 in total

1.  Evaluation of dental implant insertion torque using a manual ratchet.

Authors:  M M Goswami; Mukul Kumar; Abhinav Vats; A S Bansal
Journal:  Med J Armed Forces India       Date:  2013-09-24

2.  Multiscale analyses of the bone-implant interface.

Authors:  J Y Cha; M D Pereira; A A Smith; K S Houschyar; X Yin; S Mouraret; J B Brunski; J A Helms
Journal:  J Dent Res       Date:  2015-01-27       Impact factor: 6.116

Review 3.  Biomechanical behaviours of the bone-implant interface: a review.

Authors:  Xing Gao; Manon Fraulob; Guillaume Haïat
Journal:  J R Soc Interface       Date:  2019-07-31       Impact factor: 4.118

4.  Mechanical micromodeling of stress-shielding at the bone-implant interphase under shear loading.

Authors:  Yoann Hériveaux; Sophie Le Cann; Manon Fraulob; Elsa Vennat; Vu-Hieu Nguyen; Guillaume Haïat
Journal:  Med Biol Eng Comput       Date:  2022-09-28       Impact factor: 3.079

5.  The survival rates and risk factors of implants in the early stage: a retrospective study.

Authors:  Yong Yang; Huiting Hu; Mianyan Zeng; Hongxing Chu; Zekun Gan; Jianmin Duan; Mingdeng Rong
Journal:  BMC Oral Health       Date:  2021-06-09       Impact factor: 2.757

6.  Effect of Surgical Installation of Dental Implants on Surface Topography and Its Influence on Osteoblast Proliferation.

Authors:  Helder H M Menezes; Marina M Naves; Henara L Costa; Tarsis P Barbosa; Jéssica A Ferreira; Denildo Magalhães; Elizabeth F Martinez
Journal:  Int J Dent       Date:  2018-10-17

7.  Intraosseous Temperature Change during Installation of Dental Implants with Two Different Surfaces and Different Drilling Protocols: An In Vivo Study in Sheep.

Authors:  Michele Stocchero; Yohei Jinno; Marco Toia; Marianne Ahmad; Evaggelia Papia; Satoshi Yamaguchi; Jonas P Becktor
Journal:  J Clin Med       Date:  2019-08-11       Impact factor: 4.241

8.  Comparative study of stress characteristics in surrounding bone during insertion of dental implants of three different thread designs: A three-dimensional dynamic finite element study.

Authors:  Chaiwat Udomsawat; Pimduen Rungsiyakull; Chaiy Rungsiyakull; Pathawee Khongkhunthian
Journal:  Clin Exp Dent Res       Date:  2018-12-26

9.  Elastography of the bone-implant interface.

Authors:  Yoann Hériveaux; Vu-Hieu Nguyen; Didier Geiger; Guillaume Haïat
Journal:  Sci Rep       Date:  2019-10-02       Impact factor: 4.379

10.  Bone healing at non-submerged implants installed with different insertion torques: a split-mouth histomorphometric randomized controlled trial.

Authors:  Yoshiyuki Amari; Adriano Piattelli; Karol Alí Apaza Alccayhuaman; Natalia Fortich Mesa; Mauro Ferri; Giovanna Iezzi; Daniele Botticelli
Journal:  Int J Implant Dent       Date:  2019-12-05
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.