CT metal artefact reduction of total knee prostheses using angled gantry multiplanar reformation.

Mark Lewis1, Andoni P Toms, Karen Reid, William Bugg.   

Abstract

This study was designed to determine whether or not acquiring CT images of total knee prostheses by using an angled gantry and multiplanar reformation can reduce beam hardening artefact. A CT phantom was created with a total knee prosthesis suspended in gelatine with a known attenuation. CT data was acquired with a gantry angled at 0 degrees, 5 degrees, 10 degrees and 15 degrees in both craniocaudal oblique planes. Axial images where then reformatted from these datasets. Two independent observers selected regions of interest to measure the mean and standard deviation (SD) of attenuation in the gelatine for all reformatted axial images. Artefact was measured as SD of the background attenuation and areas under the curve of SD for each gantry angle acquisition were compared. Inter-observer reliability was excellent (ICC=0.89, CI 0.875-0.908). The most accurate mean attenuation values for tissues around a TKR were obtained with a CT gantry using 10 degrees to 15 degrees anteroinferior to posterosuperior angulation. The smallest area under the curve for SD of attenuation for the whole prosthesis, and the femoral component in isolation, was obtained with a 5 degrees gantry angle in the same direction. The smallest area under the curve for the tibial component in isolation occurred with a gantry angle of 15 degrees. We conclude that acquiring CT data with a gantry angle can reduce metal artefact around a TKR. Optimal overall metal artefact reduction can be achieved with a small angle from anteroinferior to posterosuperior. Further selective artefact reduction around the tibial component can be achieved with larger angles. Crown Copyright 2010. Published by Elsevier B.V. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20356751     DOI: 10.1016/j.knee.2010.02.007

Source DB:  PubMed          Journal:  Knee        ISSN: 0968-0160            Impact factor:   2.199


  6 in total

1.  Task-driven source-detector trajectories in cone-beam computed tomography: I. Theory and methods.

Authors:  J Webster Stayman; Sarah Capostagno; Grace J Gang; Jeffrey H Siewerdsen
Journal:  J Med Imaging (Bellingham)       Date:  2019-05-02

2.  Orthopaedic and non-orthopaedic applications of a single-energy iterative metal artefact reduction technique and other metal artefact reduction techniques explained.

Authors:  Robert Khor; Kevin Buchan; Ahilan Kuganesan; Nicholas Ardley; Kenneth K Lau
Journal:  Br J Radiol       Date:  2016-05-25       Impact factor: 3.039

3.  Quantitative analysis of metal artefacts of dental implant in CBCT image by correlation analysis to micro-CT: A microstructural study.

Authors:  Chang-Ki Min; Kyoung-A Kim
Journal:  Dentomaxillofac Radiol       Date:  2020-10-08       Impact factor: 2.419

4.  Relationship Between Glenoid Component Shift and Osteolysis After Anatomic Total Shoulder Arthroplasty: Three-Dimensional Computed Tomography Analysis.

Authors:  Eric T Ricchetti; Bong-Jae Jun; Yuxuan Jin; Jason C Ho; Thomas E Patterson; Jarrod E Dalton; Kathleen A Derwin; Joseph P Iannotti
Journal:  J Bone Joint Surg Am       Date:  2021-08-04       Impact factor: 6.558

5.  Dual-energy CT and ceramic or titanium prostheses material reduce CT artifacts and provide superior image quality of total knee arthroplasty.

Authors:  Maximilian F Kasparek; Michael Töpker; Mathias Lazar; Michael Weber; Michael Kasparek; Thomas Mang; Paul Apfaltrer; Bernd Kubista; Reinhard Windhager; Helmut Ringl
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2018-06-07       Impact factor: 4.342

6.  Metal artifact reduction on cervical CT images by deep residual learning.

Authors:  Xia Huang; Jian Wang; Fan Tang; Tao Zhong; Yu Zhang
Journal:  Biomed Eng Online       Date:  2018-11-27       Impact factor: 2.819

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.