Literature DB >> 20422186

Laparoscopic evaluation shows deficiencies in memory ring deployment during small ventral hernia repair.

Frederik Berrevoet1, Bert Van den Bossche, Luc de Baerdemaeker, Bernard de Hemptinne.   

Abstract

BACKGROUND: With the introduction of a self-expanding, memory-containing, circular hernia patch, surgeons have been enthusiastic about its use to repair ventral hernias smaller than 3 cm in diameter. The aim of this study was to evaluate the efficiency, reliability, and safety of the device laparoscopically with respect to adequate deployment of the patch.
METHODS: During 1 year all patients with small ventral hernias were treated with this memory-containing patch and were inspected by laparoscopy. Just prior to insertion of the patch, remaining adhesions on top of the peritoneum were analyzed, as was the interference of the umbilical ligament. The final position of the patch was monitored, identifying the cupping phenomenon, exposure of the polypropylene to the viscera, and the amount of tension on the straps. All patients were followed for 2 years and postoperative complications and recurrence rate were monitored.
RESULTS: Twenty-eight patients were operated on for repair of a small ventral hernia with laparoscopic control. Adhesions, not digitally palpable, that interfered with adequate patch deployment were observed in more than 80% of the cases. After a median follow-up of 25 months a 14.8% recurrence rate was observed.
CONCLUSIONS: The patch, consisting of both polypropylene and ePTFE, leads to unacceptable morbidity and a high rate of recurrences. By laparoscopic evaluation, these recurrences are probably based on a combination of material characteristics and unavoidable technical errors.

Entities:  

Mesh:

Year:  2010        PMID: 20422186     DOI: 10.1007/s00268-010-0600-7

Source DB:  PubMed          Journal:  World J Surg        ISSN: 0364-2313            Impact factor:   3.352


  14 in total

1.  A comparison of suture repair with mesh repair for incisional hernia.

Authors:  R W Luijendijk; W C Hop; M P van den Tol; D C de Lange; M M Braaksma; J N IJzermans; R U Boelhouwer; B C de Vries; M K Salu; J C Wereldsma; C M Bruijninckx; J Jeekel
Journal:  N Engl J Med       Date:  2000-08-10       Impact factor: 91.245

2.  Which repair in umbilical hernia of adults: primary or mesh?

Authors:  Ramazan Eryilmaz; Mustafa Sahin; M Hakan Tekelioglu
Journal:  Int Surg       Date:  2006 Sep-Oct

3.  Randomized clinical trial comparing suture and mesh repair of umbilical hernia in adults.

Authors:  A Arroyo; P García; F Pérez; J Andreu; F Candela; R Calpena
Journal:  Br J Surg       Date:  2001-10       Impact factor: 6.939

4.  Surgeon perspectives on options for ventral abdominal wall hernia repair: results of a postal questionnaire.

Authors:  P Witherspoon; P J O'Dwyer
Journal:  Hernia       Date:  2005-10-22       Impact factor: 4.739

5.  Retrospective comparison of mesh and sutured repair for adult umbilical hernias.

Authors:  P Sanjay; T D Reid; E L Davies; P J Arumugam; A Woodward
Journal:  Hernia       Date:  2005-05-13       Impact factor: 4.739

6.  Ventral hernia repair: a study of current practice.

Authors:  C A Courtney; A C Lee; C Wilson; P J O'Dwyer
Journal:  Hernia       Date:  2003-01-24       Impact factor: 4.739

7.  Long-term follow-up of a randomized controlled trial of suture versus mesh repair of incisional hernia.

Authors:  Jacobus W A Burger; Roland W Luijendijk; Wim C J Hop; Jens A Halm; Emiel G G Verdaasdonk; Johannes Jeekel
Journal:  Ann Surg       Date:  2004-10       Impact factor: 12.969

8.  Technique and outcomes of abdominal incisional hernia repair using a synthetic composite mesh: a report of 455 cases.

Authors:  David A Iannitti; William W Hope; H James Norton; Amy E Lincourt; Keith Millikan; Michael E Fenoglio; Mark Moskowitz
Journal:  J Am Coll Surg       Date:  2007-10-18       Impact factor: 6.113

Review 9.  Laparoscopic incisional and ventral hernia repair: complications-how to avoid and handle.

Authors:  K A LeBlanc
Journal:  Hernia       Date:  2004-12       Impact factor: 4.739

10.  Laparoscopic versus open umbilical hernia repair.

Authors:  Rodrigo Gonzalez; Edward Mason; Titus Duncan; Russell Wilson; Bruce J Ramshaw
Journal:  JSLS       Date:  2003 Oct-Dec       Impact factor: 2.172

View more
  13 in total

1.  Lessons and challenges during a 5-year follow-up of 21 Composix Kugel implantations.

Authors:  A Wiegering; N Schlegel; C Isbert; C Jurowich; S Doht; C T Germer; U A Dietz
Journal:  Hernia       Date:  2013-05-09       Impact factor: 4.739

2.  Re: Laparoscopic evaluation shows deficiencies in memory ring deployment during small ventral hernia repair.

Authors:  Christophe R Berney
Journal:  World J Surg       Date:  2012-05       Impact factor: 3.352

3.  Umbilical hernia repair with composite prosthesis: a single-centre experience.

Authors:  J L Porrero; O Cano-Valderrama; S Villar; C Sánchez-Cabezudo; B Ramos; B Porrero; M Cendrero; O Bonachia; A J Torres
Journal:  Hernia       Date:  2018-11-02       Impact factor: 4.739

4.  Infected large pore meshes may be salvaged by topical negative pressure therapy.

Authors:  F Berrevoet; A Vanlander; M Sainz-Barriga; X Rogiers; R Troisi
Journal:  Hernia       Date:  2012-07-27       Impact factor: 4.739

5.  Retrospective analysis of umbilical, epigastric, and small incisional hernia repair using the Ventralex™ hernia patch.

Authors:  T Tollens; M Den Hondt; K Devroe; C Terry; S Speybroeck; C Aelvoet; J-P Vanrykel
Journal:  Hernia       Date:  2011-04-13       Impact factor: 4.739

Review 6.  Proper mesh overlap is a key determinant in hernia recurrence following laparoscopic ventral and incisional hernia repair.

Authors:  K LeBlanc
Journal:  Hernia       Date:  2015-07-05       Impact factor: 4.739

7.  Single centre observational study to evaluate the safety and efficacy of the Proceed™ Ventral Patch to repair small ventral hernias.

Authors:  J Bontinck; I Kyle-Leinhase; P Pletinckx; V Vergucht; R Beckers; F Muysoms
Journal:  Hernia       Date:  2013-07-24       Impact factor: 4.739

8.  Intraperitoneal mesh devices for small midline hernias: mesh behavior in a porcine model.

Authors:  E Reynvoet; K Chiers; I Van Overbeke; R Troisi; F Berrevoet
Journal:  Hernia       Date:  2015-03-20       Impact factor: 4.739

9.  Abdominal wall hernia repair with a composite ePTFE/polypropylene mesh: clinical outcome and quality of life in 152 patients.

Authors:  E Iversen; A Lykke; M Hensler; L N Jorgensen
Journal:  Hernia       Date:  2010-09-21       Impact factor: 4.739

10.  Intraperitoneal behaviour of a new composite mesh (Parietex™ Composite Ventral Patch) designed for umbilical or epigastric hernia repair.

Authors:  F García-Moreno; S Sotomayor; P Pérez-López; B Pérez-Köhler; Y Bayon; G Pascual; J M Bellón
Journal:  Surg Endosc       Date:  2014-06-27       Impact factor: 4.584

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.