Markerless lung tumor tracking and trajectory reconstruction using rotational cone-beam projections: a feasibility study.

John H Lewis1, Ruijiang Li, W Tyler Watkins, Joshua D Lawson, W Paul Segars, Laura I Cerviño, William Y Song, Steve B Jiang.   

Abstract

Algorithms for direct tumor tracking in rotational cone-beam projections and for reconstruction of phase-binned 3D tumor trajectories were developed. The feasibility of the algorithm was demonstrated on a digital phantom, a physical phantom and two patients. Tracking results were obtained by comparing reference templates generated from 4DCT to rotational cone-beam projections. The 95th percentile absolute errors (e(95)) in phantom tracking results did not exceed 1.7 mm in either imager dimension, while e(95) in the patients was 3.3 mm or less. Accurate phase-binned trajectories were reconstructed in each case, with 3D maximum errors of no more than 1.0 mm in the phantoms and 2.0 mm in the patients. This work shows the feasibility of a direct tumor tracking technique for rotational images, and demonstrates that an accurate 3D tumor trajectory can be reconstructed from relatively less accurate tracking results. The ability to reconstruct the tumor's average trajectory from a 3D cone-beam CT scan on the day of treatment could allow for better patient setup and quality assurance, while direct tumor tracking in rotational projections could be clinically useful for rotational therapy such as volumetric modulated arc therapy (VMAT).

Entities:  

Mesh:

Year:  2010        PMID: 20393232     DOI: 10.1088/0031-9155/55/9/006

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  30 in total

1.  A novel technique for markerless, self-sorted 4D-CBCT: feasibility study.

Authors:  Irina Vergalasova; Jing Cai; Fang-Fang Yin
Journal:  Med Phys       Date:  2012-03       Impact factor: 4.071

2.  3D fluoroscopic image estimation using patient-specific 4DCBCT-based motion models.

Authors:  S Dhou; M Hurwitz; P Mishra; W Cai; J Rottmann; R Li; C Williams; M Wagar; R Berbeco; D Ionascu; J H Lewis
Journal:  Phys Med Biol       Date:  2015-04-23       Impact factor: 3.609

3.  A novel phantom for characterization of dual energy imaging using an on-board imaging system.

Authors:  Maksat Haytmyradov; Rakesh Patel; Hassan Mostafavi; Murat Surucu; Adam Wang; Matthew M Harkenrider; John C Roeske
Journal:  Phys Med Biol       Date:  2019-01-21       Impact factor: 3.609

4.  A Bayesian approach to real-time 3D tumor localization via monoscopic x-ray imaging during treatment delivery.

Authors:  Ruijiang Li; Benjamin P Fahimian; Lei Xing
Journal:  Med Phys       Date:  2011-07       Impact factor: 4.071

5.  Multiple anatomy optimization of accumulated dose.

Authors:  W Tyler Watkins; Joseph A Moore; James Gordon; Geoffrey D Hugo; Jeffrey V Siebers
Journal:  Med Phys       Date:  2014-11       Impact factor: 4.071

6.  Registration of clinical volumes to beams-eye-view images for real-time tracking.

Authors:  Jonathan H Bryant; Joerg Rottmann; John H Lewis; Pankaj Mishra; Paul J Keall; Ross I Berbeco
Journal:  Med Phys       Date:  2014-12       Impact factor: 4.071

7.  Implementation and experimental results of 4D tumor tracking using robotic couch.

Authors:  I Buzurovic; Y Yu; M Werner-Wasik; T Biswas; P R Anne; A P Dicker; T K Podder
Journal:  Med Phys       Date:  2012-11       Impact factor: 4.071

8.  Markerless tumor tracking using fast-kV switching dual-energy fluoroscopy on a benchtop system.

Authors:  Maksat Haytmyradov; Hassan Mostafavi; Adam Wang; Liangjia Zhu; Murat Surucu; Rakesh Patel; Arun Ganguly; Michelle Richmond; Roberto Cassetta; Matthew M Harkenrider; John C Roeske
Journal:  Med Phys       Date:  2019-06-01       Impact factor: 4.071

9.  Markerless EPID image guided dynamic multi-leaf collimator tracking for lung tumors.

Authors:  J Rottmann; P Keall; R Berbeco
Journal:  Phys Med Biol       Date:  2013-05-28       Impact factor: 3.609

10.  Robust fluoroscopic tracking of fiducial markers: exploiting the spatial constraints.

Authors:  Rui Li; Gregory Sharp
Journal:  Phys Med Biol       Date:  2013-02-26       Impact factor: 3.609

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.