Use of a hand-held meter for detecting subclinical ketosis in dairy cows.

Huseyin Voyvoda1, Hasan Erdogan.   

Abstract

The Optium Xceed is a new hand-held meter for determining blood β-hydroxybutyrate (BHBA) and glucose in human medicine. The objective of this study was to compare BHBA and glucose results obtained using the hand-held meter with those results made with a laboratory method and to evaluate its usefulness as a cowside test in the diagnosis of subclinical ketosis (SCK) in dairy cows. Seventy-eight blood samples from clinically healthy Holstein cows between 5 and 60 days post-calving were analysed. BHBA and glucose values were significantly higher with the hand-held meter versus laboratory methods. Correlation coefficients (r) for BHBA and glucose with the Optium Xceed versus laboratory methods were 0.97 and 0.63, respectively. Based on Bland-Altman plot and Passing-Bablok regression, agreement between two methods was good for BHBA but the agreement for glucose was only fair. When SCK was defined as plasma BHBA levels ≥ 1200 μmol/L, the sensitivity and specificity of the hand-held meter ketone testing in determining SCK were 85% and 94%, respectively. Raising the threshold of the laboratory method to ≥ 1400 μmol/L, the sensitivity and specificity incremented to 0.90 and 0.98, respectively. In conclusion, the blood ketone-monitoring device can be used as a rapid and reliable diagnostic test to detect SCK under field conditions.
Copyright © 2010 Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20471047     DOI: 10.1016/j.rvsc.2010.04.007

Source DB:  PubMed          Journal:  Res Vet Sci        ISSN: 0034-5288            Impact factor:   2.534


  7 in total

1.  Microfluidic biosensor for β-Hydroxybutyrate (βHBA) determination of subclinical ketosis diagnosis.

Authors:  Xuan Weng; Wenting Zhao; Suresh Neethirajan; Todd Duffield
Journal:  J Nanobiotechnology       Date:  2015-02-12       Impact factor: 10.435

2.  Evaluation of an Electrochemical Point-of-Care Meter for Measuring Glucose Concentration in Blood from Periparturient Dairy Cattle.

Authors:  A A Megahed; M W H Hiew; J R Townsend; J B Messick; P D Constable
Journal:  J Vet Intern Med       Date:  2015-10-20       Impact factor: 3.333

3.  Glucose concentration in capillary blood of dairy cows obtained by a minimally invasive lancet technique and determined with three different hand-held devices.

Authors:  B Mair; M Drillich; D Klein-Jöbstl; P Kanz; S Borchardt; L Meyer; I Schwendenwein; M Iwersen
Journal:  BMC Vet Res       Date:  2016-02-24       Impact factor: 2.741

4.  Validation of a handheld β-hydroxybutyrate acid meter to identify hyperketonaemia in ewes.

Authors:  Carolina Akiko Sato Cabral Araújo; Antonio Humberto Hamad Minervino; Rejane Santos Sousa; Francisco Leonardo Costa Oliveira; Frederico Augusto Mazzocca Lopes Rodrigues; Clara Satsuki Mori; Enrico Lippi Ortolani
Journal:  PeerJ       Date:  2020-04-10       Impact factor: 2.984

5.  Evaluation of the Freestyle Optium Neo H point-of-care device for measuring blood glucose concentrations in sick calves.

Authors:  Tolga Karapinar; Kenan Cagri Tumer; Sébastien Buczinski
Journal:  J Vet Intern Med       Date:  2020-05-18       Impact factor: 3.333

6.  Evaluation of three therapeutic alternatives for the early treatment of ovine pregnancy toxaemia.

Authors:  L Cal-Pereyra; J R González-Montaña; A Benech; J Acosta-Dibarrat; Mj Martín; S Perini; Mc Abreu; S Da Silva; P Rodríguez
Journal:  Ir Vet J       Date:  2015-10-24       Impact factor: 2.146

7.  The effect of an organic rumen-protected fat supplement on performance, metabolic status, and health of dairy cows.

Authors:  Diego Manriquez; Liang Chen; Pedro Melendez; Pablo Pinedo
Journal:  BMC Vet Res       Date:  2019-12-11       Impact factor: 2.741

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.