Computed tomography as a source of electron density information for radiation treatment planning.

Witold Skrzyński1, Sylwia Zielińska-Dabrowska, Marta Wachowicz, Wioletta Slusarczyk-Kacprzyk, Paweł F Kukołowicz, Wojciech Bulski.   

Abstract

PURPOSE: To evaluate the performance of computed tomography (CT) systems of various designs as a source of electron density (rho(el)) data for treatment planning of radiation therapy.
MATERIAL AND METHODS: Dependence of CT numbers on relative electron density of tissue-equivalent materials (HU-rho(el) relationship) was measured for several general-purpose CT systems (single-slice, multislice, wide-bore multislice), for radiotherapy simulators with a single-slice CT and kV CBCT (cone-beam CT) options, as well as for linear accelerators with kV and MV CBCT systems. Electron density phantoms of four sizes were used. Measurement data were compared with the standard HU-rhoel relationships predefined in two commercial treatment-planning systems (TPS).
RESULTS: The HU-rho(el) relationships obtained with all of the general-purpose CT scanners operating at voltages close to 120 kV were very similar to each other and close to those predefined in TPS. Some dependency of HU values on tube voltage was observed for bone- equivalent materials. For a given tube voltage, differences in results obtained for different phantoms were larger than those obtained for different CT scanners. For radiotherapy simulators and for kV CBCT systems, the information on rhoel was much less precise because of poor uniformity of images. For MV CBCT, the results were significantly different than for kV systems due to the differing energy spectrum of the beam.
CONCLUSION: The HU-rho(el) relationships predefined in TPS can be used for general-purpose CT systems operating at voltages close to 120 kV. For nontypical imaging systems (e.g., CBCT), the relationship can be significantly different and, therefore, it should always be measured and carefully analyzed before using CT data for treatment planning.

Mesh:

Year:  2010        PMID: 20458451     DOI: 10.1007/s00066-010-2086-5

Source DB:  PubMed          Journal:  Strahlenther Onkol        ISSN: 0179-7158            Impact factor:   3.621


  20 in total

1.  Relative electron density calibration of CT scanners for radiotherapy treatment planning.

Authors:  S J Thomas
Journal:  Br J Radiol       Date:  1999-08       Impact factor: 3.039

2.  Cone-beam computed tomography with a flat-panel imager: magnitude and effects of x-ray scatter.

Authors:  J H Siewerdsen; D A Jaffray
Journal:  Med Phys       Date:  2001-02       Impact factor: 4.071

3.  Physical performance and image optimization of megavoltage cone-beam CT.

Authors:  Olivier Morin; Jean-François Aubry; Michèle Aubin; Josephine Chen; Martina Descovich; Ali-Bani Hashemi; Jean Pouliot
Journal:  Med Phys       Date:  2009-04       Impact factor: 4.071

4.  The adaptation of megavoltage cone beam CT for use in standard radiotherapy treatment planning.

Authors:  T Hannah Mary Thomas; D Devakumar; S Purnima; B Paul Ravindran
Journal:  Phys Med Biol       Date:  2009-03-13       Impact factor: 3.609

5.  The direct use of CT numbers in radiotherapy dosage calculations for inhomogeneous media.

Authors:  R P Parker; P A Hobday; K J Cassell
Journal:  Phys Med Biol       Date:  1979-07       Impact factor: 3.609

6.  Assessing the effect of electron density in photon dose calculations.

Authors:  J Seco; P M Evans
Journal:  Med Phys       Date:  2006-02       Impact factor: 4.071

7.  Radiotherapy of liver metastases. Comparison of target volumes and dose-volume histograms employing CT- or MRI-based treatment planning.

Authors:  Maciej Pech; Konrad Mohnike; Gero Wieners; Ewa Bialek; Oliver Dudeck; Max Seidensticker; Nils Peters; Peter Wust; Günther Gademann; Jens Ricke
Journal:  Strahlenther Onkol       Date:  2008-05       Impact factor: 3.621

8.  Helical tomotherapy. Experiences of the first 150 patients in Heidelberg.

Authors:  Florian Sterzing; Kai Schubert; Gabriele Sroka-Perez; Jörn Kalz; Jürgen Debus; Klaus Herfarth
Journal:  Strahlenther Onkol       Date:  2008-01       Impact factor: 3.621

9.  The use of molecular imaging to evaluate radiation fields in the adjuvant setting of breast cancer: a feasibility study.

