Macular pigment optical density: repeatability, intereye correlation, and effect of ocular dominance.

Pinakin Gunvant Davey1, Silverio D Alvarez1, Jessica Y Lee1.   

Abstract

PURPOSE: To evaluate short-term repeatability, intereye correlation, and effect of ocular dominance on macular pigment optical density (MPOD) measurements obtained using the QuantifEye Heterochromatic Flicker Photometer. PATIENTS AND METHODS: A total of 72 study participants were enrolled in this prospective, cross-sectional study. Participants underwent a comprehensive ocular evaluation, including visual acuity, evaluation of ocular dominance, slit lamp examination, intraocular pressure measurement, and optic nerve head and macula analysis using optical coherence tomography and fundus photography. All study participants after initial training underwent MPOD measurement twice in both eyes in a randomized sequence. The repeatability was tested using Altman and Bland plots for first measurements with the second measurements for right eye and left eye and additionally by grouping eyes as a function of ocular dominance. The Pearson correlation coefficient was performed to assess the intereye correlation of MPOD values.
RESULTS: The mean age of study participants was 35.5 years (range 22-68 years). The mean MPOD measurements for OD (right eye) and OS (left eye) were 0.47 and 0.48, respectively, which followed a normal distribution (Shapiro-Wilk test, P=0.6 and 0.2). The 95% limits of agreement of Altman and Bland plots for the first and second measurements were -0.12 to +0.11 and -0.13 to +0.12 for OD and OS, respectively. The correlation coefficient of mean MPOD measurements of OD and OS was r statistic =0.94 (Pearson correlation coefficient P<0.0001; r (2) 0.89). The 95% limits of agreement of Altman and Bland plots when evaluated by laterality of eye or by ocular dominance were narrow, with limits of agreement ranging from -0.13 to +0.12.
CONCLUSION: The MPOD measurements obtained using the QuantifEye show good short-term repeatability. There is excellent intereye correlation, indicating that the MPOD values of one eye data can predict the fellow eye value with 89% accuracy. The ocular dominance had no bearing on the outcome of this psychophysical test in ocular healthy eyes.

Entities:  

Keywords:  QuantifEye; age-related macular degeneration; clinical protocol; heterochromatic flicker photometry; macular pigment optical density; ocular dominance; repeatability

Year:  2016        PMID: 27621586      PMCID: PMC5010153          DOI: 10.2147/OPTH.S111708

Source DB:  PubMed          Journal:  Clin Ophthalmol        ISSN: 1177-5467


Introduction

Age-related macular degeneration (AMD) is a leading cause of blindness in developed countries worldwide.1–7 AMD is broadly classified into two categories, exudative AMD and nonexudative AMD.8 Estimates indicate that in the US there were 2 million patients with late AMD and 8 million patients with intermediate AMD in 2004, and it is predicted that this number will double in the year 2024.9 There are numerous breakthroughs in the understanding of pathogenesis of wet AMD, and treatments have been developed in the form of antivascular endothelial growth factors.8 However, the understanding of the disease pathophysiology of the more common variety of AMD, the nonexudative AMD, is still lacking8 and the literature is not consistent on the efficacy of treatment.10–12 The Age-Related Eye Study-2 (AREDS-2)10 investigated the role of nutritional supplements in the progression of AMD; they particularly investigated the role of carotenoids, lutein, and zeaxanthin along with omega-3 in the prevention of progression in AMD.10 The AREDS-2 investigators concluded that taking oral supplements of carotenoids such as lutein and zeaxanthin may have beneficial effect in preventing progression to the late-stage AMD.10 Various studies have shown that oral supplementation of carotenoids can increase the measured level of macular pigment optical density (MPOD) to a variable degree,13–20 and thus MPOD measurements can be one of the ways to evaluate the treatment efficacy of oral supplementation of carotenoids. MPOD can be measured clinically using various objective methods and methods requiring subjective response.21–24 The QuantifEye (Zeavision, St Louis, MO, USA) is one of the devices that can estimate MPOD clinically.25–29 Studies have utilized the QuantifEye for measurement of MPOD repeatability;25–29 however, data on repeatability and on standardized technique that can be utilized in clinics to measure MPOD are limited. We sought to evaluate the QuantifEye in measuring MPOD in a group of ocular healthy individuals to get an understanding of the variability of measurement of MPOD due to human and ocular factors, with the idea of establishing protocols that can aid future studies. We investigated repeatability of measurements, intereye variability, and the effect of ocular dominance on MPOD measurements.

