Literature DB >> 20487445

Ex vivo evaluation of the accuracy and coefficient of repeatability of three electronic apex locators using a simple mounting model: a preliminary report.

F L C D'Assunção1, D S Albuquerque, J R Salazar-Silva, V C Dos Santos, J C N Sousa.   

Abstract

AIM: To compare ex vivo the accuracy and coefficient of repeatability of three electronic apex locators in locating the apical constriction.
METHODOLOGY: Thirty-one single-rooted teeth were used. The teeth were sectioned at the cement--enamel junction. A mounting model was used for the measurement of electronic length (EL). The Root ZX-II, the Mini Apex Locator and the Novapex were used for electronic measurements. Each electronic measurement was obtained and repeated. After the last measurement, the file was cemented in place, and the apical 4 mm of each root canal was exposed. The distance from the tip of the file to the apical constriction was determined by three investigators and compared with the corresponding ELs.
RESULTS: The coefficient of repeatability of all devices was acceptable: Root ZX-II, 0.04 mm; Mini Apex Locator, 0.10 mm; and Novapex, 0.08 mm. There was little variation in inter-examiner agreement; the rho(C) (Lin) correlation coefficient was 0.83 for examiners 1 and 2, 0.88 for examiners 1 and 3 and 0.99 for examiners 2 and 3. Using the Root ZX-II, 13 of 31 electronic measurements were located at the apical constriction (42%). Otherwise, the tip of the file was not located at the apical constriction in any of the electronic measurements with the other two devices. The Wilcoxon signed rank test did not reveal any statistical difference between the Root ZX-II measurements and the actual length (P = 0.628), but there was a statistical difference between the Mini Apex Locator and Novapex measurements and the actual length position (P < 0.05).
CONCLUSIONS: The devices tested in this study had a high coefficient of repeatability. The Root ZX-II was accurate, but the Mini Apex Locator and Novapex were not accurate in locating the apical constriction.

Entities:  

Mesh:

Year:  2010        PMID: 20487445     DOI: 10.1111/j.1365-2591.2009.01674.x

Source DB:  PubMed          Journal:  Int Endod J        ISSN: 0143-2885            Impact factor:   5.264


  5 in total

1.  The influence of periapical lesions on the repeatability of two electronic apex locators in vivo.

Authors:  Carlos G Adorno; Sandra M Solaeche; Ileana E Ferreira; Andrea Pedrozo; Patricia M Escobar; Vicente R Fretes
Journal:  Clin Oral Investig       Date:  2021-02-13       Impact factor: 3.573

2.  Accuracy of five electronic foramen locators with different operating systems: an ex vivo study.

Authors:  Bruno Carvalho de Vasconcelos; Michelli de Medeiros Bueno; Suyane Maria Luna-Cruz; Marco Antonio Hungaro Duarte; Carlos Augusto de Oliveira Fernandes
Journal:  J Appl Oral Sci       Date:  2013 Mar-Apr       Impact factor: 2.698

3.  Evaluation of Dentaport ZX and Raypex 6 electronic apex locators: an in vivo study.

Authors:  S Moscoso; K Pineda; J Basilio; C Alvarado; M Roig; F Duran-Sindreu
Journal:  Med Oral Patol Oral Cir Bucal       Date:  2014-03-01

4.  Calcium hydroxide dressing residues after different removal techniques affect the accuracy of Root-ZX apex locator.

Authors:  Emel Uzunoglu; Ayhan Eymirli; Mehmet Özgür Uyanik; Semra Çalt; Emre Nagas
Journal:  Restor Dent Endod       Date:  2014-11-05

5.  An in vitro evaluation of the accuracy of four electronic apex locators using stainless-steel and nickel-titanium hand files.

Authors:  Paras Mull Gehlot; Vinutha Manjunath; Mysore Krishnaswamy Manjunath
Journal:  Restor Dent Endod       Date:  2016-01-04
  5 in total