Literature DB >> 20508051

Clinical evaluation of the MPS 9000 Macular Pigment Screener.

Hannah Bartlett1, Louise Stainer, Sandip Singh, Frank Eperjesi, Olivia Howells.   

Abstract

BACKGROUND/AIMS: The MPS 9000 uses a psychophysical technique known as heterochromatic flicker photometry to measure macular pigment optical density (MPOD). Our aim was to determine the measurement variability (noise) of the MPS 9000.
METHODS: Forty normally sighted participants who ranged in age from 18 to 50 years (25.4+/-8.2 years) were recruited from staff and students of Aston University (Birmingham, UK). Data were collected by two operators in two sessions separated by 1 week in order to assess test repeatability and reproducibility.
RESULTS: The overall mean MPOD for the cohort was 0.35+/-0.14. There was no significant negative correlation between MPS 9000 MPOD readings and age (r=-0.192, p=0.236). Coefficients were 0.33 and 0.28 for repeatability, and 0.25 and 0.26 for reproducibility. There was no significant correlation between mean and difference MPOD values for any of the four pairs of results.
CONCLUSIONS: When MPOD is being monitored over time then any change less than 0.33 units should not be considered clinically significant as it is very likely to be due to measurement noise. The size of the coefficient appears to be positively correlated with MPOD.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20508051     DOI: 10.1136/bjo.2009.175901

Source DB:  PubMed          Journal:  Br J Ophthalmol        ISSN: 0007-1161            Impact factor:   4.638


  6 in total

1.  The effect of nutritional supplementation on the multifocal electroretinogram in healthy eyes.

Authors:  Emma J Berrow; Hannah E Bartlett; Frank Eperjesi
Journal:  Doc Ophthalmol       Date:  2016-03-17       Impact factor: 2.379

2.  Understanding reproducibility of human IVF traits to predict next IVF cycle outcome.

Authors:  Bin Wu; Juanzi Shi; Wanqiu Zhao; Suzhen Lu; Marta Silva; Timothy J Gelety
Journal:  J Assist Reprod Genet       Date:  2014-08-15       Impact factor: 3.412

3.  Improving the repeatability of heterochromatic flicker photometry for measurement of macular pigment optical density.

Authors:  Olivia Howells; Frank Eperjesi; Hannah Bartlett
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2012-08-17       Impact factor: 3.117

4.  Repeatability of the macular pigment spatial profile: A comparison of objective versus subjective classification.

Authors:  Irene Ctori; Omar A Mahroo; Katie M Williams; Christopher J Hammond; Byki Huntjens
Journal:  Acta Ophthalmol       Date:  2018-08-29       Impact factor: 3.761

5.  Associations between macular pigment, iris color and reflectance, ethnicity, and color vision: An observational study.

Authors:  Roya Garakani; Jason S Ng
Journal:  PLoS One       Date:  2019-08-08       Impact factor: 3.240

6.  Macular pigment optical density: repeatability, intereye correlation, and effect of ocular dominance.

Authors:  Pinakin Gunvant Davey; Silverio D Alvarez; Jessica Y Lee
Journal:  Clin Ophthalmol       Date:  2016-08-29
  6 in total