Transcatheter aortic valves inadequately relieve stenosis in small degenerated bioprostheses.

Ali N Azadani1, Nicolas Jaussaud, Peter B Matthews, Liang Ge, Timothy A M Chuter, Elaine E Tseng.   

Abstract

OBJECTIVES: Transcatheter aortic valves (TAVs) are a promising treatment for high risk surgical patients suffering from degeneration of previously implanted bioprostheses (valve-in-valve therapy). However, unlike native stenosed aortic valves which have accommodated Edwards SAPIEN transcatheter valves after valvuloplasty, rigid bioprostheses may prevent full TAV stent expansion and disrupt leaflet function. We hypothesized that current 23 mm TAVs would not completely relieve severe stenosis in small bioprosthetic valves. The objective of this study was to study the hemodynamics of TAVs in degenerated bioprostheses.
METHODS: Twelve TAVs designed to mimic the 23 mm SAPIEN valve were created. Using a pulse duplicator, hemodynamics of valve-in-valve implantation were measured within 19, 21, and 23 mm Carpentier-Edwards PERIMOUNT degenerated bioprostheses (n=6 each). Bioprosthetic degeneration was simulated using BioGlue to achieve a mean pressure gradient of 50 mmHg.
RESULTS: TAVs significantly reduced the mean pressure gradient (50.9+/-4.7-9.1+/-4.1 mmHg, P<0.001) and total energy loss (870.3+/-157.4-307.8+/-87.3 mJ, P<0.001) in 23 mm degenerated bioprostheses. In 21 mm bioprostheses, the pressure gradient (52.3+/-7.0-19.5+/-5.0 mmHg, P<0.001) and energy loss (785.5+/-128.1-477.8+/-123.2 mJ, P=0.007) were reduced significantly. However, no significant changes in the pressure gradient (57.1+/-4.3-46.5+/-9.3 mmHg, P=0.086) or energy loss (839.3+/-49.3-960.5+/-158.1 mJ, P=0.144) were obtained after TAVI implantation in 19 mm bioprostheses. Incomplete stent expansion resulted in leaflet distortion and central regurgitation when implanted in 19 and 21 mm bioprostheses.
CONCLUSIONS: The bioprosthetic annulus and stent posts offered a suitable landing zone for TAVs. However, oversized transcatheter valves were constrained by the rigid bioprostheses resulting in inadequate resolution of bioprosthetic stenosis. Hemodynamics of valve-in-valve intervention was worse than comparable size surgical valve replacements, particularly in 19 and 21 mm valves. Small degenerated bioprostheses require modification of current TAV design to yield acceptable hemodynamics.

Entities:  

Mesh:

Year:  2010        PMID: 20395249     DOI: 10.1510/icvts.2009.225144

Source DB:  PubMed          Journal:  Interact Cardiovasc Thorac Surg        ISSN: 1569-9285


  9 in total

1.  Transcatheter Edwards Sapien XT valve in valve implantation in degenerated aortic bioprostheses via transfemoral access.

Authors:  Martin Greif; Philipp Lange; Helmut Mair; Christoph Becker; Christoph Schmitz; Gerhard Steinbeck; Christian Kupatt
Journal:  Clin Res Cardiol       Date:  2012-06-23       Impact factor: 5.460

2.  Patient-specific simulations of transcatheter aortic valve stent implantation.

Authors:  C Capelli; G M Bosi; E Cerri; J Nordmeyer; T Odenwald; P Bonhoeffer; F Migliavacca; A M Taylor; S Schievano
Journal:  Med Biol Eng Comput       Date:  2012-02       Impact factor: 2.602

Review 3.  Valve-in-valve implantations: is this the new standard for degenerated bioprostheses? Review of the literature.

Authors:  Krys Milburn; Vinayak Bapat; Martyn Thomas
Journal:  Clin Res Cardiol       Date:  2014-01-21       Impact factor: 5.460

Review 4.  On the Mechanics of Transcatheter Aortic Valve Replacement.

Authors:  Lakshmi P Dasi; Hoda Hatoum; Arash Kheradvar; Ramin Zareian; S Hamed Alavi; Wei Sun; Caitlin Martin; Thuy Pham; Qian Wang; Prem A Midha; Vrishank Raghav; Ajit P Yoganathan
Journal:  Ann Biomed Eng       Date:  2016-11-21       Impact factor: 3.934

5.  Optimising the Haemodynamics of Aortic Valve-in-valve Procedures.

Authors:  Ren Jie Yao; Matheus Simonato; Danny Dvir
Journal:  Interv Cardiol       Date:  2017-05

6.  Bioprosthetic Valve Fracture During Valve-in-valve TAVR: Bench to Bedside.

Authors:  John T Saxon; Keith B Allen; David J Cohen; Adnan K Chhatriwalla
Journal:  Interv Cardiol       Date:  2018-01

7.  Recurrent severe aortic stenosis one year after transcatheter aortic valve-in-valve implantation: Successful treatment with balloon aortic valvuloplasty.

Authors:  Abbasali Badami; Entela B Lushaj; Kurt Jacobson; Amish Raval; Lucian Lozonschi; Takushi Kohmoto
Journal:  J Cardiol Cases       Date:  2016-04-16

Review 8.  Transcatheter Aortic Valve-in-Valve Procedure in Patients with Bioprosthetic Structural Valve Deterioration.

Authors:  Ross M Reul; Mahesh K Ramchandani; Michael J Reardon
Journal:  Methodist Debakey Cardiovasc J       Date:  2017 Jul-Sep

9.  Transcatheter heart valve in valve implantation with Edwards SAPIEN bioprosthetic valve for different degenerated bioprosthetic valve positions (First Iranian ViV report with mid-term follow up).

Authors:  Ali Mohammad Haji Zeinali; Kyomars Abbasi; Mohammad Saheb Jam; Shahrooz Yazdani; Seyedeh Hamideh Mortazavi
Journal:  J Cardiovasc Thorac Res       Date:  2017-09-30
  9 in total