Cochlear coordinates in regard to cochlear implantation: a clinically individually applicable 3 dimensional CT-based method.

Berit M Verbist1, Raoul M S Joemai, Jeroen J Briaire, Wouter M Teeuwisse, Wouter J H Veldkamp, Johan H M Frijns.   

Abstract

SETTING: Cochlear implant (CI)/tertiary referral center.
SUBJECTS: Twenty-five patients implanted with an Advanced Bionics HiRes90K HiFocus1J CI. STUDY DESIGN/MAIN OUTCOME MEASURES: A 3-dimensional cylindrical coordinate system is introduced using the basal turn of the cochlea as the x and y planes and the center of the modiolus as the z axis. The 0-degree angle is defined by the most lateral point of the horizontal semicircular canal. It is applied to both preoperative and postoperative computed tomographies in 25 patients. The angular position of the round window is examined. Interobserver reproducibility is tested by localization of all electrode contacts within the coordinate system. To observe realignment over time, electrode coordinates in postoperative images were projected on preoperative images. Additionally, comparison to existing imaging-related coordinate systems was made.
RESULTS: The angular position of the center of the round window is 34.6 +/- 0.4 degrees (standard deviation) with an intraclass coefficient of 1.00. The intraclass coefficient for interobserver reproducibility of the 16 electrode contacts ranged from 0.74 to 1 for the rotational angle (phi) and 0.77 to 1 for the distance to the modiolus (rho). In 21 of 25 patients, a perfect match or minimal displacement of up to 3 electrode contacts was seen. Comparison to existing systems showed good correlation.
CONCLUSION: A 3-dimensional cochlear coordinate system easily applicable in clinical patients is described, which fulfills the requirements set by an international consensus.

Entities:  

Mesh:

Year:  2010        PMID: 20393379     DOI: 10.1097/MAO.0b013e3181d8d3c4

Source DB:  PubMed          Journal:  Otol Neurotol        ISSN: 1531-7129            Impact factor:   2.311


  9 in total

1.  Correlation between histologic and radiographic reconstruction of intracochlear electrode position in human temporal bones.

Authors:  Jennifer T O'Malley; Barbara J Burgess; Meng-Yu Zhu; Hugh D Curtin; Joseph B Nadol
Journal:  Audiol Neurootol       Date:  2014-03-27       Impact factor: 1.854

2.  Standardization of CT depiction of cochlear implant insertion depth.

Authors:  C C Colby; N W Todd; H R Harnsberger; P A Hudgins
Journal:  AJNR Am J Neuroradiol       Date:  2014-10-22       Impact factor: 3.825

3.  Verification of computed tomographic estimates of cochlear implant array position: a micro-CT and histologic analysis.

Authors:  Jessica Teymouri; Timothy E Hullar; Timothy A Holden; Richard A Chole
Journal:  Otol Neurotol       Date:  2011-08       Impact factor: 2.311

4.  Comparison of cochlear implant relevant anatomy in children versus adults.

Authors:  Theodore R McRackan; Fitsum A Reda; Alejandro Rivas; Jack H Noble; Mary S Dietrich; Benoit M Dawant; Robert F Labadie
Journal:  Otol Neurotol       Date:  2012-04       Impact factor: 2.311

5.  Radiological Evaluation of Cochlear Orientation and Its Implications in Cochlear Implantation.

Authors:  Avvaru Satya Kiran; Ajoy Mathew Varghese; Aparna Irodi; Anjali Lepcha; John Mathew; Visalakshi Jeyaseelan
Journal:  Indian J Otolaryngol Head Neck Surg       Date:  2017-07-24

6.  Quality-assured training in the evaluation of cochlear implant electrode position: a prospective experimental study.

Authors:  Alexander Mewes; Sebastian Burg; Goetz Brademann; Jan Andreas Dambon; Matthias Hey
Journal:  BMC Med Educ       Date:  2022-05-20       Impact factor: 3.263

7.  Pitch comparisons between electrical stimulation of a cochlear implant and acoustic stimuli presented to a normal-hearing contralateral ear.

Authors:  Robert P Carlyon; Olivier Macherey; Johan H M Frijns; Patrick R Axon; Randy K Kalkman; Patrick Boyle; David M Baguley; John Briggs; John M Deeks; Jeroen J Briaire; Xavier Barreau; René Dauman
Journal:  J Assoc Res Otolaryngol       Date:  2010-06-05

8.  Hydraulic insertions of cochlear implant electrode arrays into the human cadaver cochlea: preliminary findings.

Authors:  M Geraldine Zuniga; Thomas Lenarz; Thomas S Rau
Journal:  Eur Arch Otorhinolaryngol       Date:  2021-08-14       Impact factor: 2.503

9.  New hybrid multiplanar cone beam computed tomography-laser-fluoroscopic-guided approach in cochlear implant surgery.

Authors:  Stephan Waldeck; Sandra Schmidt; Christian von Falck; René Chapot; Marc Brockmann; Daniel Overhoff
Journal:  Int J Comput Assist Radiol Surg       Date:  2022-07-11       Impact factor: 3.421

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.