Olivia M Flannery1, John R Britton, Peter O'Reilly, Nicholas Mahony, Patrick J Prendergast, Paddy J Kenny.
Abstract
BACKGROUND ANDEntities:
Mesh:
Substances:
Year: 2010 PMID: 20367418 PMCID: PMC2876831 DOI: 10.3109/17453674.2010.480936
Source DB: PubMed Journal: Acta Orthop ISSN: 1745-3674 Impact factor: 3.717
Figure 1.Impaction using the X-Change III proximal impactor 37.5 (2) attached to a biaxial hydraulic fatigue testing machine.
Figure 2.The X-Change III slap hammer used in revision hip arthroplasty was impacted against the load cell. The maximum load obtained was 3.5 kN.
Figure 3.Axial cyclic loading between 440 N (swing phase of gait) and 1,320 N (stance phase of gait) through the 28-mm femoral head, by direct contact with the hydraulic fatigue testing machine via a PTFE disc, which was free to slide along a PTFE plate.
Characteristics of the 29 femurs
| Test | Load at # (kN) | Strain at # (istrain) | Cortical thickness at # (mm) | Cortical thickness: canal diameter | BMD (g.cm²) |
|---|---|---|---|---|---|
| 1 | 7.5 | 6050 | 3.7 | 0.14 | 1.02 |
| 2 | 7.5 | 4330 | 5.0 | 0.21 | 1.11 |
| 3 | 6.5 | 2030 | 4.1 | 0.15 | 1.41 |
| 4 | 9 | 15800 | 3.7 | 0.14 | 1.04 |
| 5 | 8.5 | 2670 | 3.6 | 0.15 | 1.03 |
| 6 | 4 | 2250 | 3.6 | 0.18 | 1.15 |
| 7 | 6 | 6500 | 3.3 | 0.15 | 1.09 |
| 8 | 10 | 9190 | 3.4 | 0.14 | 1.03 |
| 9 | 7 | 1870 | 3.0 | 0.13 | 0.95 |
| 10 | 8 | 5190 | 3.7 | 0.15 | 0.96 |
| 11 | 9.5 | 4620 | 3.5 | 0.14 | 0.93 |
| 12 | 5.5 | 1820 | 3.3 | 0.14 | 1.02 |
| 13 | 8.5 | 3300 | 5.3 | 0.22 | 1.26 |
| 14 | 6 | 5950 | 3.6 | 0.16 | 1.05 |
| 15 | 8.5 | 1810 | 3.2 | 0.12 | 1.26 |
| 16 | 5.5 | 7330 | 3.0 | 0.13 | 1.12 |
| 17 | 9.5 | 5960 | 4.6 | 0.19 | 1.35 |
| 18 | 6.5 | 5120 | 3.6 | 0.20 | 1.04 |
| 19 | 6.5 | 6900 | 3.6 | 0.15 | 1.39 |
| 20 | 8 | 2700 | 3.8 | 0.17 | 1.40 |
| 21 | 6 | 4580 | 4.1 | 0.16 | 1.33 |
| 22 | 5 | 1040 | 4.9 | 0.20 | 1.11 |
| 23 | 8.5 | 4390 | 4.3 | 0.15 | 1.04 |
| 24 | 8.5 | 5030 | 4.4 | 0.18 | 1.22 |
| 25 | 10 | 4100 | 3.9 | 0.16 | 1.25 |
| 26 | 8 | 8940 | 4.3 | 0.18 | 1.06 |
| 27 | 9.5 | 3910 | 4.0 | 0.15 | 1.17 |
| 28 | 10 | 11000 | 3.6 | 0.14 | 1.29 |
| 29 | 10 | 2660 | 4.0 | 0.18 | 1.20 |
Figure 4.Axial displacement of implant during swing phase of gait for each femur.