| Literature DB >> 20471047 |
Huseyin Voyvoda1, Hasan Erdogan.
Abstract
The Optium Xceed is a new hand-held meter for determining blood β-hydroxybutyrate (BHBA) and glucose in human medicine. The objective of this study was to compare BHBA and glucose results obtained using the hand-held meter with those results made with a laboratory method and to evaluate its usefulness as a cowside test in the diagnosis of subclinical ketosis (SCK) in dairy cows. Seventy-eight blood samples from clinically healthy Holstein cows between 5 and 60 days post-calving were analysed. BHBA and glucose values were significantly higher with the hand-held meter versus laboratory methods. Correlation coefficients (r) for BHBA and glucose with the Optium Xceed versus laboratory methods were 0.97 and 0.63, respectively. Based on Bland-Altman plot and Passing-Bablok regression, agreement between two methods was good for BHBA but the agreement for glucose was only fair. When SCK was defined as plasma BHBA levels ≥ 1200 μmol/L, the sensitivity and specificity of the hand-held meter ketone testing in determining SCK were 85% and 94%, respectively. Raising the threshold of the laboratory method to ≥ 1400 μmol/L, the sensitivity and specificity incremented to 0.90 and 0.98, respectively. In conclusion, the blood ketone-monitoring device can be used as a rapid and reliable diagnostic test to detect SCK under field conditions.Entities:
Mesh:
Substances:
Year: 2010 PMID: 20471047 DOI: 10.1016/j.rvsc.2010.04.007
Source DB: PubMed Journal: Res Vet Sci ISSN: 0034-5288 Impact factor: 2.534