Patients and methods

The study was approved by the institutional review board at Western University of Health Sciences, Pomona, California, USA, and conducted in accordance with the tenets of the Declaration of Helsinki. A group of 72 individuals were evaluated for the study and all participants provided signed informed consent for the study. All participants were at least 18 years of age and were deemed ocular healthy by a dilated comprehensive examination in the last 1 year. Additionally, all individuals had a study-specific eye examination that included visual acuity measurement using the logMAR charts, ocular examination using the slit lamp biomicroscope, tonometry using the Ocular Response Analyzer, and optic disk and macula examination using the optical coherence tomography and fundus photography of the posterior pole. Participants with diabetic retinopathy, macular degeneration, macular edema, prior history of retinal detachment or vitreoretinal surgeries, or other congenital retinal pathologies or significant media opacity that could influence the outcome of the study were not included. The study participants in the immediate postoperative follow-up period (3 months) of cataract surgery or invasive ocular surgery were also not included. The presence of retinal disease or glaucoma was evaluated objectively using optical coherence tomography scans and fundus photographs during the study visit. All participants had a logMAR visual acuity of +0.2 (20/32) or better. The mean visual acuity of the study population was 0.003 (standard deviation [SD] 0.08) and 0.002 (SD 0.09) for the right and left eyes, respectively. The mean logMAR visual acuity was not significantly different on the basis of laterality of the eye and ocular dominance (P-value paired samples t-test P=0.96 and P=0.19, respectively).

Identification of the dominant eye

The Miles technique30 of identifying ocular dominance was utilized in the study. Briefly, all participants were asked to look at a small target at a distance of 10 ft. Participants were instructed to view the target with both their eyes through a small aperture made by their hands at arm’s length. The eyes were closed one at a time and the participants were asked to report which eye maintained the fixation and was seeing the distant target. The eye that maintained the fixation was deemed as the dominant eye.

MPOD measurements

MPOD measurements were obtained using the QuantifEye by trained observers who were part of the study group. The QuantifEye is a heterochromatic flicker photometer (HFP) that can measure MPOD in vivo. The equipment details about heterochromatic flicker photometry can be found elsewhere.24,25,31 Briefly, the mode of action of heterochromatic flicker photometry is based on the macular pigment’s absorption spectrum at the retina, specifically the macula and fovea.24,25,31 The heterochromatic flicker photometry determines the MPOD by displaying two light stimuli of different wavelengths, which the patient perceives as a flicker.24,25,31 The light stimuli alternate between a blue light of short wavelength, which is maximally absorbed by the macular pigments, and a green-yellow light of longer wavelength, which is not absorbed by the pigments.24,25,31 First, the flickering light is targeted to the fovea; upon perceiving the stimulus the subject will respond that “the target is flickering”. The first five targets determine the threshold. The testing is continued and the machine alters the radiance of blue light versus the green-yellow light until the patient perceives no flicker and is recorded as the lowest point on the graph given in Figure 1. The machine continues to increase the blue radiance until a flicker is perceived three times consecutively. The machine has a built-in correction factor to account for the age-related yellowing of the lens and a final MPOD measurement is obtained. If the individual being examined had an intraocular lens implant, the age of participant is recorded as 21 years where the machines’ correction factor for the yellowing of the lens is zero.
Figure 1

MPOD graphs showing acceptable and unacceptable examples.