Authors:  Samuel Bral; Vincent Vinh-Hung; Hendrik Everaert; Peter De Coninck; Guy Storme
Journal:  Strahlenther Onkol       Date:  2008-02       Impact factor: 3.621

10.  Effect of contrast enhanced CT scans on heterogeneity corrected dose computations in the lung.

Authors:  Nichola A Burridge; Carl G Rowbottom; Paul A Burt
Journal:  J Appl Clin Med Phys       Date:  2006-11-28       Impact factor: 2.102

View more
  15 in total

1.  Phase-specific cone beam computed tomography reduces reconstructed volume loss of moving phantom.

Authors:  H-L Chao; W-L Chen; C-C Hu; J-K Wu; C-J Wu; J C-H Cheng
Journal:  Strahlenther Onkol       Date:  2011-12-24       Impact factor: 3.621

2.  RapidArc combined with the active breathing coordinator provides an effective and accurate approach for the radiotherapy of hepatocellular carcinoma.

Authors:  G Z Gong; Y Yin; L G Xing; Y J Guo; T Liu; J Chen; J Lu; C Ma; T Sun; T Bai; G Zhang; R Wang
Journal:  Strahlenther Onkol       Date:  2012-02-08       Impact factor: 3.621

3.  Hounsfield units variations: impact on CT-density based conversion tables and their effects on dose distribution.

Authors:  B Zurl; R Tiefling; P Winkler; P Kindl; K S Kapp
Journal:  Strahlenther Onkol       Date:  2013-11-09       Impact factor: 3.621

4.  Early changes of parotid density and volume predict modifications at the end of therapy and intensity of acute xerostomia.

Authors:  Maria Luisa Belli; Elisa Scalco; Giuseppe Sanguineti; Claudio Fiorino; Sara Broggi; Nicola Dinapoli; Francesco Ricchetti; Vincenzo Valentini; Giovanna Rizzo; Giovanni Mauro Cattaneo
Journal:  Strahlenther Onkol       Date:  2014-04-23       Impact factor: 3.621

5.  Can CT scan protocols used for radiotherapy treatment planning be adjusted to optimize image quality and patient dose? A systematic review.

Authors:  Anne T Davis; Antony L Palmer; Andrew Nisbet
Journal:  Br J Radiol       Date:  2017-05-23       Impact factor: 3.039

6.  Practically acquired and modified cone-beam computed tomography images for accurate dose calculation in head and neck cancer.

Authors:  Chih-Chung Hu; Wen-Tao Huang; Chiao-Ling Tsai; Jian-Kuen Wu; Hsiao-Ling Chao; Guo-Ming Huang; Chun-Wei Wang; Chien-Jang Wu; Jason Chia-Hsien Cheng
Journal:  Strahlenther Onkol       Date:  2011-09-23       Impact factor: 3.621

7.  Pretreatment verification of dose calculation and delivery by means of measurements with PLEXITOM™ phantom.

Authors:  Paweł Wołowiec; Paweł Kukołowicz; Krzysztof Lis
Journal:  Rep Pract Oncol Radiother       Date:  2013-02-04

8.  Procedure for creating a three-dimensional (3D) model for superficial hyperthermia treatment planning.

Authors:  Marianne Linthorst; Tomas Drizdal; Hans Joosten; Gerard C van Rhoon; Jacoba van der Zee
Journal:  Strahlenther Onkol       Date:  2011-11-25       Impact factor: 3.621

9.  CT-myelography for high-dose irradiation of spinal and paraspinal tumors with helical tomotherapy: revival of an old tool.

Authors:  Matthias Uhl; Florian Sterzing; Gregor Habl; Kai Schubert; Gabriele Sroka-Perez; Jürgen Debus; Klaus Herfarth
Journal:  Strahlenther Onkol       Date:  2011-06-27       Impact factor: 3.621

10.  MRI-Based Proton Treatment Planning for Base of Skull Tumors.

Authors:  Ghazal Shafai-Erfani; Yang Lei; Yingzi Liu; Yinan Wang; Tonghe Wang; Jim Zhong; Tian Liu; Mark McDonald; Walter J Curran; Jun Zhou; Hui-Kuo Shu; Xiaofeng Yang
Journal:  Int J Part Ther       Date:  2019-09-30
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.