Literature DB >> 20454914

Basic study of a transcutaneous information transmission system using intra-body communication.

Eiji Okamoto1, Yusuke Sato, Kazuyuki Seino, Takashi Kiyono, Yoshikuni Kato, Yoshinori Mitamura.   

Abstract

The transcutaneous communication system (TCS) is one of the key technologies for monitoring and controling artificial hearts and other artificial organs in the body. In this study, we have developed a new TCS that uses the human body as a conductive medium. Having no energy conversion from electric currents into electromagnetic waves and light provides energy-saving data transmission with a simple electrical circuit. Each unit of the TCS mainly consists of two electrodes, an amplitude shift keying (ASK) modulator and an ASK demodulator (carrier frequency: 4 and 10 MHz). A resonant frequency of an L-C tank circuit including the capacitance component of the body is tuned into each carrier frequency in order to apply the data current effectively into the body. Performance of the TCS was evaluated by a communication test on the surface of a human body. The TCS was able to transmit 3,315 bytes of data bi-directionally at a transmission rate of 115 kbps from a left wrist to a right forearm, to an abdomen and to a left calf without communication error. The power consumption of each TCS unit was 125 mW with an ASK modulated current of 7 mA (RMS). While further study is required to secure its safety, the TCS promises to be a next-generation transcutaneous communication device.

Entities:  

Mesh:

Year:  2010        PMID: 20454914     DOI: 10.1007/s10047-010-0502-3

Source DB:  PubMed          Journal:  J Artif Organs        ISSN: 1434-7229            Impact factor:   1.731


  4 in total

1.  Development of a bidirectional transcutaneous optical data transmission system for artificial hearts allowing long-distance data communication with low electric power consumption.

Authors:  Eiji Okamoto; Yoshiro Yamamoto; Yusuke Inoue; Tsutomu Makino; Yoshinori Mitamura
Journal:  J Artif Organs       Date:  2005       Impact factor: 1.731

2.  An attempt to model the human body as a communication channel.

Authors:  Marc Simon Wegmueller; Andreas Kuhn; Juerg Froehlich; Michael Oberle; Norbert Felber; Niels Kuster; Wolfgang Fichtner
Journal:  IEEE Trans Biomed Eng       Date:  2007-10       Impact factor: 4.538

3.  Transcutaneous optical telemetry system with infrared laser diode.

Authors:  K Inoue; K Shiba; E Shu; K Koshiji; K Tsukahara; T Oh-umi; T Masuzawa; E Tatsumi; Y Taenaka; H Takano
Journal:  ASAIO J       Date:  1998 Nov-Dec       Impact factor: 2.872

4.  Remote monitoring of patients with biventricular defibrillators through the CareLink system improves clinical management of arrhythmias and heart failure episodes.

Authors:  Massimo Santini; Renato P Ricci; Maurizio Lunati; Maurizio Landolina; Giovanni B Perego; Maurizio Marzegalli; Milena Schirru; Chiara Belvito; Roberto Brambilla; Giuseppe Guenzati; Serena Gilardi; Sergio Valsecchi
Journal:  J Interv Card Electrophysiol       Date:  2008-10-31       Impact factor: 1.900

  4 in total
  5 in total

Review 1.  Journal of Artificial Organs 2010: the year in review.

Authors: 
Journal:  J Artif Organs       Date:  2011-03-16       Impact factor: 1.731

2.  Histological investigation of the titanium fiber mesh with one side sealed with non-porous material for its application to the artificial heart system.

Authors:  Eiji Okamoto; Kyoko Arimura; Yoshinori Mitamura
Journal:  J Artif Organs       Date:  2018-08-20       Impact factor: 1.731

3.  Electrical characteristic of the titanium mesh electrode for transcutaneous intrabody communication to monitor implantable artificial organs.

Authors:  Eiji Okamoto; Sakiko Kikuchi; Yoshinori Mitamura
Journal:  J Artif Organs       Date:  2016-02-26       Impact factor: 1.731

4.  Interface of data transmission for a transcutaneous communication system using the human body as transmission medium.

Authors:  Eiji Okamoto; Yoshikuni Kato; Kazuyuki Seino; Yoshinori Mitamura
Journal:  J Artif Organs       Date:  2011-08-21       Impact factor: 1.731

5.  Study of channel characteristics for galvanic-type intra-body communication based on a transfer function from a quasi-static field model.

Authors:  Xi Mei Chen; Peng Un Mak; Sio Hang Pun; Yue Ming Gao; Chan-Tong Lam; Mang I Vai; Min Du
Journal:  Sensors (Basel)       Date:  2012-11-27       Impact factor: 3.576

  5 in total