Notes: (A–C) Left panels show S-shaped, V-shaped, or U-shaped MPOD output graphs, respectively, that have a lowest point and were considered acceptable. Right panels (D–F) show MPOD graphs that were considered to have data with excessive variability, no downward trend and/or no definite low point and was considered unacceptable.

Abbreviation: MPOD, macular pigment optical density.

All participants had a brief tutorial explaining the procedure of testing and were shown a picture of what target will be visible during testing. The eye undergoing testing first was randomly selected and then the fellow eye was examined. Then the procedure was repeated in both eyes. Participants were given rest when desired between tests. Each eye testing takes ∼2 minutes to perform the procedure. The device outputs include a graph and a value of the MPOD. The graphs (Figure 1) are evaluated for reliability using a set criterion in evaluating the graphs. For a result to be deemed acceptable the graph should have the following criteria: 1) there has to be a downward slope of data points, 2) a distinct lowest point, and 3) three points in upward direction from the lowest point.

Statistical analysis

The difference in number of male and female participants was tested using the binomial test with hypothesized proportions 0.5. The normality of distribution was tested using the Shapiro–Wilk test. The repeatability of the first and second set of MPOD measurements obtained using the right and left eyes and the dominant and nondominant eyes was assessed using the Altman and Bland plots. The paired samples t-test was utilized to evaluate the differences between first and second measurements of MPOD obtained from right eye and left eye and dominant eye and the nondominant eye. The coefficient of repeatability was also calculated. Pearson correlation coefficient was utilized to evaluate the correlation between mean MPOD obtained between the right and left eyes. One-way analysis of variance (ANOVA) was utilized to evaluate the difference in mean MPOD as a function of sex. The statistical analysis was performed using Analyze-it for Microsoft Excel v2.26.

Results

The mean age of the study participants was 35.5 (SD 12.4, range 22–68 years). The study participants consisted of approximately equal number of males (29) and females (43); the difference in the number of males and females was not statistically significant (binomial test proportion 0.40, P=0.12). Table 1 provides the mean MPOD measurements of the study population grouped by laterality, ocular dominance, sex, and coefficient of repeatability. The mean MPOD measurements was not significantly different when participants were grouped by ocular dominance or laterality (paired samples t-test, t statistic =0.42, P=0.68 and t statistic = −1.08, P=0.28, respectively). The mean MPOD measurements in the dominant eye were 0.45 (SD 0.19) and 0.44 (SD 0.14) for males and females, respectively, and not significantly different (one-way ANOVA F=0.05, P=0.8).
Table 1

Values of macular pigment optical density as measured using the QuantifEye Heterochromatic Flicker Photometer

Eye testedMacular pigment optical density
MeanSDRangeLimits of agreementCoefficient of repeatability
Right eye0.430.150.1–0.81−0.11 to +0.110.11
Left eye0.440.170.1–0.91−0.13 to +0.120.12
Dominant eye0.440.170.1–0.91−0.12 to +0.110.12
Nondominant eye0.430.160.1–0.81−0.13 to +0.120.13

Abbreviation: SD, standard deviation.

Assessment of repeatability of MPOD measurements

Figure 1 shows the Altman and Bland plots of the first and second measurements obtained on the right eye and the left eyes, respectively. There is no significant systematic bias between the first and second measurements of the right and left eyes, respectively, with the mean difference between the first and second measurements being close to zero (bias = −0.002 and −0.001, respectively, for the right and left eyes). The difference in the first and second measurements obtained was not significantly different for the data obtained on the right and left eyes (paired samples t-test, t statistic −0.26, P=0.80 and t statistic −0.09, P=0.93 for the right and left eyes, respectively). The limits of agreement were narrow, indicating good repeatability of measurements (Figure 2A and B for values) when evaluating measurements grouped by laterality. The coefficient of repeatability was small and in the range of 0.11 and 0.12 for the dominant and nondominant eyes, respectively.
Figure 2

Repeatability of MPOD measurements.

