Comparison of patient megavoltage cone beam CT images acquired with an unflattened beam from a carbon target and a flattened treatment beam.

Bruce A Faddegon1, Michele Aubin, Ali Bani-Hashemi, Bijumon Gangadharan, Alexander R Gottschalk, Olivier Morin, Daren Sawkey, Vincent Wu, Sue S Yom.   

Abstract

PURPOSE: To use an imaging beam line (IBL) to obtain the first megavoltage cone-beam computed tomography (MV CBCT) images of patients with a low atomic number (Z) target, and to compare these images to those taken of the same patients with the 6 MV flattened beam from the treatment beam line (TBL).
METHODS: The IBL, which produces a 4.2 MV unflattened beam from a carbon target, was installed on a linear accelerator in use for radiotherapy. Provision was made for switching between the IBL and TBL for imaging the same patient with beams from the low-Z and high-Z targets. Dose was quoted as monitor units times the dose per monitor unit for the standard calibration geometry. Images were acquired with institutional approval and patient consent with both the IBL and TBL on a series of 23 patients undergoing radiotherapy. Patients were imaged daily to weekly and aligned to the planning CT using the images. Doses were reduced over the course of treatment to determine the minimum doses required for alignment. Images were assessed offline.
RESULTS: IBL MV CBCT images of prostate, head and neck, lung, and abdomen showed improvement in soft tissue contrast for the same dose as the TBL images. Bony anatomy, air cavities, and fiducial markers were sharper. CBCT with a dose of 1 cGy was sufficient for alignment of prostate and head and neck patients based on bony anatomy or implanted gold seeds, 2-4 cGy for lung, abdomen, and pelvis. Photon scatter in the patient had minimal effect on image quality. The metallic hip prosthesis in one patient showed reduced artifacts compared to diagnostic CT.
CONCLUSIONS: The IBL has the advantage of improved image quality at the same dose, or reduced dose for the same image quality, over the TBL.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20443494     DOI: 10.1118/1.3359822

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  8 in total

1.  Super-resolution imaging in a multiple layer EPID.

Authors:  Haijian Chen; Joerg Rottmann; Stephen Sf Yip; Daniel Morf; Rony Füglistaller; Josh Star-Lack; George Zentai; Ross Berbeco
Journal:  Biomed Phys Eng Express       Date:  2017-02-21

2.  Dosimetric properties and commissioning of cone-beam CT image beam line with a carbon target.

Authors:  Y Dzierma; F G Nuesken; N P Licht; C Ruebe
Journal:  Strahlenther Onkol       Date:  2013-05-30       Impact factor: 3.621

3.  [Organizational models of inpatient emergency and acute care].

Authors:  A Kaltwasser; C Dodt
Journal:  Med Klin Intensivmed Notfmed       Date:  2014-10       Impact factor: 0.840

4.  Image quality and dose distributions of three linac-based imaging modalities.

Authors:  Yvonne Dzierma; Evemarie Ames; Frank Nuesken; Jan Palm; Norbert Licht; Christian Rübe
Journal:  Strahlenther Onkol       Date:  2014-12-20       Impact factor: 3.621

5.  Prostate image-guided radiotherapy by megavolt cone-beam CT.

Authors:  Sergio Zucca; Barbara Carau; Ignazio Solla; Elisabetta Garibaldi; Paolo Farace; Giancarlo Lay; Gianfranco Meleddu; Pietro Gabriele
Journal:  Strahlenther Onkol       Date:  2011-07-22       Impact factor: 3.621

6.  Feasibility of using respiratory correlated mega voltage cone beam computed tomography to measure tumor motion.

Authors:  Mingqing Chen; R Alfredo Siochi
Journal:  J Appl Clin Med Phys       Date:  2011-01-31       Impact factor: 2.102

7.  Imaging dose and secondary cancer risk in image-guided radiotherapy of pediatric patients.

Authors:  Yvonne Dzierma; Katharina Mikulla; Patrick Richter; Katharina Bell; Patrick Melchior; Frank Nuesken; Christian Rübe
Journal:  Radiat Oncol       Date:  2018-09-05       Impact factor: 3.481

8.  Positioning errors of metal localization devices with motion artifacts on kV and MV cone beam CT.

Authors:  Teh Lin; Chang-Ming Charlie Ma
Journal:  BJR Open       Date:  2019-03-15
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.