[Transpalpebral measurement of axial eye length. Use of contact B-scan sonography].

P Martius1, S Tech, O Stachs, R F Guthoff.   

Abstract

PURPOSE: In ophthalmology, it is very often necessary to determine the axial eye length. When standard measuring techniques fail, B-scan sonography may be used. However, coupling problems inhibit exact measurement. The aim of this study was to counter these problems. PATIENTS AND METHODS: Correction values were created for the ultrasonic devices (1) Master-Vu (Sonomed), (2) CineScan (Quantel Medical), and (3) System-ABD (Innovative Imaging), including the position of the coupling layer of the hand pieces and the lid thickness. For evaluation of the method, the axial eye lengths of 35 eyes were measured transpalpebrally. These data were compared to those from the (4) IOL Master (Carl Zeiss Meditec). The usefulness of the developed method was shown by two patients.
RESULTS: The correction values were (1) 3.0 mm, (2) 1.5 mm, and (3) 1.5 mm. The medial aberrations to (4) were (1) 0.4+/-0.5 mm, (2) 0.9+/-0.7 mm, and (3) 0.2+/-0.7 mm.
CONCLUSIONS: The introduced method enables transpalpebral eye length measurement and is also applicable in cases of missing compliance or in eyes with an obfuscated optical medium.

Entities:  

Mesh:

Year:  2010        PMID: 20376463     DOI: 10.1007/s00347-009-2096-9

Source DB:  PubMed          Journal:  Ophthalmologe        ISSN: 0941-293X            Impact factor:   1.059


  2 in total

1.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

2.  [Congenital clinical anophthalmia and blind microphthalmia].

Authors:  M P Schittkowski; K K H Gundlach; R F Guthoff
Journal:  Ophthalmologe       Date:  2003-06-21       Impact factor: 1.059

  2 in total
  1 in total

1.  Measuring myofiber orientations from high-frequency ultrasound images using multiscale decompositions.

Authors:  Xulei Qin; Baowei Fei
Journal:  Phys Med Biol       Date:  2014-06-24       Impact factor: 3.609

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.