Notes: (A) and (B) shows Altman and Bland plots with repeat MPOD measurements obtained using the right and left eyes, respectively. The data shows narrow limits of agreement and no significant bias.

Abbreviation: MPOD, macular pigment optical density.

Assessment of intereye correlation of MPOD measurements

Pearson correlation coefficient of mean MPOD values of the right and left eyes shows excellent correlation with r statistic 0.94 (t statistic =22.55, P=<0.0001). The equation allows us to calculate left eye MPOD values from right eye values with an accuracy of 89%. where MPOD is the macular pigment optical density measurement obtained using the QuantifEye device.

Assessment of ocular dominance on MPOD measurements

The repeatability of the MPOD is not influenced if eyes are grouped by ocular dominance. Figure 3A and B shows the Altman and Bland plots of first and second measurements obtained on the dominant and nondominant eyes, respectively. There is no significant systematic bias between the first and second measurements when grouped by ocular dominance with the mean difference between the first and second measurements being close to zero (bias = −0.002 and −0.001, respectively, for the dominant and nondominant eyes, respectively). The difference in the first and second measurements obtained was not significantly different for the data grouped by dominance (paired samples t-test, t statistic −0.26, P=0.80 and t statistic −0.09, P=0.93 for the dominant and nondominant eyes, respectively). Furthermore, the difference in the first and second measurements was not significantly different between the dominant and nondominant eyes (paired t-test, t statistic =0.15, P=0.88).
Figure 3

Repeatability of measures in dominant and nondominant eye.

Notes: (A) and (B) shows Altman and Bland plots with repeat MPOD measurements as grouped by ocular dominance. The data shows narrow limits of agreement and no significant bias.

Abbreviation: MPOD, macular pigment optical density.

The limits of agreement were narrow, indicating good repeatability of measurements (Table 1 for values) when evaluating measurements grouped by dominance. The mean MPOD measured in both the dominant and nondominant eyes was not significantly different in males and females (one-way ANOVA F statistic =0.05, P=0.82 and F statistic =0.01, P=0.94). The mean MPOD measurements for the dominant and nondominant eyes were 0.447 and 0.444, respectively, which were not significantly different (paired samples t-test =0.42, P=0.68).

Discussion

The results of this study show that by utilizing standardized techniques and minimal training, novice individuals can perform testing with the HFP (QuantifEye) and repeatable measurements can be obtained. This study also looked at various factors that could be important to consider when examining patients clinically. This study also establishes protocols for future studies. Prior studies have looked at repeatability of MPOD measurements obtained using the QuantifEye HFP. The literature was not consistent with some studies finding the repeatability not acceptable,25,28 whereas some other finding the repeatability acceptable.26,29 All these studies25,28 have smaller sample sizes ranging from 23 to 40 individuals, whereas the present study had a relatively larger sample size. The possible explanation of the difference in results lies in the quantity and quality of data collected and understanding of the tests by the individuals. Howells et al27 found that removing the suboptimal graphs allowed for improvement in coefficient of repeatability by 0.04 and intersession repeatability by 0.08. The QuantifEye HFP is a psychophysical test and requires some degree of understanding of the tests. Thus, an output of threshold that is graphed also needs to be interpreted by trained observers. In the present study, we found that standardized minimal explanation to individuals performing the MPOD testing and quality check of the MPOD graphs obtained, yields reproducible MPOD results with limits of agreement being narrow and the coefficient of repeatability being 0.11–0.12. The coefficient of repeatability is clinically useful information when augmenting supplements for improving MPOD in individuals at risk of AMD. An increase in MPOD values >0.12 can be considered to be significant. The results of the present study indicate that the measurements obtained in both the right and left eyes of an individual are repeatable. There was no significant difference in measurements obtained on both right and left eyes. This is consistent with prior studies.32–34 Furthermore, the MPOD values obtained from one eye can predict the values of the fellow eye with 89% accuracy. The MPOD values between eyes are not significantly different because the carotenoids, lutein and zeaxanthin, are obtained from oral route and perhaps deposited equally in both eyes. The clinical relevance of the results overall indicates that clinically in ocular healthy individuals it will be sufficient to obtain results in one eye as an indicator of an individual’s MPOD measurements. The present study evaluated the issue of MPOD in ocular dominant eye versus nondominant eye in detail. It was probable that the individuals may be better at performing a psychophysical test with their ocular dominant eye compared with the nondominant eye. It was also possible that the MPOD levels could be greater in the dominant eye and it could in part be responsible for slightly greater visual function and may explain ocular dominance. Although the mean MPOD of the dominant eye was marginally higher than the mean MPOD of the nondominant eye (0.447 vs 0.444, respectively), the difference was not statistically significant and unlikely of any clinical significance. The difference in first set of MPOD measurements and second set of measurements, repeatability, and the mean MPOD values was not related to ocular dominance. It can be safely concluded that the MPOD values can be obtained from either eye, although clinically the patients may feel comfortable using the dominant eye over nondominant eye when performing the test. The known risk factors of the nonexudative AMD include age, smoking, sunlight exposure, race, and low macular pigment.8 Of the known risk factors, one of the modifiable risk factors is the level of MPOD in the retina. It is postulated that increasing the MPOD allows for the absorption of blue light and quenching of the free radicals in the retina, which may harm the retina. The AREDS-210 trial showed that oral supplementation of carotenoids may aid and could be beneficial in prevention of AMD progression to advanced stages. Unfortunately, the trial did not measure MPOD. The baseline level of MPOD and change in MPOD with oral supplementation may aid in clinical decision making and adjusting the dose of oral supplementation to achieve desired effect and not experience unnecessary side effects due to oral supplementation of these carotenoids. Therefore, it is clinically important to accurately measure MPOD. The current study shows that the device QuantifEye can yield repeatable MPOD measurements with low variability. The present study also outlines an examination protocol that can be utilized in future studies that evaluate MPOD.
  34 in total

1.  Macular pigment and risk for age-related macular degeneration in subjects from a Northern European population.

Authors:  S Beatty; I J Murray; D B Henson; D Carden; H Koh; M E Boulton
Journal:  Invest Ophthalmol Vis Sci       Date:  2001-02       Impact factor: 4.799

Review 2.  What does the dominant eye dominate? A brief and somewhat contentious review.

Authors:  Alistair P Mapp; Hiroshi Ono; Raphael Barbeito
Journal:  Percept Psychophys       Date:  2003-02

3.  Macular pigment optical density measurements: evaluation of a device using heterochromatic flicker photometry.

Authors:  R de Kinkelder; R L P van der Veen; F D Verbaak; D J Faber; T G van Leeuwen; T T J M Berendschot
Journal:  Eye (Lond)       Date:  2010-11-05       Impact factor: 3.775

Review 4.  Macular pigment: new clinical methods of detection and the role of carotenoids in age-related macular degeneration.

Authors:  Ivan Y-F Leung
Journal:  Optometry       Date:  2008-05

Review 5.  Measuring macular pigment optical density in vivo: a review of techniques.

Authors:  Olivia Howells; Frank Eperjesi; Hannah Bartlett
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2011-01-08       Impact factor: 3.117

6.  The effect of lutein and zeaxanthin supplementation on metabolites of these carotenoids in the serum of persons aged 60 or older.

Authors:  Frederick Khachik; Fabiana F de Moura; Emily Y Chew; Larry W Douglass; Frederick L Ferris; Jonghyeon Kim; Darby J S Thompson
Journal:  Invest Ophthalmol Vis Sci       Date:  2006-12       Impact factor: 4.799

Review 7.  Age and gender variations in age-related macular degeneration prevalence in populations of European ancestry: a meta-analysis.

Authors:  Alicja R Rudnicka; Zakariya Jarrar; Richard Wormald; Derek G Cook; Astrid Fletcher; Christopher G Owen
Journal:  Ophthalmology       Date:  2011-12-15       Impact factor: 12.079

8.  LAST II: Differential temporal responses of macular pigment optical density in patients with atrophic age-related macular degeneration to dietary supplementation with xanthophylls.

Authors:  Stuart Richer; Jenny Devenport; John C Lang
Journal:  Optometry       Date:  2007-05

9.  Lutein + zeaxanthin and omega-3 fatty acids for age-related macular degeneration: the Age-Related Eye Disease Study 2 (AREDS2) randomized clinical trial.

Authors: 
Journal:  JAMA       Date:  2013-05-15       Impact factor: 56.272

Review 10.  Mechanisms of age-related macular degeneration.

Authors:  Jayakrishna Ambati; Benjamin J Fowler
Journal:  Neuron       Date:  2012-07-12       Impact factor: 17.173

View more
  7 in total

1.  Reply to Green-Gomez et al. Comment on "Richer et al. Night Vision and Carotenoids (NVC): A Randomized Placebo Controlled Clinical Trial on Effects of Carotenoid Supplementation on Night Vision in Older Adults. Nutrients 2021, 13, 3191".

Authors:  Stuart Richer; Steven Novil; Taylor Gullett; Avni Dervishi; Sherwin Nassiri; Co Duong; Robert Davis; Pinakin Gunvant Davey
Journal:  Nutrients       Date:  2022-07-05       Impact factor: 6.706

2.  Correlations between internal and external ocular factors and macular pigment optical density.

Authors:  Ruxandra Tudosescu; Cristina Mihaela Alexandrescu; Sânziana Luminiţa Istrate; Alexandra Diana Vrapciu; Radu Constantin Ciuluvică; Liliana Voinea
Journal:  Rom J Ophthalmol       Date:  2018 Jan-Mar

3.  Differences in macular pigment optical density across four ethnicities: a comparative study.

Authors:  Pinakin Gunvant Davey; Christopher Lievens; Stephanie Ammono-Monney
Journal:  Ther Adv Ophthalmol       Date:  2020-06-15

Review 4.  Carotenoids in the Management of Glaucoma: A Systematic Review of the Evidence.

Authors:  Drake W Lem; Dennis L Gierhart; Pinakin Gunvant Davey
Journal:  Nutrients       Date:  2021-06-06       Impact factor: 5.717

5.  Macular pigment optical density in a Brazilian sample.

Authors:  Letícia Pinto Coelho Jorge; Carlos Eduardo Gonçalves Pereira; Eduardo Jorge; Marcos Pereira de Ávila
Journal:  Int J Retina Vitreous       Date:  2018-01-22

6.  Evaluation of Macular Pigment Optical Density in Healthy Eyes Based on Dual-Wavelength Autofluorescence Imaging in South Indian Population.

Authors:  Ramyaa Srinivasan; Michel M Teussink; Kenneth R Sloan; Janani Surya; Rajiv Raman
Journal:  Transl Vis Sci Technol       Date:  2020-07-28       Impact factor: 3.283

7.  Visual Function and Macular Carotenoid Changes in Eyes with Retinal Drusen-An Open Label Randomized Controlled Trial to Compare a Micronized Lipid-Based Carotenoid Liquid Supplementation and AREDS-2 Formula.

Authors:  Pinakin Gunvant Davey; Thomas Henderson; Drake W Lem; Rebecca Weis; Stephanie Amonoo-Monney; David W Evans
Journal:  Nutrients       Date:  2020-10-26       Impact factor: 5.717

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.